Intuitive Surgical, Inc.

Medical device manufacturer in Sunnyvale, CA.

197510(k) clearances
0PMA approvals
276recalls

Device types

510(k) clearances

510(k)DeviceDecision
K261823da Vinci Surgical System, Model IS5000NAY · Traditional 09/08/2026SE
K253702Intuitive Endoluminal Gastrointestinal System (IG1000)FDS · Traditional 08/07/2026SE
K261824da Vinci Surgical System, Model IS5000NAY · Traditional 07/31/2026SE
K261249Universal Seal (5-12 mm)GCJ · Traditional 07/14/2026SE
K260695da Vinci Firefly Imaging SystemNAY · Traditional 05/22/2026SE
K253986da Vinci Force Feedback InstrumentsNAY · Traditional 03/11/2026SE
K253978Universal Seal (5-12 mm)GCJ · Traditional 02/24/2026SE
K253556da Vinci SP Surgical System (SP1098); da Vinci E-200 GeneratorNAY · Traditional 01/16/2026SE
K251739da Vinci Surgical System (IS5000)NAY · Traditional 01/16/2026SE
K251763IRISegQIH · Traditional 12/16/2025SE
K252675da Vinci SP Surgical System (SP1098)NAY · Traditional 12/15/2025SE
K252069da Vinci SP Surgical System (SP1098)NAY · Traditional 12/02/2025SE
K251426da Vinci SP Surgical System (SP1098)NAY · Traditional 10/03/2025SE
K252045Ion Endoluminal System (IF1000)EOQ · Traditional 09/23/2025SE
K251817da Vinci SP Surgical System (SP1098)NAY · Traditional 09/10/2025SE
K251227da Vinci Surgical System (IS5000)NAY · Traditional 09/08/2025SE
K252528Ion Endoluminal System (IF1000)EOQ · Special 09/05/2025SE
K251324da Vinci SP Surgical System (SP1098)NAY · Traditional 08/29/2025SE
K252350da Vinci SP Firefly Imaging SystemNAY · Special 08/28/2025SE
K250442da Vinci Surgical System, IS5000NAY · Traditional 07/29/2025SE
K251213Da Vinci Surgical System (IS5000)NAY · Traditional 07/23/2025SE
K251202da Vinci Surgical System (IS5000)NAY · Traditional 07/16/2025SE
K2514908 mm Assist Cannula; 12 mm Assist CannulaGCJ · Traditional 07/02/2025SE
K250674Vessel Sealer Curved (480522)NAY · Traditional 06/13/2025SE
K240852da Vinci X Surgical System (IS4200); da Vinci Xi Surgical System (IS4000)NAY · Traditional 06/11/2025SE
K250786SP Endoscope, 0° (430600)NAY · Traditional 05/13/2025SE
K243714da Vinci SP Surgical System (SP1098)NAY · Traditional 04/24/2025SE
K242427da Vinci Xi Surgical System (IS4000); da Vinci X Surgical System (IS4200)NAY · Traditional 04/01/2025SE
K243596da Vinci SP SureForm 45 Staplers and Reloads (SP1098)GAG · Traditional 03/10/2025SE
K243582da Vinci SP Advanced Access Port Kit (432701)NAY · Traditional 02/28/2025SE
K243641da Vinci Surgical System (IS5000)NAY · Traditional 12/16/2024SE
K242461IRISegQIH · Traditional 12/10/2024SE
K242318da Vinci SP Surgical System (SP1098)NAY · Traditional 11/26/2024SESU
K241814da Vinci SP Instruments (SP1098)NAY · Traditional 11/18/2024SESU
K242070Ion Endoluminal System (IF 1000)EOQ · Traditional 09/24/2024SE
K2416388mm SureForm 30 Curved-Tip Stapler; 8mm SureForm 30 Stapler; 8mm SureForm 30 ReloadsNAY · Traditional 09/10/2024SE
K241635da Vinci E-200 Electrosurgical GeneratorGEI · Traditional 08/05/2024SE
K241360Universal Seal (5-12 mm)GCJ · Traditional 08/05/2024SE
K241621Da Vinci Monopolar and Bipolar Adapters (378896)GEI · Traditional 08/02/2024SE
