Leica Microsystems, Inc.

Medical device manufacturer in deerfield, IL.

2510(k) clearances
0PMA approvals
59recalls

Device types

510(k) clearances

510(k)DeviceDecision
K061871Leica Model FL800IZI · Traditional 09/20/2006SE
K990869Leica AT550HKX · Traditional 06/07/1999SE

Recalls

RecallProductInitiated
Z-0477-2023 Class IILeica MICROSYSTEMS M530 OHX, Part Number 10448737, surgical microscope systemThere is a potential issue resulting in the Power Supply potentially becoming overheated due to internal… 11/21/2022Open, Classified
Z-0476-2023 Class IILeica MICROSYSTEMS Provido, Part Numbers 10448950 (Premium), 10448976 (Standard).…There is a potential issue resulting in the Power Supply potentially becoming overheated due to internal… 11/21/2022Open, Classified
Z-0475-2023 Class IILeica MICROSYSTEMS Proveo 8, Part Number 10449058, surgical microscope systemThere is a potential issue resulting in the Power Supply potentially becoming overheated due to internal… 11/21/2022Open, Classified
Z-1851-2022 Class IILeica Surgical Operating Microscopes, PROvido Part Numbers: a. 10448976, b. 10448950During internal testing, Leica Microsystems has become aware of a component change on the photodiodes inside… 08/26/2022Open, Classified
Z-1850-2022 Class IILeica Surgical Operating Microscopes, M530 OHX, Part Number 10448737During internal testing, Leica Microsystems has become aware of a component change on the photodiodes inside… 08/26/2022Open, Classified
Z-1849-2022 Class IILeica Surgical Operating Microscopes, Arveo, Part Number 10448999During internal testing, Leica Microsystems has become aware of a component change on the photodiodes inside… 08/26/2022Open, Classified
Z-1848-2022 Class IILeica Surgical Operating Microscopes, ARveo 8, Part Number 10449063During internal testing, Leica Microsystems has become aware of a component change on the photodiodes inside… 08/26/2022Open, Classified
Z-0412-2022 Class IILeica Microsystems M220 F12 Microscope SystemsIn May 2021, Leica Microsystems received one (1) complaint pertaining to the M220 F12 Surgical Microscope… 11/03/2021Terminated
Z-2481-2021 Class IIBOND Ready-To-Use Primary Antibody CDX2 (EP25), REF PA0375Product may not perform as specified in IFU. 07/30/2021Terminated
Z-2399-2020 Class IILeica ARveo and M530 OH6 microscope systems.Possible unintended interruption of surgical procedures due to equipment shutdown, related to changes in… 05/20/2020Terminated
Z-0673-2020 Class IIBOND Polymer Refine Detection Kit, Catalog No. DS9800. This detection system is for in…BOND Polymer Refine Detection kits (DS9800) were held out of refrigerated storage longer than specified… 11/18/2019Open, Classified
Z-1729-2019 Class IILeica Cryostats, used to produce frozen sample sections for biological, medical and…The current labelling does not include any specific warning about usage of flammable freezing sprays. Such… 05/03/2019Terminated
Z-0772-2019 Class IILeica M220 F12 surgical microscope The Leica M220 F12 surgical microscope is an optical…The M220 optics may unintentionally drop into the surgical field, risking contact with the patient. 01/02/2019Terminated
Z-0281-2019 Class IIBOND Slide Labeler Zebra Printer Model TLP3842 and Model GX4301, for use with BOND-III…The power supply unit for the printer used with the device can potentially overheat or cause a fire hazard… 09/17/2018Terminated
Z-3175-2018 Class IILeica M525 F20 Surgical Microscope SystemThe firm received complaints that the instrument tipped over when the swing arm was fully extended. 09/04/2018Terminated
Z-0019-2019 Class IILeica HistoCore SPECTRA CV Coverslipper, used in preparation of histological and…Product was assembled with an isolating fiberglass hose within the oven in the wrong position. This results… 09/04/2018Terminated
Z-2659-2018 Class IILeica M530 OHX Microscope System Supplied with 100 - 120 Vac.Recall is due to a design weakness of the power insert module. 06/25/2018Terminated
Z-0064-2019 Class IILeica Biosystems HistoCore SPECTRA ST, Slide Stainer, Automated Product Usage: The…These devices have an incorrect specification against safety standard ISO61010-1. Should a single fault… 06/13/2018Terminated
Z-2555-2018 Class IILeica Microsystems products labeled as the following: Leica M320 Microscope Systems with…Possible failure of screws which hold the gas spring fixation bracket in place. Due to this failure, the gas… 06/04/2018Terminated
