Recall Z-0074-2020
Becton, Dickinson and Company, BD Biosciences · initiated August 19, 2019
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
BD FACSLyric Flow Cytometer, 3-Laser 12 Color Instrument, REF: 663029, when using BD Trucount Tubes
Codes / lots: BD FACSLyric Flow Cytometer serial numbers: R663029000001, R663029000002, R663029000003, R663029000004, R663029000005, R663029000006, R663029000007, R663029000008, R663029000010, R663029000011, R663029000012, R663029000014, R663029000015, R663029000016, R663029000017, R663029000018, R663029000021, R663029000025, R663029000031, R659180000136, R659180000160, R659180000238.
Reason for recall
Flow cytometers, when used with specific tubes may experience excessive electronic abort counts (electronic abort count of beads exceeds 1.0% of the total processed events), which may affect the ratio of cell population events to bead events, potentially resulting in falsely high absolute counts. This may affect the following: initiation of prophylactic therapy for Opportunistic Infections (HIV patients), decisions regarding care/management (PID/immune suppressed patients), laboratory developed tests/experiments, specimen engraftment potency, and use of leucoreduced blood products for transfusion.
FDA-determined cause: Under Investigation by firm
Action taken
On 08/19/19, the recalling firm started calling and distributing Urgent Medical Device Recall notices to customers. Laboratory managers were asked to do the following: 1) Run the Abort Count Quantification Protocol. 2) Record the percentage of aborted events on the attached Customer Response Form and submit via fax, email, or online to the recalling firm regardless of the result. 3) For percent aborted events less-than 1.0 percent: your instrument is performing as expected and no further action is required once the percent aborted events has been provided to via Customer Response Form. 4) For percent aborted events greater-than-or-equal-to 1.0 percent: suspend all patient testing. A Technical Representative will contact you within 24 hours (Mon-Fri) of receiving the data (response form) to schedule a service visit. 5) Refer to the Frequently Asked Questions document for more information. Recalling firm is investigating the root cause and to identify actions to prevent recurrence. Customers with additional questions are encouraged to call: Customer/Technical Support at 844-918-0554.
Details
- Recall number
- Z-0074-2020
- Initiated
- August 19, 2019
- Terminated
- April 12, 2024
- Product code
- OYE – Flow Cytometric Reagents And Accessories.
- Quantity
- 22
- Distribution
- U.S.: CA, KS, NJ, MI, NY, PA, IL, IN, VA, WA, MD, IA, OK, RI, FL, CO, OH, TX, TN, GA, MO, MN, HI, VT, MA, NE. O.U.S. (Foreign): Germany, Italy, South Korea, Japan, United Kingdom, Poland, France, Belgium, Canada, Latvia, India, Taiwan, Ireland, Algeria, Israel, Iraq, Netherlands, Malaysia, Australia, Sweden, Colombia, Switzerland, Slovakia, Spain, Brazil, Austria, Peru, Iran, Croatia, Lithuania…
- Recalling firm
- Becton, Dickinson and Company, BD Biosciences, San Jose, CA
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-0066-2020 Class II | BD FACSLyric Flow Cytometer, 2-Laser 4 Color Instrument, REF: 662875, when using BD…Flow cytometers, when used with specific tubes may experience excessive electronic abort counts (electronic… | 08/19/2019Terminated |
| Z-0067-2020 Class II | BD FACSLyric Flow Cytometer, 2-Laser 6 Color Instrument, REF: 662876, when using BD…Flow cytometers, when used with specific tubes may experience excessive electronic abort counts (electronic… | 08/19/2019Terminated |
| Z-0068-2020 Class II | BD FACSLyric Flow Cytometer, 3-Laser 8 Color Instrument, REF: 662877, when using BD…Flow cytometers, when used with specific tubes may experience excessive electronic abort counts (electronic… | 08/19/2019Terminated |
| Z-0069-2020 Class II | BD FACSLyric Flow Cytometer, 3-Laser 10 Color Instrument, REF: 662878, and 66287808 when…Flow cytometers, when used with specific tubes may experience excessive electronic abort counts (electronic… | 08/19/2019Terminated |
| Z-0070-2020 Class II | BD FACSLyric Flow Cytometer, 2-Laser 4 Color Instrument, REF: 651164, when using BD…Flow cytometers, when used with specific tubes may experience excessive electronic abort counts (electronic… | 08/19/2019Terminated |
| Z-0071-2020 Class II | BD FACSLyric Flow Cytometer, 2-Laser 6 Color Instrument, REF: 651165, when using BD…Flow cytometers, when used with specific tubes may experience excessive electronic abort counts (electronic… | 08/19/2019Terminated |
| Z-0072-2020 Class II | BD FACSLyric Flow Cytometer, 3-Laser 8 Color Instrument, REF: 654587, when using BD…Flow cytometers, when used with specific tubes may experience excessive electronic abort counts (electronic… | 08/19/2019Terminated |
| Z-0073-2020 Class II | BD FACSLyric Flow Cytometer,3-Laser 10 Color Instrument, REF: 659180 and 65918001, when…Flow cytometers, when used with specific tubes may experience excessive electronic abort counts (electronic… | 08/19/2019Terminated |