Recall Z-3286-2011

Hospira Inc. · initiated September 16, 2011

Terminated

Product

Plum A+ Hyperbaric Single Channel Infusion Pumps; Hospira, Inc., Lake Forest, IL 60045; list 11005. Hospira has received reports of incorrect seating of the regulator closer on the Plum A+ Infusion Pumps. If the regulator closer does not close the flow regulator actuator when the pump door is opened, and if the clinician has not engaged the clamp prior to opening the cassette door, unrestricted flow may occur.

Codes / lots: list 11005, serial numbers 0016792207 through 0018807700

Reason for recall

Hospira has received reports of incorrect seating of the regulator closer on the Plum A+ Infusion Pumps. If the regulator closer does not close the flow regulator actuator when the pump door is opened, and if the clinician has not engaged the clamp prior to opening the cassette door, unrestricted flow may occur.

FDA-determined cause: Nonconforming Material/Component

Action taken

Hospira, Inc. sent an "URGENT DEVICE RECALL" letter dated September 16, 2011 to all affected customers. The letter describes the product, problem and actions to be taken by the customers. The letter instructs customers on how to prevent occurrences of continuous recycling/rebooting during start-up. A reply form was attached to the letter for customers to complete and return via fax to 1-888-345-5358. Contact Hospira Advanced Knowledge Center at 1-800-241-4002, option 4, for technical assistance or information regarding this notice.

Details

Recall number
Z-3286-2011
Initiated
September 16, 2011
Posted by FDA
September 29, 2011
Terminated
November 7, 2016
Product code
FRN – Pump, Infusion
Quantity
120 pumps
Distribution
Worldwide Distribution-USA (nationwide) including Puerto Rico and the U. S. Virgin Islands and countries of Argentina, Australia, Bahamas, Barbados, Belgium, Brazil, Canada, Chile, Colombia, Costa Rica, Denmark, Dominican Republic, Ecuador, France, Greece, Hong Kong, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Korea, Kuwait, Lebanon, Libya, Malaysia, Mexico, Netherlands, New Zealand, Oman…
Recalling firm
Hospira Inc., Lake Forest, IL

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-3284-2011 Plum A+ Single Channel Infusion Pumps; Hospira, Inc., Lake Forest, IL 60045; the pumps…Hospira has received reports of incorrect seating of the regulator closer on the Plum A+ Infusion Pumps. If… 09/16/2011Terminated
Z-3285-2011 Plum A+3 Triple Channel Infusion Pumps; Hospira, Inc., Lake Forest, IL 60045; the pumps…Hospira has received reports of incorrect seating of the regulator closer on the Plum A+ Infusion Pumps. If… 09/16/2011Terminated