Recall Z-2660-2017

Leica Microsystems, Inc. · initiated June 19, 2017

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Nussloch GmbH ASP6025 Tissue Processor

Codes / lots: Serial No. 200, 202, 204, 206, 208, 210, 212, 214, 216, 218, 220, 222, 224, 226, 228, 230, 232, 234, 238, 240, 242, 248, 250, 252, 254, 256, 258, 260, 262, 264, 266, 268, 270, 272, 274, 276, 278, 280, 282, 284, 286, 290, 292, 294, 296, 298, 300, 302, 304, 308, 312, 320, 334, 336, 338, 340, 342, 352, 354, 356, 358, 360, 362, 364, 366, 368, 370, 372, 374, 378, 380, 382, 384, 386, 388, 396, 400, 402, 406, 408, 418, 420, 422, 424, 426, 428, 432, 434, 438, 440, 442, 444, 450, 452, 454, 456, 460, 462, 464, 468, 478, 480, 482, 484, 486, 488, 490, 492, 498, 502, 504, 516, 526, 530, 536, 544, 546, 548, 550, 554, 556, 558, 560, 562, 564, 570, 572, 574, 576, 580, 582, 584, 586, 588, 590, 598, 604, 612, 614, 628, 632, 636, 640, 642, 646, 650, 654, 660, 662, 666, 668, 674, 676, 684, 686, 692, 694, 696, 698, 700, 702, 704, 706, 712, 718, 720, 726, 728, 730, 732, 734, 736, 738, 742, 744, 746, 770, 784, 794, 796, 808, 810, 814, 816, 818, 820, 824, 834, 858, 860, 864, 866, 868, 870, 874, 878, 886, 892, 314, 316, 448, 538, 594, 622, 712, 778.

Reason for recall

Incorrect labeling for specified voltage for the Alarm Connectors on the rear side of the unit.

FDA-determined cause: Labeling mix-ups

Action taken

Customers were notified via email on 06/19/2017. Instructions to customers included to clean surface around the Alarm Connectors and place the newly provided label at the Alarm Connectors (over the engraved specification). They were also instructed to update their current Instructions For Use and update the online help function in the software, as well as return the response form.

Details

Recall number
Z-2660-2017
Initiated
June 19, 2017
Product code
IEO – Processor, Tissue, Automated
Quantity
812 units
Distribution
Domestic: AR, AZ, CA, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MN, MO, MT, NC, NE, NJ, NM, NV, NY, OH, PA, SC, TN, TX, UT, VA, WA, WI, WV, & DC. Foreign: Australia, Austria, Bangladesh, Belgium, Brazil, Canada, Chile, China, Colombia, Costa Rica, Czech Republic, England, France, Georgia, Germany, India, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Korea, Kuwait, Lithuania…
Recalling firm
Leica Microsystems, Inc., Buffalo Grove, IL