Recall Z-3249-2026

Philips Ultrasound, LLC · initiated August 7, 2026

Open, Classified

Product

Philips EPIQ Elite Diagnostic Ultrasound System Model # 795234

Codes / lots: All Philips EPIQ Elite Ultrasound Systems running software version 14.0, UDI(s): (01)00884838102514 (01)00884838107540 (01)00884838114159 (01)00884838124912 (01)00884838128125

Reason for recall

Due to software issue with ultrasound system, doppler sample is displayed shallower than its actual measurement location, potentially causing inaccurate blood flow velocity measurements and misdiagnosis.

FDA-determined cause: Software Design Change

Action taken

On August 7, 2026, Philips issued a Urgent Medical Device Correction Notification to affected consignees. Philips requested consignees take the following actions: 1. Due to possibility of a weak PW Doppler Signal on EPIQ Elite running 14.0 SW, customers are advised not to use C5-1 Abd Vasc Tissue Specific Preset (TSP). Please use a different TSP (e.g., Abd Gen, Abd Renal) within C5-1. Please note that this issue (i.e., Weak PW Doppler signal) is restricted to the Abd Vasc TSP with C5-1 on EPIQ Elite running SW 14.0 only. 2. Otherwise, you may continue to use your device in accordance with the IFU. 3.Circulate this notice to all users of the device so that they are aware of the product issue and associated hazard / harm. 4. Post this notice near the affected unit(s) for ease of reference 5. Please return the completed response form at the end of this notice, indicating that information on this issue has been read and understood. Philips asks that you return the completed response form no later than 30 days from receipt of this letter.

Details

Recall number
Z-3249-2026
Initiated
August 7, 2026
Posted by FDA
September 18, 2026
Product code
IYN – System, Imaging, Pulsed Doppler, Ultrasonic
Quantity
219 units
Distribution
Worldwide - US Nationwide distribution in the states of AL, CA, CT, FL, GA, KS, MD, MN, NJ, NY, OH, PA, SC, TN, TX, VA, WA, WI and the countries of Austria, Denmark, Finland, France, Germany, Greece, Italy, Japan, Lithuania, Netherlands, Norway, Philippines, Poland, Romania, Slovenia, South Korea, Spain, Sweden, Switzerland, Taiwan , United Kingdom, Vietnam.
Recalling firm
Philips Ultrasound, LLC, Bothell, WA

Official FDA recall record