K240502da Vinci SP Surgical System (SP1098)NAY · Traditional 07/08/2024SESU
K241284da Vinci Surgical System (IS5000)NAY · Traditional 07/03/2024SE
K232773da Vinci SP Surgical System (SP1098)NAY · Traditional 06/04/2024SE
K240723da Vinci Xi Surgical System (IS4000); da Vinci X Surgical System (IS4200)NAY · Traditional 05/24/2024SE
K240135Ion™ Endoluminal System (Ion™ Fully Articulating Catheter) (IF1000)EOQ · Traditional 04/03/2024SE
K232610da Vinci Surgical System Model IS5000, da Vinci Insufflator and Tube Set with Smoke…NAY · Traditional 03/14/2024SE
K232984Ion Endoluminal System (IF1000)EOQ · Traditional 12/18/2023SE
K231798da Vinci SP Surgical SystemNAY · Traditional 11/08/2023SESU
K231358Universal Seal (5-12 mm)GCJ · Traditional 07/05/2023SE
K231212da Vinci E-200 Electrosurgical GeneratorGEI · Traditional 06/21/2023SE
K231224da Vinci Xi Surgical System (IS4000), da Vinci X Surgical System (IS4200)NAY · Special 05/12/2023SE
K230033Da Vinci SP Surgical System (SP1098)NAY · Traditional 04/05/2023SESU
K222839EndoWrist Stapler 30 System, EndoWrist Stapler 45 System, SureForm 45 System, SureForm…NAY · Special 03/16/2023SE
K223039da Vinci E-200 Electrosurgical GeneratorGEI · Traditional 11/28/2022SE
K223080da Vinci X Surgical System (IS4200), da Vinci Xi Surgical System (IS4000)NAY · Traditional 11/22/2022SE
K222827da Vinci Firefly Imaging SystemNAY · Special 10/20/2022SE
K214095da Vinci X/Xi (IS4200/IS4000) 8mm Reusable InstrumentsNAY · Traditional 08/15/2022SE
K220299Ion Endoluminal System (IF1000)EOQ · Traditional 06/16/2022SE
K213710da Vinci Fluorescence Imaging Vision System, da Vinci Firefly Imaging SystemNAY · Traditional 02/17/2022SE
K2200238mm Monopolar Curved ScissorsNAY · Special 01/31/2022SE
K2119978 mm SureForm 30 Curved-Tip Stapler, 8 mm SureForm 30 Stapler, SureForm 30 ReloadsNAY · Traditional 12/10/2021SE
K212101da Vinci SP Firefly Imaging SystemNAY · Traditional 11/23/2021SE
K212747Da Vinci SP Surgical SystemNAY · Special 09/29/2021SESU
K212048Ion Endoluminal System (Ion Fully Articulating Catheter), Ion Endoluminal System (Ion…EOQ · Traditional 08/26/2021SE
K211784da Vinci Xi Surgical System (IS4000), da Vinci X Surgical System (IS4200)NAY · Traditional 08/06/2021SE
K211316da Vinci SP Surgical System (SP1098)NAY · Traditional 07/23/2021SESU
K211595da Vinci SP Surgical System (SP1098)NAY · Special 06/23/2021SESU
K210918da Vinci Fluorescence Imaging Vision System, da Vinci Firefly Imaging SystemNAY · Special 04/26/2021SE
K203632da Vinci S/Si (IS2000/IS3000) 5mm and 8mm Reusable Instruments, da Vinci Xi/X…NAY · Traditional 02/10/2021SE
K202968Da Vinci SP Surgical SystemNAY · Traditional 12/22/2020SESU
K202834da Vinci Xi Surgical System (IS4000), da Vinci X Surgical System (IS4200)NAY · Traditional 12/10/2020SE
K202370Ion Endoluminal SystemEOQ · Traditional 11/16/2020SE
K202571da Vinci SP Surgical System, Model SP1098, EndoWrist SP Instruments, and AccessoriesNAY · Traditional 11/12/2020SESU
K192717da Vinci SP Surgical SystemNAY · Traditional 09/28/2020SESU
K201146Ion Endoluminal SystemEOQ · Traditional 06/26/2020SE
K192803da Vinci Xi Surgical System, da Vinci X Surgical SystemNAY · Traditional 04/29/2020SE
