Recall Z-0070-2020

Becton, Dickinson and Company, BD Biosciences · initiated August 19, 2019

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

BD FACSLyric Flow Cytometer, 2-Laser 4 Color Instrument, REF: 651164, when using BD Trucount Tubes

Codes / lots: BD FACSLyric Flow Cytometer serial numbers: R651164000007, R651164000016, R651164000017, R651164000019, R651164000020, R651164000021, R651164000022, R651164000023, R651164000024, R651164000025, R651164000026, R651164000027, R651164000028, R651164000029, R651164000030, R651164000031, R651164000032, R651164000033, R651164000034, R651164000035, R651164000036.

Reason for recall

Flow cytometers, when used with specific tubes may experience excessive electronic abort counts (electronic abort count of beads exceeds 1.0% of the total processed events), which may affect the ratio of cell population events to bead events, potentially resulting in falsely high absolute counts. This may affect the following: initiation of prophylactic therapy for Opportunistic Infections (HIV patients), decisions regarding care/management (PID/immune suppressed patients), laboratory developed tests/experiments, specimen engraftment potency, and use of leucoreduced blood products for transfusion.

FDA-determined cause: Under Investigation by firm

Action taken

On 08/19/19, the recalling firm started calling and distributing Urgent Medical Device Recall notices to customers. Laboratory managers were asked to do the following: 1) Run the Abort Count Quantification Protocol. 2) Record the percentage of aborted events on the attached Customer Response Form and submit via fax, email, or online to the recalling firm regardless of the result. 3) For percent aborted events less-than 1.0 percent: your instrument is performing as expected and no further action is required once the percent aborted events has been provided to via Customer Response Form. 4) For percent aborted events greater-than-or-equal-to 1.0 percent: suspend all patient testing. A Technical Representative will contact you within 24 hours (Mon-Fri) of receiving the data (response form) to schedule a service visit. 5) Refer to the Frequently Asked Questions document for more information. Recalling firm is investigating the root cause and to identify actions to prevent recurrence. Customers with additional questions are encouraged to call: Customer/Technical Support at 844-918-0554.

Details

Recall number
Z-0070-2020
Initiated
August 19, 2019
Terminated
April 12, 2024
Product code
OYE – Flow Cytometric Reagents And Accessories.
Quantity
21
Distribution
U.S.: CA, KS, NJ, MI, NY, PA, IL, IN, VA, WA, MD, IA, OK, RI, FL, CO, OH, TX, TN, GA, MO, MN, HI, VT, MA, NE. O.U.S. (Foreign): Germany, Italy, South Korea, Japan, United Kingdom, Poland, France, Belgium, Canada, Latvia, India, Taiwan, Ireland, Algeria, Israel, Iraq, Netherlands, Malaysia, Australia, Sweden, Colombia, Switzerland, Slovakia, Spain, Brazil, Austria, Peru, Iran, Croatia, Lithuania…
Recalling firm
Becton, Dickinson and Company, BD Biosciences, San Jose, CA

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0066-2020 Class IIBD FACSLyric Flow Cytometer, 2-Laser 4 Color Instrument, REF: 662875, when using BD…Flow cytometers, when used with specific tubes may experience excessive electronic abort counts (electronic… 08/19/2019Terminated
Z-0067-2020 Class IIBD FACSLyric Flow Cytometer, 2-Laser 6 Color Instrument, REF: 662876, when using BD…Flow cytometers, when used with specific tubes may experience excessive electronic abort counts (electronic… 08/19/2019Terminated
Z-0068-2020 Class IIBD FACSLyric Flow Cytometer, 3-Laser 8 Color Instrument, REF: 662877, when using BD…Flow cytometers, when used with specific tubes may experience excessive electronic abort counts (electronic… 08/19/2019Terminated
Z-0069-2020 Class IIBD FACSLyric Flow Cytometer, 3-Laser 10 Color Instrument, REF: 662878, and 66287808 when…Flow cytometers, when used with specific tubes may experience excessive electronic abort counts (electronic… 08/19/2019Terminated
Z-0071-2020 Class IIBD FACSLyric Flow Cytometer, 2-Laser 6 Color Instrument, REF: 651165, when using BD…Flow cytometers, when used with specific tubes may experience excessive electronic abort counts (electronic… 08/19/2019Terminated
Z-0072-2020 Class IIBD FACSLyric Flow Cytometer, 3-Laser 8 Color Instrument, REF: 654587, when using BD…Flow cytometers, when used with specific tubes may experience excessive electronic abort counts (electronic… 08/19/2019Terminated
Z-0073-2020 Class IIBD FACSLyric Flow Cytometer,3-Laser 10 Color Instrument, REF: 659180 and 65918001, when…Flow cytometers, when used with specific tubes may experience excessive electronic abort counts (electronic… 08/19/2019Terminated
Z-0074-2020 Class IIBD FACSLyric Flow Cytometer, 3-Laser 12 Color Instrument, REF: 663029, when using BD…Flow cytometers, when used with specific tubes may experience excessive electronic abort counts (electronic… 08/19/2019Terminated