Recall Z-0265-2014

Medtronic Vascular · initiated October 21, 2013

Class I Terminated

Class I – dangerous or defective products that could cause serious health problems or death

Product

ATTAIN HYBRID Guide Wire; Models: GWR419478, GWR419488, GWR419688 The Attain Hybrid guide wire is intended to aid in the placement of Medtronic transvenous left ventricular leads in the coronary vasculature.

Codes / lots: 0006752371 0006752383 0006752385 0006759937 0006759939 0006759941 0006769772 0006781453 0006781454 0006781455 0006808160 0006808161 0006808164 0006808165 0006808166 0006808167 0006831243 0006840866 0006840878 0006840879 0006866694 0006866696 0006866697 0006873688 0006873692 0006873694

Reason for recall

Potential for PTFE (polytetrafluroethylene) coating to delaminate and detach from Guidewire.

FDA-determined cause: Nonconforming Material/Component

Action taken

The firm, Medtronic, sent a "URGENT MEDICAL DEVICE RECALL" letter dated October 21, 2013 (in the US via UPS overnight service, 1-day delivery) to its customers. The letter described the product, problem and actions to be taken. The customers were instructed to immediately remove and quarantine all potentially affected guidewires that remain in inventory; return the affected product to Medtronic; if replacement product is needed, your Medtronic representative can assist you with identifying suitable replacement product, and to complete and return the Customer Confirmation Certificate confirming receipt of the notification via fax to Medtronic at 651-367-0612 to the attention of Customer Focused Quality (100% customer confirmations will be required). Should have any questions, please contact your Medtronic Representative or call 763-526-2513. Notifications to customers concerning the additional affected product began December 19, 2013 (in the US via UPS overnight service, 1-day delivery). Customers are requested to return a Confirmation Certificate confirming their receipt of the notification. 100% customer confirmations will be required from the additional consignees.

Details

Recall number
Z-0265-2014
Initiated
October 21, 2013
Posted by FDA
November 15, 2013
Terminated
December 15, 2014
Product code
DQX – Wire, Guide, Catheter
Quantity
1674
Distribution
Worldwide distribution: US (Nationwide) including states of: AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, and WV; and countries of: Albania, Algeria, Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Croatia, Czech Republic, Denmark…
Recalling firm
Medtronic Vascular, Danvers, MA

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0261-2014 Class ICougar Nitinol Workhorse Guidewire: Models: CGRLS190HJ, CGRLS190HS, CGRLS190S…Potential for PTFE (polytetrafluroethylene) coating to delaminate and detach from Guidewire. 10/21/2013Terminated
Z-0262-2014 Class IZinger Stainless Steel Workhorse Guidewire; Models: ZNGRLS180HJ, ZNGRLS180HS…Potential for PTFE (polytetrafluroethylene) coating to delaminate and detach from Guidewire. 10/21/2013Terminated
Z-0263-2014 Class ITHUNDER Extra Support Guidewire; Models: THNDR190J, THNDR190S, THNDR300J, THNDR300S…Potential for PTFE (polytetrafluroethylene) coating to delaminate and detach from Guidewire. 10/21/2013Terminated
Z-0264-2014 Class IProVia CROSSING GUIDEWIRE; Models: 15PROV300HS, 3PROV180HS, 3PROV180SS, 3PROV300HS…Potential for PTFE (polytetrafluroethylene) coating to delaminate and detach from Guidewire. Medtronic guide… 10/21/2013Terminated
Z-0266-2014 Class ICOUGAR Steerable Guidewire; Models: LVCLS190J, LVCLS190S, LVCXT190J, LVCXT190S Medtronic…Potential for PTFE (polytetrafluroethylene) coating to delaminate and detach from Guidewire. 10/21/2013Terminated
Z-0267-2014 Class ITHUNDER Steerable Guidewire; Model: LVTNDR190S Medtronic guide wires are steerable guide…Potential for PTFE (polytetrafluroethylene) coating to delaminate and detach from Guidewire. 10/21/2013Terminated
Z-0268-2014 Class IZINGER Steerable Guidewire; Models: LVZRLS180J, LVZRLS180S, LVZRMS180J, LVZRMS180S…Potential for PTFE (polytetrafluroethylene) coating to delaminate and detach from Guidewire. 10/21/2013Terminated