Recall Z-0636-2012
ConMed Corporation · initiated April 15, 2011
Product
STAT 2 Primary Administration Set Gravity Flow Controller (Needle-Free), 60 Drops/cc, 84 in. length, Catalog/REF No. S2V-60 N, STERILE, ASSEMBLED IN MEXICO, CONMED CORPORATION, 525 FRENCH ROAD, UTICA, NY 13502 --- This is a disposable device with a sterile, nonpyrogenic fluid path. The STAT 2® Secondary and Primary Administration Sets are for use in attachment to an I.V. bag and patient catheter inserted into a vein to control the infusion of intravenous fluids into the human body.
Codes / lots: Beginning Lot Code 1007054, Ending Lot Code 1011104
Reason for recall
ConMed received complaints of some units exhibiting inaccurate or inconsistent flow rates. It was determined that certain units were assembled with an incorrect component which could result in the devices exhibiting a different flow rate and might reduce the ability of the user to regulate the device at low flow rates.
FDA-determined cause: Mixed-up of materials/components
Action taken
ConMed Corporation sent Urgent Medical Device Recall Letters and Response Forms dated April 14, 2011, via UPS Priority Mail pouches to all domestic consignees. The letter identified the product, the problem, and the action to be taken by the consignee. The international consignees were sent Urgent Medical Device Recall Letters and Response Forms dated April 22, 2011, via FedEx International Priority. Users were instructed to immediately stop use of the device and return it to ConMed Corporation. --- RECALL EXPANSION: Consignees were also instructed to contact all of those organization within their facility and any other facilities that they may have supplied or given these affected products to. ConMed Corp. sent out Amended Urgent Medical Device Recall Letters and Response Forms dated May 10, 2011, to the 4 consignees via UPS PriorIty Mail for the expansion of the recall for Catalog No. S2-12N, Lot 1006234. For any questions call 315-624-3237 or email ivcontroller@conmed.com.
Details
- Recall number
- Z-0636-2012
- Initiated
- April 15, 2011
- Posted by FDA
- January 11, 2012
- Terminated
- January 15, 2013
- Product code
- FPA – Set, Administration, Intravascular
- Distribution
- Worldwide Distribution - USA (nationwide) and the countries of Canada, Panama and India
- Recalling firm
- ConMed Corporation, Utica, NY
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-0629-2012 | STAT 2 I.V. Controller, Catalog/REF No. S2, ASSEMBLED IN MEXICO, CONMED CORPORATION, 525…ConMed received complaints of some units exhibiting inaccurate or inconsistent flow rates. It was determined… | 04/15/2011Terminated |
| Z-0630-2012 | STAT 2 Extension Set Gravity Flow Controller, 12 in. length, Catalog/REF No. S2-12…ConMed received complaints of some units exhibiting inaccurate or inconsistent flow rates. It was determined… | 04/15/2011Terminated |
| Z-0631-2012 | STAT 2 Extension Set Gravity Flow Controller (Needle-Free), 12 in. length, Catalog/REF…ConMed received complaints of some units exhibiting inaccurate or inconsistent flow rates. It was determined… | 04/15/2011Terminated |
| Z-0632-2012 | STAT 2 Primary Administration Set Gravity Flow Controller, 20 Drops/cc, 84 in. length…ConMed received complaints of some units exhibiting inaccurate or inconsistent flow rates. It was determined… | 04/15/2011Terminated |
| Z-0633-2012 | STAT 2 Primary Administration Set Gravity Flow Controller (Needle-Free), 20 Drops/cc, 84…ConMed received complaints of some units exhibiting inaccurate or inconsistent flow rates. It was determined… | 04/15/2011Terminated |
| Z-0634-2012 | STAT 2 Secondary Set Gravity Flow Controller, 20 Drops/cc, 13 in. length, Catalog/REF…ConMed received complaints of some units exhibiting inaccurate or inconsistent flow rates. It was determined… | 04/15/2011Terminated |
| Z-0635-2012 | STAT 2 Primary Administration Set Gravity Flow Controller, 60 Drops/cc, 84 in. length…ConMed received complaints of some units exhibiting inaccurate or inconsistent flow rates. It was determined… | 04/15/2011Terminated |