Recall Z-0653-2016

Intuitive Surgical, Inc. · initiated November 23, 2015

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

30 deg, 8mm Endoscope, IS4000: Model numbers: 470027-34, 470027-40, 470027-41. General and Plastic Surgery: The da Vinci Xi endoscope produces high definition (HD), visible light (VIS). The HD video is processed by the system electronics in the Vision Cart and displayed on the Surgeon Console 3D viewer and Vision Cart touchscreen. Endoscopes are available with either a 0 degree or 30 degree tips.

Codes / lots: Model numbers: 470027-34, 470027-40, 470027-41. All lots.

Reason for recall

This correction to the da Vinci Xi I&A User Manual communicates proper use of the da Vinci Xi 8 mm Endoscopes in combination with the da Vinci 8 mm and 12 mm cannulae, cannula reducers, and monopolar energy instruments. It informs customers of scenarios that may result in unintended patient harm due to capacitive coupling from monopolar energy sources.

FDA-determined cause: Device Design

Action taken

Intuitive Surgical sent an Urgent Medical Device Correction letter to affected customers on December 4, 2015,. Letters explained the issue and provided action to follow to prevent the issue. Customers were instructed to do the following: Distribute a copy of this letter to all da Vinci Xi users at your facility. 3. Place a copy of this letter with your user manual. 4. Complete the attached Acknowledgement Form and return it to Intuitive Surgical as instructed. 5. Retain a copy of this letter and the Acknowledgement Form for your files. If you need further information or support concerning this Medical Device Notification , please contact your Clinical Sales Representative or contact Intuitive Surgical Customer Service at the numbers listed below: North and South America: (800) 876-1310, Option 3 (6 AM to 5 PM PST) or mail: customersupport-servicesupport@intusurg.com Europe, Middle East, Asia and Africa: +800 0821 2020 or +41 21 821 2020 (8 AM to 6 PM CET) or ics@intusurg.com South Korea: 02-3271-3200 (9 AM to 6 PM KSTJ) Japan: 0120-56-5635 or 003-5575-1362 (9 AM to 6 PM JST)

Details

Recall number
Z-0653-2016
Initiated
November 23, 2015
Posted by FDA
January 14, 2016
Terminated
April 5, 2016
Product code
NAY – System, Surgical, Computer Controlled Instrument
Quantity
2624 total all models
Distribution
Worldwide Distribution - US (nationwide) and Internationally to Australia, Austria, Belgium, Cyprus, Czech Republic, Denmark, France, Germany, Hong Kong, India, Israel, Italy, Japan, Monaco, Netherlands, Norway, Puerto Rico, Qatar, Romania, Saudi Arabia, Singapore, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Kingdom, and Venezuela. .
Recalling firm
Intuitive Surgical, Inc., Sunnyvale, CA

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0652-2016 Class II0 deg, 8mm Endoscope, IS4000: Model numbers: 470026-34, 470026-40, 470026-41. General…This correction to the da Vinci Xi I&A User Manual communicates proper use of the da Vinci Xi 8 mm… 11/23/2015Terminated