Recall Z-0268-2014

Medtronic Vascular · initiated October 21, 2013

Class I Terminated

Class I – dangerous or defective products that could cause serious health problems or death

Product

ZINGER Steerable Guidewire; Models: LVZRLS180J, LVZRLS180S, LVZRMS180J, LVZRMS180S, LVZRXT180J, LVZRXT180S Medtronic guide wires are steerable guide wires that are used for the introduction and placement of diagnostics or interventional devices in the coronary and peripheral vasculature, and may be used to reach and cross a target lesion. Medtronic guide wires are not intended for cerebral vasculature. Medtronic steerable exchange guide wires are used to facilitate the substitution of one diagnostic or interventional device for another.

Codes / lots: G13A00118 G13A00715 G13A00716 G13A00843 G13A00937 G13A01601 G13A01620 G13A01621 G13A01638 G13A01639 G13A01650 GV854 GW761

Reason for recall

Potential for PTFE (polytetrafluroethylene) coating to delaminate and detach from Guidewire.

FDA-determined cause: Nonconforming Material/Component

Action taken

The firm, Medtronic, sent a "URGENT MEDICAL DEVICE RECALL" letter dated October 21, 2013 (in the US via UPS overnight service, 1-day delivery) to its customers. The letter described the product, problem and actions to be taken. The customers were instructed to immediately remove and quarantine all potentially affected guidewires that remain in inventory; return the affected product to Medtronic; if replacement product is needed, your Medtronic representative can assist you with identifying suitable replacement product, and to complete and return the Customer Confirmation Certificate confirming receipt of the notification via fax to Medtronic at 651-367-0612 to the attention of Customer Focused Quality (100% customer confirmations will be required). Should have any questions, please contact your Medtronic Representative or call 763-526-2513. Notifications to customers concerning the additional affected product began December 19, 2013 (in the US via UPS overnight service, 1-day delivery). Customers are requested to return a Confirmation Certificate confirming their receipt of the notification. 100% customer confirmations will be required from the additional consignees.

Details

Recall number
Z-0268-2014
Initiated
October 21, 2013
Posted by FDA
November 15, 2013
Terminated
December 15, 2014
Product code
DQT – Occluder, Catheter Tip
Quantity
5
Distribution
Worldwide distribution: US (Nationwide) including states of: AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, and WV; and countries of: Albania, Algeria, Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Croatia, Czech Republic, Denmark…
Recalling firm
Medtronic Vascular, Danvers, MA

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0261-2014 Class ICougar Nitinol Workhorse Guidewire: Models: CGRLS190HJ, CGRLS190HS, CGRLS190S…Potential for PTFE (polytetrafluroethylene) coating to delaminate and detach from Guidewire. 10/21/2013Terminated
Z-0262-2014 Class IZinger Stainless Steel Workhorse Guidewire; Models: ZNGRLS180HJ, ZNGRLS180HS…Potential for PTFE (polytetrafluroethylene) coating to delaminate and detach from Guidewire. 10/21/2013Terminated
Z-0263-2014 Class ITHUNDER Extra Support Guidewire; Models: THNDR190J, THNDR190S, THNDR300J, THNDR300S…Potential for PTFE (polytetrafluroethylene) coating to delaminate and detach from Guidewire. 10/21/2013Terminated
Z-0264-2014 Class IProVia CROSSING GUIDEWIRE; Models: 15PROV300HS, 3PROV180HS, 3PROV180SS, 3PROV300HS…Potential for PTFE (polytetrafluroethylene) coating to delaminate and detach from Guidewire. Medtronic guide… 10/21/2013Terminated
Z-0265-2014 Class IATTAIN HYBRID Guide Wire; Models: GWR419478, GWR419488, GWR419688 The Attain Hybrid…Potential for PTFE (polytetrafluroethylene) coating to delaminate and detach from Guidewire. 10/21/2013Terminated
Z-0266-2014 Class ICOUGAR Steerable Guidewire; Models: LVCLS190J, LVCLS190S, LVCXT190J, LVCXT190S Medtronic…Potential for PTFE (polytetrafluroethylene) coating to delaminate and detach from Guidewire. 10/21/2013Terminated
Z-0267-2014 Class ITHUNDER Steerable Guidewire; Model: LVTNDR190S Medtronic guide wires are steerable guide…Potential for PTFE (polytetrafluroethylene) coating to delaminate and detach from Guidewire. 10/21/2013Terminated