Recall Z-3327-2018

B. Braun Medical, Inc. · initiated July 20, 2018

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

CONT STIM BASIC KIT W/4 IN INSUL TUOHY, Material Number 331758 Connection device used by physicians to provide various anesthetic and fluid administration devices with a single, common access point to a catheter for delivery of anesthetics. The connector is used in conjunction with the catheter for continuous administration of anesthetic fluids.

Codes / lots: Batch: 61606946, 61613536, 61619213

Reason for recall

The catheter connectors may not stay closed during in and in some cases, leakage or disconnection of the catheters from the catheter connectors has been observed.

FDA-determined cause: Under Investigation by firm

Action taken

The firm, B. Braun, sent an "URGENT MEDICAL DEVICE CORRECTION RECALL NOTIFICATION" letters dated 7/20/18 to its customers. The letters described the product, problem and actions to be taken. The letters instruct customers to do the following: 1. Review this Medical Device Correction Notification in its entirety and ensure that all users of the mentioned products in your organization and other concerned persons are informed about this Medical Device Correction Recall Notification. If you are a distributor, please forward this recall notification to your customers. 2. For continued safe use of the Perifix Catheter Connector as part, review and follow the instructions described in Attachment 2, Step-by-Step Correction of the Perifix Catheter Connector using Label. 3. For continued safe use of the Perifix Catheter Connector, review and follow the instructions described in Attachment 3, Step-by-Step Correction of the Perifix Catheter Connector using Cloth/Silk Medical Tape. This method can be applied to the Perifix Catheter Connector in Procedural Trays. 4. Utilizing the attached Medical Device Correction Recall Notification Acknowledgement form, please acknowledge that you have received and reviewed this information. 5. Return the completed Medical Device Correction Recall Notification Acknowledgement form to B. Braun Medical Inc. Quality Assurance department by faxing to: 610-849-1197 or e-mailing to: PA_QualityAssurance.BBMUS_Service@bbraunusa.com the form within two (2) weeks of receipt of this notice. It is important this form is returned so B. Braun Medical Inc. can meet regulatory requirements of the United States Food and Drug Administration. 6. If you have any affected products as identified in attachment 1 that you choose to return instead of following the continued safe use instructions within Attachment 2 and 3, please identify on the Medical Device Correction Recall Notification Acknowledgement form that you will be returning t

Details

Recall number
Z-3327-2018
Initiated
July 20, 2018
Terminated
May 13, 2020
Product code
BSO – Catheter, Conduction, Anesthetic
Quantity
430
Distribution
Worldwide Distribution: US (nationwide) and countries of: Australia, Canada, Guatemala and Kuwait.
Recalling firm
B. Braun Medical, Inc., Allentown, PA