Z-0838-2019 Class IIPROVEO 8 Ophthalmic Microscope System optical instrument for improving the visibility of…The observed issues can lead to unexpected interruptions of the optics carrier focus drive movement… 09/25/2017Terminated
Z-2660-2017 Class IINussloch GmbH ASP6025 Tissue ProcessorIncorrect labeling for specified voltage for the Alarm Connectors on the rear side of the unit. 06/19/2017Open, Classified
Z-1877-2016 Class IINovocastr Liquid Mouse Monoclonal Antibody CDX2 (CDX2) is recommended for the assessment…Testing of retained units indicated that Lot numbers of CDX2 indicated are not stable up to the expiry date… 04/05/2016Terminated
Z-1876-2016 Class IINovocastr Liquid Mouse Monoclonal Antibody Renal Cell Carcinoma Marker (RCC) is…Testing of retained units indicated that Lot numbers of CDX2 and RCC indicated are not stable up to the… 04/05/2016Terminated
Z-1704-2016 Class IINovocastra Liquid Mouse Monoclonal Antibody Calretinin, NCL-L-CALRET-566. Intended for…Novocastra Liquid Mouse Monoclonal Antibody Calretinin when used in combination with specific Lots of… 03/22/2016Terminated
Z-1499-2016 Class IILeica Microsystems (Schweiz) Ag. The common name of the device is surgical microscope.…An electronic component of the Leica M525 OH4 and Leica M720 OH5 surgical microscopes with 400W illumination… 01/11/2016Terminated
Z-0823-2016 Class IIThis reagent is for in vitro diagnostic use. Pax-5 (1EW) monoclonal antibody is intended…The Bond" Polymer Refine Detection and Novolink" Polymer Detection System may not provide adequate staining… 12/30/2015Terminated
Z-0808-2016 Class IIBond Polymer Refine Red Detection DS9390 kits. The common name of the device is…The Firm informed the importer/distributor, Leica Microsystems, a manufacturing error occurred during the… 11/27/2015Terminated
Z-0341-2016 Class IINovocastra Lyophilized Mouse Monoclonal Antibody CD10 (CD10). Immunohistochemistry. in…These Lot numbers are not stable up to the expiry date on the product labeling. 09/03/2015Terminated
Z-2779-2015 Class IIIThe Leica ASP200S/ASP300S Tissue Processor. Pathology and Hematology Departments. The…The Leica ASP200S/ASP300S Tissue Processor have an incorrect internal wiring to the remote alarm plug, which… 08/05/2015Terminated
Z-2804-2015 Class IIThe Leica ASP6025 Tissue Processor. Pathology and Hematology Departments. The Leica…The ASP6025 Tissue Processor instrument is incorrectly getting into a state of bottle empty during retort… 07/27/2015Terminated
Z-1315-2015 Class IILiquid Mouse Monoclonal Antibody CD15 used is laboratories to perform…The staining intensity decreases over the shelf life. 12/16/2014Terminated
Z-0179-2015 Class IIThyroglobulin (1D4) is a mouse anti-human monoclonal antibody produced as a tissue…Bond Ready to Use Primary Antibody Thyroglobulin does not function as intended up to the expiry date on the… 04/18/2014Terminated
Z-1698-2014 Class IINovocastra Liquid Mouse Monoclonal Antibody Glutathione S-Transferase pi…Novocastra Liquid Mouse Monoclonal Antibody Glutathione S-Transferase pi (Product Code: NCL-L-GSTpi-438)… 02/26/2014Terminated
Z-1710-2014 Class IILiquid Mouse Monoclonal Antibody CD15 (NCL-L-CD15), a liquid tissue culture supernatant…Product does not function as intended up to the expiry date on the product labeling. There is a link between… 01/10/2014Terminated
Z-1726-2014 Class IILeica Ceiling Mount MS-1F & MC-1F for use with the M500 N operating microscope. Article…The ceiling mount supplier encountered an event in which the drive screw in the motor unit, which lifts and… 09/11/2013Terminated
Z-1725-2014 Class IILeica Ceiling Mount MSF-1 for use with M841 and M501 operating microscopes. Article…The ceiling mount supplier encountered an event in which the drive screw in the motor unit, which lifts and… 09/11/2013Terminated
Z-1446-2013 Class IILPC Fine Mesh Cassette for the Leica LPC Printer. Part No: 14060546848; 14060546846…An LPC Fine Mesh Cassette failed to stay closed during processing operations. 04/04/2013Terminated
Z-0521-2013 Class IINovocastra Ready-to-Use Mouse Monoclonal Antibody Terminal Deoxynuceotidyl Transferase…RTU-TdT-339 is not stable up to the expiry date on the product labeling, affecting the staining intensity. 11/14/2012Terminated
Z-1815-2013 Class IILeica ASP6025 Vacuum Tissue Processor The Leica ASP6025 is a modular tissue processor…The Instruction for Use 1v9 Ref F and all previous versions specify incorrect dimensions for the Ready To… 11/13/2012Terminated
