Recall Z-0155-2020

Vital Scientific N.V. · initiated January 10, 2012

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Viva-E, Catalog # 6002-380, UDI # 03661540600050 Product Usage: automated chemistry analyzer intended for clinical use

Codes / lots: serial numbers: 10-4582 till 10-4584; 10-4586 till 10-4594; 11-2309; 11-2320 t through 11-2328; 11-2330 through-2334; 11-2336 through 11-2375; 11-2383; 11-2384; 11-2388; 11-2395; 12-2314; 12-2315; 12-2324 12-2325.

Reason for recall

The action was initiated due power supply failure. During a power supply failure the instruments malfunctions and no patient results are produced.

FDA-determined cause: Component design/selection

Action taken

The firm reached out to consignees via phone to inform them of the affected product.

Details

Recall number
Z-0155-2020
Initiated
January 10, 2012
Terminated
April 19, 2024
Product code
JJF – Analyzer, Chemistry, Micro, For Clinical Use
Quantity
75 units
Distribution
US Nationwide in the states of NY, and UT
Recalling firm
Vital Scientific N.V., Dieren

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0154-2020 Class IIViva - JR, catalog # 6002-940, UDI # 03661540600210 Product Usage: automated chemistry…The action was initiated due power supply failure. During a power supply failure the instruments… 01/10/2012Terminated
Z-0156-2020 Class IIV-Twin, Catalog # 6002-800, UDI # 03661540600180 Product Usage: automated chemistry…The action was initiated due power supply failure. During a power supply failure the instruments… 01/10/2012Terminated
Z-0157-2020 Class IISelectra ProM Analyzer, Catalog # 6003-435/UDI # 03661540600395 and Catalog #…The action was initiated due power supply failure. During a power supply failure the instruments… 01/10/2012Terminated
Z-0158-2020 Class IISelectra ProS Analyzer, Catalog # 6003-600/UDI # 03661540600418 Product Usage: :clinical…The action was initiated due power supply failure. During a power supply failure the instruments… 01/10/2012Terminated