Recall Z-3252-2024

Cook Biotech, Inc. · initiated August 7, 2024

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Biodesign Fistula Plug C-FPS-0.2 - For implantation to reinforce soft tissue for repair of recto-vaginal or anorectal fistulas. Order Number (GPN): G54612

Codes / lots: UDI: (01)00827002546127(17) Lot Number: LB1524461 LB1524469 LB1524472 LB1524486 LB1529402 LB1531398 LB1537840 LB1537858

Reason for recall

Products expire prior to the expiration date printed on the product labeling

FDA-determined cause: Incorrect or no expiration date

Action taken

Cook Biotech Inc. issued Urgent: Medical Device Recall Letter on 8/7/24. Letter states reason for recall, health risk and action to take: 1. Examine inventory immediately to determine if you have affected product(s), and quarantine affected product(s). Immediately cease all distribution and use of Biodesign Fistula Plugs with Lot Numbers provided in the Product Information section of this letter. 2. Return the affected product(s) to Cook Medical with a copy of the Acknowledgement and Receipt Form to receive a product credit. Note: Unaffected products that are returned will not be credited. 3. Please complete the Acknowledgement and Receipt Form within 5 business days of receiving this letter. Even if you do not have affected product(s) on hand, you must still complete the Acknowledgement and Receipt Form and return via fax (812.339.7316) or email (FieldActionsNA@CookMedical.com). 4. This notice must be shared with appropriate personnel, including down to the user level, within your organization or with any organization where the potentially affected devices have been transferred. 5. Immediately report adverse events to Cook Medical Customer Relations by phone at 800.457.4500 or 812.339.2235, Monday through Friday between 7:30am and 5:00pm (Eastern Time), or by email to: CustomerRelationsNA@CookMedical.com. Cook Biotech initiated CAPA 2024-CR-002 for this issue. If you have any questions or concerns, please contact Cook Medical Customer Relations at 800.457.4500 or 812.339.2235.

Details

Recall number
Z-3252-2024
Initiated
August 7, 2024
Posted by FDA
September 25, 2024
Product code
FTM – Mesh, Surgical
Quantity
35 units
Distribution
Worldwide distribution - US Nationwide and the countries of Australia, Austria, Canada, China, Finland, France, Germany, Ireland, Luxembourg, Netherlands, New Zealand, Norway, South Africa, United Kingdom.
Recalling firm
Cook Biotech, Inc., W Lafayette, IN

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-3250-2024 Class IIBiodesign Fistula Plug C-FPB-0.4 - For implantation to reinforce soft tissue for repair…Products expire prior to the expiration date printed on the product labeling 08/07/2024Open, Classified
Z-3251-2024 Class IIBiodesign Fistula Plug C-FPB-0.7 - For implantation to reinforce soft tissue for repair…Products expire prior to the expiration date printed on the product labeling 08/07/2024Open, Classified
Z-3253-2024 Class IIBiodesign Fistula Plug C-FPS-0.4 - For implantation to reinforce soft tissue for repair…Products expire prior to the expiration date printed on the product labeling 08/07/2024Open, Classified
Z-3254-2024 Class IIBiodesign Fistula Plug C-FPS-0.7 - For implantation to reinforce soft tissue for repair…Products expire prior to the expiration date printed on the product labeling 08/07/2024Open, Classified
Z-3255-2024 Class IIBiodesign Fistula Plug C-FPS-0.2-2 - For implantation to reinforce soft tissue for…Products expire prior to the expiration date printed on the product labeling 08/07/2024Open, Classified
Z-3256-2024 Class IIBiodesign Fistula Plug C-FPS-0.4-2 - For implantation to reinforce soft tissue for…Products expire prior to the expiration date printed on the product labeling 08/07/2024Open, Classified
Z-3257-2024 Class IIBiodesign Fistula Plug C-FPS-0.7-2 - For implantation to reinforce soft tissue for…Products expire prior to the expiration date printed on the product labeling 08/07/2024Open, Classified
Z-3258-2024 Class IIBiodesign Recto-Vaginal Fistula Plug RVP-0.2 - For implantation to reinforce soft tissue…Products expire prior to the expiration date printed on the product labeling 08/07/2024Open, Classified
Z-3259-2024 Class IIBiodesign Recto-Vaginal Fistula Plug RVP-0.4 - For implantation to reinforce soft tissue…Products expire prior to the expiration date printed on the product labeling 08/07/2024Open, Classified
Z-3260-2024 Class IIBiodesign Recto-Vaginal Fistula Plug RVP-0.7 - For implantation to reinforce soft tissue…Products expire prior to the expiration date printed on the product labeling 08/07/2024Open, Classified
Z-3261-2024 Class IIBiodesign Inguinal Hernia Graft C-IHM-10X15 - To reinforce soft tissues in the inguinal…Products expire prior to the expiration date printed on the product labeling 08/07/2024Open, Classified