Recall Z-0158-2020

Vital Scientific N.V. · initiated January 10, 2012

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Selectra ProS Analyzer, Catalog # 6003-600/UDI # 03661540600418 Product Usage: :clinical chemistry system intended for use in clinical laboratories.

Codes / lots: serial numbers: 11-9609; 11-9690; 11-9694; 11-9695; 12-9618; 12-9620.

Reason for recall

The action was initiated due power supply failure. During a power supply failure the instruments malfunctions and no patient results are produced.

FDA-determined cause: Component design/selection

Action taken

The firm reached out to consignees via phone to inform them of the affected product.

Details

Recall number
Z-0158-2020
Initiated
January 10, 2012
Terminated
April 19, 2024
Product code
JGS – Electrode, Ion Specific, Sodium
Quantity
6 instruments
Distribution
US Nationwide in the states of NY, and UT
Recalling firm
Vital Scientific N.V., Dieren

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0154-2020 Class IIViva - JR, catalog # 6002-940, UDI # 03661540600210 Product Usage: automated chemistry…The action was initiated due power supply failure. During a power supply failure the instruments… 01/10/2012Terminated
Z-0155-2020 Class IIViva-E, Catalog # 6002-380, UDI # 03661540600050 Product Usage: automated chemistry…The action was initiated due power supply failure. During a power supply failure the instruments… 01/10/2012Terminated
Z-0156-2020 Class IIV-Twin, Catalog # 6002-800, UDI # 03661540600180 Product Usage: automated chemistry…The action was initiated due power supply failure. During a power supply failure the instruments… 01/10/2012Terminated
Z-0157-2020 Class IISelectra ProM Analyzer, Catalog # 6003-435/UDI # 03661540600395 and Catalog #…The action was initiated due power supply failure. During a power supply failure the instruments… 01/10/2012Terminated