K183086da Vinci X and Xi Surgical SystemNAY · Traditional 03/31/2020SE
K191529da Vinci Xi Surgical System, da Vinci X Surgical SystemNAY · Traditional 02/06/2020SE
K192367Ion Endoluminal SystemEOQ · Traditional 11/26/2019SE
K191280E-100 Electrosurgical Generator, SynchroSealNAY · Traditional 11/14/2019SE
K191736da Vinci X/Xi 8mm Endoscope Plus, 0, da Vinci X/Xi 8mm Endoscope Plus, 30NAY · Special 07/26/2019SE
K190999SureForm 45 Curved Tip, SureForm 45 Gray ReloadNAY · Traditional 07/12/2019SE
K191043da Vinci Handheld CameraGCJ · Traditional 06/06/2019SE
K182371da Vinci SP Surgical System, EndoWrist SP Instruments, and AccessoriesNAY · Traditional 03/14/2019SESU
K182643IRIS 1.0 SystemLLZ · Traditional 02/22/2019SE
K182188Ion Endoluminal System ; Flexision Biopsy NeedleEOQ · Traditional 02/14/2019SE
K183139Endoscope Sterilization TrayKCT · Abbreviated 02/01/2019SE
K183224SureForm 45, SureForm 45 ReloadsNAY · Traditional 01/18/2019SE
K183107Intuitive Surgical Vessel Sealer ExtendNAY · Traditional 12/11/2018SE
K182140da Vinci Xi Surgical System, da Vinci X Surgical SystemNAY · Traditional 10/24/2018SE
K180964Endoscope Sterilization TrayKCT · Traditional 09/28/2018SE
K180351EndoWrist Mercury Bipolar GrasperNAY · Traditional 08/07/2018SE
K181395Universal Cannula Seal (5-12 mm)GCJ · Traditional 07/27/2018SE
K173585da Vinci Xi Surgical System, da Vinci X Surgical SystemNAY · Traditional 07/19/2018SE
K173721SureForm 60 and SureForm 60 ReloadsNAY · Traditional 07/05/2018SE
K173906da Vinci SP Surgical System, EndoWrist SP Instruments, and AccessoriesNAY · Traditional 05/31/2018SESU
K173415EndoWrist 5mm Thoracic GrasperNAY · Traditional 05/31/2018SE
K173337Intuitive Surgical EndoWrist Vessel Sealer ExtendNAY · Traditional 04/26/2018SE
K173842da Vinci Xi Surgical System, da Vinci X Surgical SystemNAY · Traditional 04/23/2018SE
K1800338mm Monopolar Curved ScissorsNAY · Traditional 04/06/2018SE
K172643da Vinci Xi Surgical System; da Vinci X Surgical SystemNAY · Traditional 01/31/2018SE
K170879IS4000 Stapler 45 Instrument and its Reusable Accessories, IS4000 EndoWrist Stapler 30…NAY · Traditional 09/21/2017SE
K170641da Vinci S/Si Endoscopes, da Vinci Xi EndoscopesNAY · Traditional 09/21/2017SE
K171632da Vinci Xi Surgical SystemNAY · Traditional 09/19/2017SE
K170640Endoscope Sterilization TrayKCT · Traditional 09/17/2017SE
K170875da Vinci Si Single-Site Instruments and Accessories, da Vinci Xi Single-Site Instruments…NAY · Traditional 09/12/2017SE
K170645da Vinci Xi EndoWrist Instruments and AccessoriesNAY · Traditional 09/11/2017SE
K170644da Vinci S/Si EndoWrist Instruments and Accessories, Harmonic ACE Curved Shears (5mm &…NAY · Traditional 09/11/2017SE
K171699da Vinci Xi Surgical System, da Vinci Si Surgical System, da Vinci X Surgical SystemNAY · Traditional 07/28/2017SE
K171426da Vinci Xi 8mm Endoscope, 0 degree, da Vinci Xi 8mm Endoscope, 30 degreeNAY · Special 06/13/2017SE
K170713da Vinci Xi Surgical SystemNAY · Traditional 06/13/2017SE
K171388EndoWrist Stapler 45 System and Stapler 45 ReloadsNAY · Special 05/31/2017SE
K171294da Vinci X Surgical SystemNAY · Special 05/26/2017SE
