Recall Z-0632-2012

ConMed Corporation · initiated April 15, 2011

Terminated

Product

STAT 2 Primary Administration Set Gravity Flow Controller, 20 Drops/cc, 84 in. length, Catalog/REF No. S2V-20, STERILE, ASSEMBLED IN MEXICO, CONMED CORPORATION, 525 FRENCH ROAD, UTICA, NY 13502 --- This is a disposable device with a sterile, nonpyrogenic fluid path. The STAT 2® Secondary and Primary Administration Sets are for use in attachment to an I.V. bag and patient catheter inserted into a vein to control the infusion of intravenous fluids into the human body.

Codes / lots: Beginning Lot Code 1006094, Ending Lot Code 1103214

Reason for recall

ConMed received complaints of some units exhibiting inaccurate or inconsistent flow rates. It was determined that certain units were assembled with an incorrect component which could result in the devices exhibiting a different flow rate and might reduce the ability of the user to regulate the device at low flow rates.

FDA-determined cause: Mixed-up of materials/components

Action taken

ConMed Corporation sent Urgent Medical Device Recall Letters and Response Forms dated April 14, 2011, via UPS Priority Mail pouches to all domestic consignees. The letter identified the product, the problem, and the action to be taken by the consignee. The international consignees were sent Urgent Medical Device Recall Letters and Response Forms dated April 22, 2011, via FedEx International Priority. Users were instructed to immediately stop use of the device and return it to ConMed Corporation. --- RECALL EXPANSION: Consignees were also instructed to contact all of those organization within their facility and any other facilities that they may have supplied or given these affected products to. ConMed Corp. sent out Amended Urgent Medical Device Recall Letters and Response Forms dated May 10, 2011, to the 4 consignees via UPS PriorIty Mail for the expansion of the recall for Catalog No. S2-12N, Lot 1006234. For any questions call 315-624-3237 or email ivcontroller@conmed.com.

Details

Recall number
Z-0632-2012
Initiated
April 15, 2011
Posted by FDA
January 11, 2012
Terminated
January 15, 2013
Product code
FPA – Set, Administration, Intravascular
Distribution
Worldwide Distribution - USA (nationwide) and the countries of Canada, Panama and India
Recalling firm
ConMed Corporation, Utica, NY

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0629-2012 STAT 2 I.V. Controller, Catalog/REF No. S2, ASSEMBLED IN MEXICO, CONMED CORPORATION, 525…ConMed received complaints of some units exhibiting inaccurate or inconsistent flow rates. It was determined… 04/15/2011Terminated
Z-0630-2012 STAT 2 Extension Set Gravity Flow Controller, 12 in. length, Catalog/REF No. S2-12…ConMed received complaints of some units exhibiting inaccurate or inconsistent flow rates. It was determined… 04/15/2011Terminated
Z-0631-2012 STAT 2 Extension Set Gravity Flow Controller (Needle-Free), 12 in. length, Catalog/REF…ConMed received complaints of some units exhibiting inaccurate or inconsistent flow rates. It was determined… 04/15/2011Terminated
Z-0633-2012 STAT 2 Primary Administration Set Gravity Flow Controller (Needle-Free), 20 Drops/cc, 84…ConMed received complaints of some units exhibiting inaccurate or inconsistent flow rates. It was determined… 04/15/2011Terminated
Z-0634-2012 STAT 2 Secondary Set Gravity Flow Controller, 20 Drops/cc, 13 in. length, Catalog/REF…ConMed received complaints of some units exhibiting inaccurate or inconsistent flow rates. It was determined… 04/15/2011Terminated
Z-0635-2012 STAT 2 Primary Administration Set Gravity Flow Controller, 60 Drops/cc, 84 in. length…ConMed received complaints of some units exhibiting inaccurate or inconsistent flow rates. It was determined… 04/15/2011Terminated
Z-0636-2012 STAT 2 Primary Administration Set Gravity Flow Controller (Needle-Free), 60 Drops/cc, 84…ConMed received complaints of some units exhibiting inaccurate or inconsistent flow rates. It was determined… 04/15/2011Terminated