Recall Z-0166-2011

Siemens Medical Solutions USA, Inc · initiated March 31, 2010

Terminated

Product

Siemens brand 550 TxT Patient Table, Part No. 7346534, Distributed by and/or Manufactured by Siemens AG, Medical Solutions, Germany

Codes / lots: All codes.

Reason for recall

An increase in voltage may cause a short circuit of a power transistor that may cause a smoldering and smoking effect of the electrical wiring and also an unexpected lowering of the Table. Both issues present the potential for injury to a patient or the operator.

FDA-determined cause: Other

Action taken

Siemens sent a Customer Information Letter on March 31, 2010, to all consignees. The letter identified the product, the problem, and the action that should be taken. Siemens implemented minor modifications to address the issues. No further action was required for the continual operation of the 550TxT Table. For questions regarding this recall call 925-602-8083.

Details

Recall number
Z-0166-2011
Initiated
March 31, 2010
Posted by FDA
October 28, 2010
Terminated
March 1, 2012
Product code
JAI – Couch, Radiation Therapy, Powered
Quantity
326 units
Distribution
Worldwide Distribution including USA
Recalling firm
Siemens Medical Solutions USA, Inc, Concord, CA