Recall Z-0418-2023

Integrity Implants Inc. · initiated January 27, 2022

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

LineSider Spinal System 5.5/6.0mm, Non-mod Standard 7.1mm - Dual Lead, Non-cannulated. Thoracolumbosacral pedicle screw system; Spinal interlaminal fixation orthosis. Product sold as non-sterile and shipped in protective sterilization trays. Screws can also be shipped individually, as tray replenishments. Labeled as the following sizes: a. 5.5mm x 35mm; b. 5.5mm x 40mm; c. 5.5mm x 45mm; d. 5.5mm x 50mm; e. 5.5mm x 55mm; f. 6.5mm x 35mm; g. 6.5mm x 40mm; h. 6.5mm x 45mm; i. 6.5mm x 50mm; j. 6.5mm x 55mm; k. 6.5mm x 60mm; l. 7.5mm x 35mm; m. 7.5mm x 40mm; n. 7.5mm x 50mm; o. 7.5mm x 55mm; p. 7.5mm x 60mm; q. 8.5mm x 35mm; r. 8.5mm x 40mm; s. 8.5mm x 45mm; t. 8.5mm x 50mm; u.8.5mm x 55mm; v. 8.5mm x 60mm; Used in spinal fixation surgery.

Codes / lots: a. 5.5mm x 35mm, Part Number: LS-N6SA5535, UDI: (01)00818613027716(10)02210117, lot Number: 2210117; Part Number: LS-N6SA7545, UDI: (01)00818613027884(10)02210130, Lot Number: 2210130. b. 5.5mm x 40mm, Part Number: LS-N6SA5540, UDI: (01)00818613027723(10)02210118, Lot Number: 2210118. c. 5.5mm x 45mm, Part Number: LS-N6SA5545, UDI: (01)00818613027730(10)02210119, Lot Number: 2210119. d. 5.5mm x 50mm, Part Number: LS-N6SA5550, UDI: (01)00818613027747(10)02210120, Lot Number: 2210120. e. 5.5mm x 55mm, Part Number: LS-N6SA5555, UDI: (01)00818613027754(10)02210121, Lot Number: 2210121. f. 6.5mm x 35mm, Part Number: LS-N6SA6535, UDI: (01)00818613027792(10)02210122, Lot Number: 2210122. g. 6.5mm x 40mm, Part Number: LS-N6SA6540, UDI: (01)00818613027808(10)02210123, Lot Number: 2210123. h. 6.5mm x 45mm, Part Number: LS-N6SA6545, UDI: (01)00818613027815(10)02210124, Lot Number: 2210124. i. 6.5mm x 50mm, Part Number: LS-N6SA6550, UDI: (01)00818613027822(10)02210125, Lot Number: 2210125. j. 6.5mm x 55mm, Part Number: LS-N6SA6555, UDI: (01)00818613027839(10)02210126, Lot Number: 2210126. k. 6.5mm x 60mm, Part Number: LS-N6SA6560, UDI: (01)00818613027846(10)02210127, Lot Number: 2210127. l.…

Reason for recall

Potential for weld separation of the Upper Tulip Head component from the Lower Tulip Head component, which would result in the separation of the Tulip Head from the Screw Shank.

FDA-determined cause: Device Design

Action taken

Integrity Implants notified consignees on 01/28/2022, via emailed letter. The letter instructed consignees to quarantine affected product, complete and return the Product Tracking & Verification Form, and return affected product. On 02/05/2022 the recall was expanded, and consignees were notified via emailed letter. The listed the additional affected product and contained the same instructions to consignees.

Details

Recall number
Z-0418-2023
Initiated
January 27, 2022
Posted by FDA
December 5, 2022
Terminated
October 23, 2023
Product code
NKB – Thoracolumbosacral Pedicle Screw System
Quantity
1,227 screws
Distribution
Nationwide distribution to AZ, CA, CT, DC, FL, IL, IN, LA, MA, MN, NJ, OK, PR, TN, TX, VA.
Recalling firm
Integrity Implants Inc., Palm Beach Gardens, FL

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0419-2023 Class IILineSider Spinal System 5.5/6.0mm, Non-mod Standard 7.1mm - Dual Lead, Cannulated.…Potential for weld separation of the Upper Tulip Head component from the Lower Tulip Head component, which… 01/27/2022Terminated
Z-0420-2023 Class IILineSider Spinal System 5.5/6.0mm, Non-mod MIS 7.1mm - Dual Lead, Cannulated.…Potential for weld separation of the Upper Tulip Head component from the Lower Tulip Head component, which… 01/27/2022Terminated