K170865EndoWrist Vessel Sealer, 8 mm Harmonic ACE Curved Shears, da Vinci Single-Site…NAY · Traditional 04/21/2017SE
K170508EndoWrist Stapler 45 Instrument, EndoWrist Stapler 45 Reloads, EndoWrist Stapler 30…NAY · Traditional 03/10/2017SE
K162973EndoWrist Suction IrrigatorNAY · Traditional 02/06/2017SE
K161178da Vinci Xi Surgical SystemNAY · Traditional 01/19/2017SE
K162411da Vinci Xi 12 – 8 mm ReducerNAY · Special 09/21/2016SE
K153276da Vinci Xi Surgical SystemNAY · Traditional 08/07/2016SE
K161271da Vinci Xi Surgical SystemNAY · Special 07/11/2016SE
K152892da Vinci Xi Surgical SystemNAY · Traditional 04/29/2016SE
K152578da Vinci Surgical System, Model IS4000NAY · Traditional 03/30/2016SE
K152448da Vinci Single-Site Instrument and AccessoriesNAY · Traditional 03/09/2016SE
K152421IS4000 Stapler 30 Instrument and Stapler 30 ReloadsNAY · Traditional 03/04/2016SE
K151794da Vinci Xi Surgical SystemNAY · Traditional 01/15/2016SE
K151450Endoscope Sterilization TrayKCT · Traditional 01/12/2016SE
K153126da Vinci Xi Hasson ConeGCJ · Traditional 12/09/2015SE
K152663Bladeless Optical ObturatorsGCJ · Special 10/16/2015SE
K150284IS4000 da Vinci EndoWrist InstrumentsNAY · Special 05/15/2015SE
K150837IS4000 Small Clip Applier, IS4000 Long Bipolar ForcepsNAY · Special 04/29/2015SE
K143132IS4000 8mm Harmonic ACE Curved ShearsNAY · Traditional 04/02/2015SE
K142937Endoscope Sterilization TrayKCT · Traditional 02/24/2015SE
K14268312 mm Endoscopes and AccessoriesNAY · Traditional 12/10/2014SE
K14321712 mm & Stapler Bladeless ObturatorsNAY · Special 12/03/2014SE
K141075Single-Site Wristed Needle DriverNAY · Traditional 09/26/2014SESU
K123329Da Vinvi Surgical SystemNAY · Traditional 09/17/2014SESU
K141077Da Vinci Firefly Imaging SystemNAY · Traditional 08/12/2014SE
K133942Endoscope Sterilization TrayKCT · Abbreviated 08/01/2014SE
K140553Endowrist Stapler 45 and Stapler 45 ReloadsNAY · Traditional 07/25/2014SE
K140189Endowrist Vessel SealerNAY · Traditional 06/05/2014SE
K133203Single-Site PortGCJ · Traditional 05/09/2014SESU
K131962Da Vinci SP Surgical System, Endowrist SP Instruments, and AccessoriesNAY · Traditional 04/17/2014SESU
K131861Da Vinci Surgical System, Endowrist Instruments and AccessoriesNAY · Traditional 03/28/2014SE
K1338458MM Trocar KitGCJ · Traditional 02/19/2014SE
K133167Monopolar Cautery Cord, Bipolar Cautery CordGEI · Traditional 01/07/2014SE
K124031Da Vinci Fluorescence Imaging Vision SystemOWN · Traditional 09/13/2013SE
K130266Endowrist One Vessel SealerNAY · Traditional 08/29/2013SE
K122532Da Vinci Single-Site Instruments and AccessoriesNAY · Traditional 07/30/2013SESU
K130726Da Vinci Single-Site Permanent Cautery HookNAY · Special 06/07/2013SESU
K123840Connect for Da Vinci Surgical System(S)NAY · Special 02/14/2013SE
K123463Intuitive Surgical Da Vinci Si Surgical System SmartpedalsNAY · Special 12/03/2012SE
K121921Intuitive Surgical Onsite for Da Vinci Surgical SystemsNAY · Traditional 10/25/2012SE
K113706Endowrist Stapler SystemNAY · Traditional 10/17/2012SE
K120215Single-Site Medium-Large Clip Applier, Single-Site Cadiere Grasper, Single-Site Fundus…NAY · Special 04/30/2012SESU
