Recall Z-0419-2023
Integrity Implants Inc. · initiated January 27, 2022
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
LineSider Spinal System 5.5/6.0mm, Non-mod Standard 7.1mm - Dual Lead, Cannulated. Thoracolumbosacral pedicle screw system; Spinal interlaminal fixation orthosis. Product sold as non-sterile and shipped in protective sterilization trays. Screws can also be shipped individually, as tray replenishments. Labeled as the following sizes: a. 5.5mm x 35mm; b. 5.5mm x 40mm; c. 5.5mm x 45mm; d. 5.5mm x 50mm; e. 5.5mm x 55mm; f. 6.5mm x 35mm; g. 6.5mm x 40mm; h. 6.5mm x 45mm; i. 6.5mm x 50mm; j. 6.5mm x 55mm; k. 6.5mm x 60mm; l. 7.5mm x 35mm; m. 7.5mm x 40mm; n. 7.5mm x 45mm; o. 7.5mm x 50mm; p. 7.5mm x 55mm; q. 7.5mm x 60mm; r. 8.5mm x 35mm; s. 8.5mm x 40mm; t. 8.5mm x 45mm; u. 8.5mm x 50mm; v.8.5mm x 55mm; w. 8.5mm x 60mm; Used in spinal fixation surgery.
Codes / lots: a. 5.5mm x 35mm, Part Number: LS-N6SG5535, UDI: (01)00818613022568(10)02210139, lot Number: 2210139. b. 5.5mm x 40mm, Part Number: LS-N6SG5540, UDI: (01)00818613022575(10)02210140, Lot Number: 2210140. c. 5.5mm x 45mm, Part Number: LS-N6SG5545, UDI: ((01)00818613022582(10)02210141, Lot Number: 2210141. d. 5.5mm x 50mm, Part Number: LS-N6SG5550, UDI: (01)00818613022599(10)02210142, Lot Number: 2210142. e. 5.5mm x 55mm, Part Number: LS-N6SG5555, UDI: (01)00818613022605(10)02210143, Lot Number: 2210143. f. 6.5mm x 35mm, Part Number: LS-N6SG6535, UDI: (01)00818613022643(10)02210144, Lot Number: 2210144. g. 6.5mm x 40mm, Part Number: LS-N6SG6540, UDI: (01)00818613022650(10)02210145, Lot Number: 2210145. h. 6.5mm x 45mm, Part Number: LS-N6SG6545, UDI: (01)00818613022667(10)02210146, Lot Number: 2210146, UDI: (01)00818613022667(10)02210146-1, Lot Number: 02210146-1. i. 6.5mm x 50mm, Part Number: LS-N6SG6550, UDI: (01)00818613022674(10)02210147, Lot Number: 2210147. j. 6.5mm x 55mm, Part Number: LS-N6SG6555, UDI: (01)00818613022681(10)02210148, Lot Number: 2210148. k. 6.5mm x 60mm, Part Number: LS-N6SG6560, UDI: (01)00818613022698(10)02210149, Lot Number: 2210149. l. 7.5mm x 35mm, Part…
Reason for recall
Potential for weld separation of the Upper Tulip Head component from the Lower Tulip Head component, which would result in the separation of the Tulip Head from the Screw Shank.
FDA-determined cause: Device Design
Action taken
Integrity Implants notified consignees on 01/28/2022, via emailed letter. The letter instructed consignees to quarantine affected product, complete and return the Product Tracking & Verification Form, and return affected product. On 02/05/2022 the recall was expanded, and consignees were notified via emailed letter. The listed the additional affected product and contained the same instructions to consignees.
Details
- Recall number
- Z-0419-2023
- Initiated
- January 27, 2022
- Posted by FDA
- December 5, 2022
- Terminated
- October 23, 2023
- Product code
- NKB – Thoracolumbosacral Pedicle Screw System
- Quantity
- 1,702 screws
- Distribution
- Nationwide distribution to AZ, CA, CT, DC, FL, IL, IN, LA, MA, MN, NJ, OK, PR, TN, TX, VA.
- Recalling firm
- Integrity Implants Inc., Palm Beach Gardens, FL
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-0418-2023 Class II | LineSider Spinal System 5.5/6.0mm, Non-mod Standard 7.1mm - Dual Lead, Non-cannulated.…Potential for weld separation of the Upper Tulip Head component from the Lower Tulip Head component, which… | 01/27/2022Terminated |
| Z-0420-2023 Class II | LineSider Spinal System 5.5/6.0mm, Non-mod MIS 7.1mm - Dual Lead, Cannulated.…Potential for weld separation of the Upper Tulip Head component from the Lower Tulip Head component, which… | 01/27/2022Terminated |