Z-2053-2012 Class IILeica M525 F50 Surgical Microscope; Article No. 10338422; Leica Microsystems (Schweiz)…Leica Microsystems evaluated data showing that a malfunction resulting in an illumination reduction can be… 06/04/2012Terminated
Z-1473-2012 ASP 6025 Vacuum Tissue Processor; an automated modular tissue processor; Leica…There is a potential malfunction with the device in autorotation mode, which can lead to incorrect reagent… 03/30/2012Terminated
Z-1269-2012 Touch Panel/Control Unit Carrier of the Swingarm M820/M844 C19, article no. 10448194MO…The screws, which lock the bottom plate to the spindle, pull through allowing the carrier arm with touch… 02/27/2012Terminated
Z-1054-2012 Novocastra Liquid Concentrated Mouse Monoclonal Antibody CD7 (LP15) Reagent; an in vitro…The primary antibodies to CD2 (clone 11F11) and CD7 (clone LP15) have been adulterated. 01/10/2012Terminated
Z-1053-2012 Leica Bond Ready-to-Use Primary Antibody CD7 (LP15) Reagent; 7 mL vial; an in vitro…The primary antibodies to CD2 (clone 11F11) and CD7 (clone LP15) have been adulterated. 01/10/2012Terminated
Z-1052-2012 Leica Bond Ready-to-Use Primary Antibody CD2 (11F11); 7 mL vial; an in vitro diagnostic…The primary antibodies to CD2 (clone 11F11) and CD7 (clone LP15) have been adulterated. 01/10/2012Terminated
Z-1042-2012 Leica M822 Surgical Microscope Device Controller Software (MDC version 3.6, pack F…Leica Microsystems received complaints stating that the zoom function of the Leica M822 surgical microscope… 12/21/2011Terminated
Z-3182-2011 Leica Bond Ready-to-Use Primary Antibody Epithelial Membrane Antigen (GP1.4); an in…There was an issue with the controlled temperature storage unit where the products were stored prior to… 07/26/2011Terminated
Z-3181-2011 Leica Bond Ready-to-Use Primary Antibody CDX2 (AMT28); an in vitro diagnostic…There was an issue with the controlled temperature storage unit where the products were stored prior to… 07/26/2011Terminated
Z-3180-2011 Leica Bond Aspirating Probe Cleaning System; an in vitro diagnostic immunohistochemistry…There was an issue with the controlled temperature storage unit where the products were stored prior to… 07/26/2011Terminated
Z-2597-2011 Peloris II Rapid Tissue Processor; Leica Biosystems Melbourne Pty Ltd, Mount Waverly…The processor potentially has a faulty impellor which could lead to reagent carry-over from one processing… 04/11/2011Terminated
Z-1065-2011 Bond Polymer Refine Detection Kit; an in vitro diagnostic immunohistochemistry reagent…Users of the Bond Polymer Refine Detection System, DS9800, may experience instances of "weak staining" due… 11/16/2010Terminated
Z-2499-2010 Liquid Mouse Monoclonal Antibody Thyroid Transcription Factor-1; 1 mL liquid tissue…In combination with a sensitive polymer detection system, the TTF-1 antibodies may stain additional tumor… 08/02/2010Terminated
Z-2498-2010 Lyophilized Mouse Monoclonal Antibody Thyroid Transcription Factor-1; 0.1 mL and 1 mL…In combination with a sensitive polymer detection system, the TTF-1 antibodies may stain additional tumor… 08/02/2010Terminated
Z-2497-2010 Bond Ready-To-Use Primary Antibody Thyroid Transcription Factor-1 (SPT24); 7 mL vial; an…In combination with a sensitive polymer detection system, the TTF-1 antibodies may stain additional tumor… 08/02/2010Terminated
Z-1633-2010 Leica Wireless Footswitch, Type B used in conjunction with Leica Surgical Microscopes -…The pedals of the Footswitch Type B stick and the Zoom/Focus moves to the end position. 03/29/2010Terminated
Z-1632-2010 Leica Footswitch, Type B used in conjunction with Leica Surgical Microscopes - Models…The pedals of the Footswitch Type B stick and the Zoom/Focus moves to the end position. 03/29/2010Terminated
Z-1551-2010 Peloris Rapid Tissue Processor; Leica Biosystems Melbourne Pty Ltd, Mount Waverly, 3149…Confusion during manually changing reagents, general users running unvalidated protocols, and wax valve… 02/09/2010Terminated
Z-1221-2010 Shutter HX SP5, a component of the Leica Point Scanning Confocal Microscope Models TCS…A manufacturing defect was identified that could potentially result in the laser shutter not closing as… 11/12/2009Terminated
Z-1040-2009 QCA and iQCA software, a video microscopy software system for quantitative estrogen…A change in the software to open up the software architecture to include other analytes other than Estrogen… 11/06/2008Terminated