K110639Endowrist One Vessel SealerNAY · Traditional 12/28/2011SE
K112208Intutive Surgical Da Vinci Single Site Instruments and AccessoriesNAY · Traditional 12/08/2011SESU
K112263Monopolar Curved Scissors Tip Cover AccessoryNAY · Special 10/07/2011SE
K1125845MM/8MM Harmonic Ace(Tm) Curved Shears, Disposable Harmonic Ace(Tm) Insert, Disposable…NAY · Special 09/29/2011SE
K110451Endowrist One Suction/IrrigatorNAY · Traditional 08/26/2011SE
K101581Intuitive Surgical Da Vinci S Surgical System with Da Vinci Connect & Da Vinci Onsite…NAY · Traditional 04/08/2011SE
K1017435MM Flared Cannula Model 420262, 8MM Flared Cannula Model 420319NAY · Traditional 02/04/2011SE
K101077Da Vinci Fluorescence Imaging Vision System, Model FF 100GCJ · Traditional 02/04/2011SE
K0932175 MM Harmonic Ace Instrument (Used with Da Vinci IS1200 & IS2000/IS3000 System)NAY · Traditional 01/21/2010SE
K090993Intuitive Surgical Endoscopic Instrument Control Systems, Models IS1200, IS2000 and IS3000NAY · Traditional 12/16/2009SE
K082497Intuitive Surgical Endowrist One Hot Shears InstrumentNAY · Traditional 05/07/2009SE
K081137Intuitive Surgical Da Vinci Si Surgical System: Model IS3000NAY · Traditional 02/18/2009SE
K081207Intuitive Surgical Da Vinci S Surgical System, Model IS2000, with Da Vinci Connect and…NAY · Traditional 12/19/2008SE
K073112Da Vinci Sonicpro Cleaning System, Model 400253NVE · Traditional 08/01/2008SE
K080291Intuitive Surgical Da Vinci Endoscopic Instruments and Control System and Endowrist…NAY · Traditional 03/19/2008SE
K080155Modification To:Intuitive Surgical Stereo View Endoscopic SystemGCJ · Special 03/06/2008SE
K072211Intuitive Surgical Instrument and Accessories Sterilization TraysKCT · Traditional 02/28/2008SE
K070947Intuitive Surgical Da Vinci Surgical System and Endoscopic Instruments and Endowrist…NAY · Traditional 02/14/2008SE
K072627Intuitive Surgical Da Vinci and Da Vinci S Surgical System and Endoscopic Instruments…NAY · Traditional 02/07/2008SE
K071405Probe Holder SystemLKF · Traditional 10/09/2007SE
K063220Da Vinci S Surgical SYSTEM-V1.1, Model IS2000NAY · Traditional 12/01/2006SE
K061260Intuitive Surgical Endowrist PK Dissecting Forceps, Models 400214 & 420214NAY · Special 05/18/2006SE
K060391Intuitive Surgical Endowrist Stabilizer, Model 420182NAY · Special 04/10/2006SE
K050802Modification to Intuitive Surgical Da Vinci Surgical System and Endoscopic InstrumentsNAY · Traditional 06/29/2005SE
K050369Intuitive Surgical Da Vinci Surgical System, Model IS2000NAY · Traditional 04/29/2005SE
K050404Intuitive Surgical Da Vinci Surgical System and Endoscopic InstrumentsHET · Traditional 04/21/2005SE
K043288Modification to Intuitive Surgical Da Vinci Surgical System and Endoscopic InstrumentsNAY · Traditional 03/03/2005SE
K050005Intuitive Surgical Monopolar Curved Scissors, Model 400179; Tip Cover Accessory, Model…NAY · Special 01/25/2005SE
K043153Intuitive Surgical Da Vinci Surgical System and Endoscopic Instruments, Models IS1200 &…NAY · Special 12/15/2004SE
K042855Intuitive Surgical Harmonic Curved Shears InstrumentNAY · Special 11/12/2004SE
K040237Intuitive Surgical Da Vinci Endoscopic Instrument Control System and Endoscopic…NAY · Traditional 07/07/2004SE