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-3313-2018 Class IICONTI TUOHY ULTRA 360 2 IN CONT PNB SET, Material Number 331640 Connection device used…The catheter connectors may not stay closed during in and in some cases, leakage or disconnection of the… 07/20/2018Terminated
Z-3314-2018 Class IICNBFX350O CONTIPLEX CONT NERVE BLOCK, Material Number 331670 Connection device used by…The catheter connectors may not stay closed during in and in some cases, leakage or disconnection of the… 07/20/2018Terminated
Z-3315-2018 Class IICNBFX350C CONTIPLEX CONT NERVE BLOCK, Material Number 331672 Connection device used by…The catheter connectors may not stay closed during in and in some cases, leakage or disconnection of the… 07/20/2018Terminated
Z-3316-2018 Class IICONTIPLEX TUOHY ULTRA 2 IN CONT. PNB SET, Material Number 331673 Connection device used…The catheter connectors may not stay closed during in and in some cases, leakage or disconnection of the… 07/20/2018Terminated
Z-3317-2018 Class IICNB200T CONTIPLEX TUOHY SET 2 IN, Material Number 331691 Connection device used by…The catheter connectors may not stay closed during in and in some cases, leakage or disconnection of the… 07/20/2018Terminated
Z-3318-2018 Class IICONTIPLEX TUOHY ULTRA 4 IN CNB400TU, LF, Material Number 331691 Connection device used…The catheter connectors may not stay closed during in and in some cases, leakage or disconnection of the… 07/20/2018Terminated
Z-3319-2018 Class IICONTIPLEX TUOHY SET, 18GX4", 1.3X100MM, Material Number 331693 Connection device used by…The catheter connectors may not stay closed during in and in some cases, leakage or disconnection of the… 07/20/2018Terminated
Z-3320-2018 Class IICONTIPLEX TUOHY SET, 18GX6", 1.3X150MM, Material Number 331694 Connection device used by…The catheter connectors may not stay closed during in and in some cases, leakage or disconnection of the… 07/20/2018Terminated
Z-3321-2018 Class IICONTIPLEX TUOHY SET, 18GX11/2X 1,3X40MM, Material Number 331695 Connection device used…The catheter connectors may not stay closed during in and in some cases, leakage or disconnection of the… 07/20/2018Terminated
Z-3322-2018 Class IICNB200S CONTIPLEX INSULATED TUOHY NDL, Material Number 331697 Connection device used by…The catheter connectors may not stay closed during in and in some cases, leakage or disconnection of the… 07/20/2018Terminated
Z-3323-2018 Class IICNB200TK CONTIPLEX CONT NERVE BLOCK, Material Number 331701 Connection device used by…The catheter connectors may not stay closed during in and in some cases, leakage or disconnection of the… 07/20/2018Terminated
Z-3324-2018 Class IICNB400TK CONTIPLEX CONT NERVE BLOCK, Material Number 331706 Connection device used by…The catheter connectors may not stay closed during in and in some cases, leakage or disconnection of the… 07/20/2018Terminated
Z-3325-2018 Class IISCNB4TK CONT STIM W/4 INCH INSUL TUOHY, Material Number 331756 Connection device used by…The catheter connectors may not stay closed during in and in some cases, leakage or disconnection of the… 07/20/2018Terminated
Z-3326-2018 Class IICONT STIM BASIC KIT W/2 IN INSUL TUOHY, Material Number 331757 Connection device used by…The catheter connectors may not stay closed during in and in some cases, leakage or disconnection of the… 07/20/2018Terminated
Z-3328-2018 Class IICONTIPLEX ECHO OT W/4" TUOHY ULTRA SET, Material Number 331762 Connection device used by…The catheter connectors may not stay closed during in and in some cases, leakage or disconnection of the… 07/20/2018Terminated
Z-3329-2018 Class IICONTI ECHO CT W/2IN TUOHY ULTRA 360 SET, Material Number 331767 Connection device used…The catheter connectors may not stay closed during in and in some cases, leakage or disconnection of the… 07/20/2018Terminated
Z-3330-2018 Class IICE17TKFCY 17GA SAFETY TUOHY NEEDLE TRAY, Material Number 332075 Connection device used…The catheter connectors may not stay closed during in and in some cases, leakage or disconnection of the… 07/20/2018Terminated
Z-3331-2018 Class IICE17TBFC EPIDURAL BASIC TRAY 17G TOUHY, Material Number 332077 Connection device used by…The catheter connectors may not stay closed during in and in some cases, leakage or disconnection of the… 07/20/2018Terminated
Z-3332-2018 Class IICE17TKFCPS EPIDURAL FULL TRAY, Material Number 332078 Connection device used by…The catheter connectors may not stay closed during in and in some cases, leakage or disconnection of the… 07/20/2018Terminated
Z-3333-2018 Class IICE17TKFC EPIDURAL TRAY W/ 17GA TUOHY, Material Number 332079 Connection device used by…The catheter connectors may not stay closed during in and in some cases, leakage or disconnection of the… 07/20/2018Terminated
Z-3334-2018 Class IICE17TKFSA FULL TRAY W/17G TUOHY 19G SPRW, Material Number 332080 Connection device used…The catheter connectors may not stay closed during in and in some cases, leakage or disconnection of the… 07/20/2018Terminated
Z-3335-2018 Class IICE18TKST CONT EPID W/SOFT TIP, Material Number 332081 Connection device used by…The catheter connectors may not stay closed during in and in some cases, leakage or disconnection of the… 07/20/2018Terminated
Z-3336-2018 Class IICE17TKST CONT EPID W/SOFT TIP, Material Number 332082 Connection device used by…The catheter connectors may not stay closed during in and in some cases, leakage or disconnection of the… 07/20/2018Terminated
Z-3337-2018 Class IICE18HKST CONT EPID 20G SOFTIP, Material Number 332084 Connection device used by…The catheter connectors may not stay closed during in and in some cases, leakage or disconnection of the… 07/20/2018Terminated
Z-3338-2018 Class IICE17TKF PERIFIX CONTIN. EPIDURAL TRAY, Material Number 332086 Connection device used by…The catheter connectors may not stay closed during in and in some cases, leakage or disconnection of the… 07/20/2018Terminated
Z-3339-2018 Class IICE17TKFPS PERIFIX FX CONT. EPID TRAY-LF, Material Number 332087 Connection device used…The catheter connectors may not stay closed during in and in some cases, leakage or disconnection of the… 07/20/2018Terminated
Z-3340-2018 Class IICE17TKFSDT EPID TRAY W/17GA FIXED WING, Material Number 332093 Connection device used by…The catheter connectors may not stay closed during in and in some cases, leakage or disconnection of the… 07/20/2018Terminated
Z-3341-2018 Class IICE17TKFSDTD EPID TRAY W/17GA FIXED WING, Material Number 332095 Connection device used…The catheter connectors may not stay closed during in and in some cases, leakage or disconnection of the… 07/20/2018Terminated
Z-3342-2018 Class IICE17TKFMHDT 17 GA X 4 FULL EPID KIT, Material Number 332096 Connection device used by…The catheter connectors may not stay closed during in and in some cases, leakage or disconnection of the… 07/20/2018Terminated
Z-3343-2018 Class IICE17TKFCS CONT EPID TRAY W/17G TUOHY-LF, Material Number 332097 Connection device used…The catheter connectors may not stay closed during in and in some cases, leakage or disconnection of the… 07/20/2018Terminated