Processor, Tissue, Automated

FDA product code IEO · Class I (low risk) · 21 CFR 864.3875 · Pathology

12510(k) clearances
0PMA approvals
11recalls

Manufacturers

510(k) clearances

510(k)DeviceApplicantDecision
K924590Microwave Automated Tissue ProcessorIEO · Traditional Innovative Medical Systems, Inc.12/29/1992SE
K883599LX-80 Automated Tissue Processor (Plastic)IEO · Traditional Innovative Medical Systems, Inc.09/02/1988SE
K881165Sakura HEM-480 Electron Microscopy ProcessorIEO · Traditional Sakura Finetek U.S.A., Inc.07/01/1988SE
K853938LX 300 Tissue ProcessorIEO · Traditional Innovative Medical Systems, Inc.10/10/1985SE
K851054LX120 Tissue ProcessorIEO · Traditional Innovative Medical Systems, Inc.05/15/1985SE
K843506LKB 2189 UltroprocessorIEO · Traditional Lkb Instruments, Inc.11/30/1984SE
K834599Autotechnicon V Tissue Process SysIEO · Traditional Technicon Instruments Corp.02/04/1984SE
K831313Tissue-Tek V.I.P. 100/200IEO · Traditional Miles Laboratories, Inc.07/28/1983SE
K801513Tissue Processing SystemIEO · Traditional Innovative Medical Systems, Inc.07/28/1980SE
K801427Tissue-Tek III Mini-Rapid ProcessorIEO · Traditional Miles Laboratories, Inc.07/28/1980SE
K771308Tissue-Tek III V.I.P.IEO · Traditional Miles Laboratories, Inc.08/22/1977SE
K761062Histo-Tek W/ Automatic CoverslipperIEO · Traditional Miles Laboratories, Inc.03/07/1977SE

Recalls

RecallProductFirmInitiated
Z-1683-2025 Class IIHistoCore PELORIS 3, Model/Catalog Number: 45.0005/45.7512.501 A11, Software Version:…There is a leakage issue associated with the tubing in the manifold of the instrument. Leica Biosystems Melbourne Pty Ltd04/02/2025Open, Classified
Z-1947-2024 Class IILeica HistoCore PEGASUS Tissue Processor, REF 14048858005.Poorly processed and/or damaged biopsy tissue specimens on the device resulting from reagent levels… Leica Biosystems Nussloch GmbH04/10/2024Open, Classified
Z-0590-2024 Class IILeica HistoCore PEGASUS (PROCESSOR, TISSUE, AUTOMATED)There is an issue regarding poorly processed and/or damaged biopsy tissue specimens on the HistoCore PEGASUS… Leica Biosystems Nussloch GmbH11/20/2023Open, Classified
Z-2660-2017 Class IINussloch GmbH ASP6025 Tissue ProcessorIncorrect labeling for specified voltage for the Alarm Connectors on the rear side of the unit. Leica Microsystems, Inc.06/19/2017Open, Classified
Z-2779-2015 Class IIIThe Leica ASP200S/ASP300S Tissue Processor. Pathology and Hematology Departments. The…The Leica ASP200S/ASP300S Tissue Processor have an incorrect internal wiring to the remote alarm plug, which… Leica Microsystems, Inc.08/05/2015Terminated
Z-2804-2015 Class IIThe Leica ASP6025 Tissue Processor. Pathology and Hematology Departments. The Leica…The ASP6025 Tissue Processor instrument is incorrectly getting into a state of bottle empty during retort… Leica Microsystems, Inc.07/27/2015Terminated
Z-1815-2013 Class IILeica ASP6025 Vacuum Tissue Processor The Leica ASP6025 is a modular tissue processor…The Instruction for Use 1v9 Ref F and all previous versions specify incorrect dimensions for the Ready To… Leica Microsystems, Inc.11/13/2012Terminated
Z-1473-2012 ASP 6025 Vacuum Tissue Processor; an automated modular tissue processor; Leica…There is a potential malfunction with the device in autorotation mode, which can lead to incorrect reagent… Leica Microsystems, Inc.03/30/2012Terminated
Z-2597-2011 Peloris II Rapid Tissue Processor; Leica Biosystems Melbourne Pty Ltd, Mount Waverly…The processor potentially has a faulty impellor which could lead to reagent carry-over from one processing… Leica Microsystems, Inc.04/11/2011Terminated
Z-1551-2010 Peloris Rapid Tissue Processor; Leica Biosystems Melbourne Pty Ltd, Mount Waverly, 3149…Confusion during manually changing reagents, general users running unvalidated protocols, and wax valve… Leica Microsystems, Inc.02/09/2010Terminated
Z-1320-2010 ThermoFisher Scientific Microm STP 420D automated tissue processor.The firm has received complaints regarding specimen integrity (higher rate of specimen fragmentation) as… Thermo Fisher Scientific06/03/2009Terminated