Recall Z-0476-2023

Leica Microsystems, Inc. · initiated November 21, 2022

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Leica MICROSYSTEMS Provido, Part Numbers 10448950 (Premium), 10448976 (Standard). Surgical microscope system

Codes / lots: ***EXPANSION 09/08/2023 UPDATE a. Premium Part Number 10448950; Serial Numbers: 160222001, 280622001, 50422001, 50422002, 91221001, 180622001, 170622001, 61221001, 130922001, 120623001, 150623001, 81022001, 200623001, 51022001, 61022001. b. Standard Part Number 10448976; Serial Numbers: 120322001, 300122001, 40222002, 301221001, 40122001, 140522001, 151221002, 81221002, 100222002, 300122002, 110322001, 50322001, 90622001, 120422001, 90522001, 100522001, 250122001, 100222001, 171022001, 80922001, 70623001, 30922001, 91022001.

Reason for recall

There is a potential issue resulting in the Power Supply potentially becoming overheated due to internal components overstressing. Consequently, due to the overstressing of the internal components, one specific component can fail which will cause the device to unexpectedly shutdown.

FDA-determined cause: Process control

Action taken

The firm issued an URGENT FIELD SAFETY NOTICE to its consignees by email on 11/21/2022. The notice explained the issue and stated that a Leica Microsystems representative would be contacting each facility to arrange an update to the the affected devices and advised the following: Consignees can continue to use the system pending the field update; Ensure that all users of the affected microscope are aware of the Safety notes as described in section 3 of the respective user manual; Like any other instrument in the operating theater, the system may fail and Leica Microsystem (Schweiz) AG recommends that a backup system be kept available during the operation; Check before the operation : Connect the power cable. + Switch on the microscope. UPDATE 09/12/2023: Recall was expanded, and customers were sent the notification on about 09/11/2023. Each consignee is requested to acknowledge receipt of the Field Safety Notice letter outlining the potential manufacturing defect, perform an initial inspection, indicate the results of the initial inspection, and provide contact information for further Leica Microsystems Representative verification of conformity/defect.

Details

Recall number
Z-0476-2023
Initiated
November 21, 2022
Posted by FDA
December 9, 2022
Product code
EPT – Microscope, Surgical
Quantity
38 units
Distribution
US Nationwide
Recalling firm
Leica Microsystems, Inc., deerfield, IL

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0475-2023 Class IILeica MICROSYSTEMS Proveo 8, Part Number 10449058, surgical microscope systemThere is a potential issue resulting in the Power Supply potentially becoming overheated due to internal… 11/21/2022Open, Classified
Z-0477-2023 Class IILeica MICROSYSTEMS M530 OHX, Part Number 10448737, surgical microscope systemThere is a potential issue resulting in the Power Supply potentially becoming overheated due to internal… 11/21/2022Open, Classified