K040948Intuitive Surgical Endopass Endoscopic Delivery Instrument, Model P/N 400170NAY · Special 05/05/2004SE
K022574Intuitive Surgical Endoscopic Instrument Control System & Endoscopic Instruments, Model…NAY · Traditional 11/12/2002SE
K021036Intuitive Surgical Da Vinci Surgical System, Model IS1000NAY · Traditional 06/26/2002SE
K013952Intuitive Surgical Stero View Endoscopic SystemGCJ · Traditional 02/28/2002SE
K013416Intuitive Surgical Endowrist Endoscopic Instrument FamilyGEI · Traditional 01/10/2002SE
K012833Intuitive Surgical Bipolar ForcepsNAY · Traditional 11/16/2001SE
K011281Intuitive Surgical Ultrasonic ShearsNAY · Traditional 07/24/2001SE
K011002Intuitive Surgical Da Vinci Surgical System, Model Isi 1000NAY · Traditional 05/30/2001SE
K002489Intuitive Surgical Da Vinci Endoscopic Control SystemNAY · Traditional 03/02/2001SE
K001666Modification to Intuitive Surgical Stereo View Endoscopic SystemGCJ · Traditional 08/24/2000SE
K990144Intuitive Surgical Endoscopic Instruments, Intuitive Surgical Endoscopic Instrument…NAY · Traditional 07/11/2000SE
K990188Intuitive Surgical Stereo View Endoscopic System, Model SVS-1GCJ · Traditional 04/01/1999SE
K965001Intuitive Surgical Monarch Laparoscopic ManipulatorGCJ · Traditional 07/31/1997SE

Recalls (latest 200)

RecallProductInitiated
Z-2310-2026 Class IIBrand Name: Da Vinci 5 Product Name: ASSY,DV5 TOWER,IS5000 Model/Catalog Number: 380746…Due to unintentionally activated near infrared handheld camera when attempting to turn off illumination… 04/24/2026Open, Classified
Z-2521-2026 Class IIDa Vinci ASSY, PSS, SP1098, Part Number: 380601Operative components of surgical systems potentially distributed with E-brake retainer pin, so in event of a… 04/22/2026Open, Classified
Z-2046-2026 Class IIBrand Name: Da Vinci X and Da Vinci Xi Product Name: da Vinci X, Xi Surgical System…Due to manipulator containing screws with in the arm sub-assembly that my be susceptible to breaking. 04/02/2026Open, Classified
Z-1855-2026 Class IIntuitive 8 mm SureForm 30 Gray Reloads Reference Numbers: 48230M-05 and 48230M-06Reloads for curved-tipped stapler may potentially produce an incomplete staple line. 03/11/2026Open, Classified
Z-1355-2026 Class IIda Vinci 5 Surgeon Console Viewer Display, ASSY, DV5 CONSOLE, IS5000, Part Number:…Robotic-assisted surgical system has Viewer Display manufactured with outdated firmware, may result in loss… 12/30/2025Open, Classified
Z-1096-2026 Class IIda Vinci SP, SP Access Port Kit, Large Incision Reference Numbers:430073…Access Port System tray may develop cracks potentially resulting in a sterility breach. 12/15/2025Open, Classified
Z-1701-2026 Class IIBrand Name: da Vinci S, Si Permanent Cautery Hook Instrument Product Name: da Vinci S…Due to increased in complaints regarding frayed or broken pitch cables on reusable surgical instruments 12/09/2025Open, Classified
Z-1700-2026 Class IIBrand Name: da Vinci S, Si Tenaculum Forceps Instrument Product Name: da Vinci S, Si…Due to increased in complaints regarding frayed or broken pitch cables on reusable surgical instruments 12/09/2025Open, Classified
Z-1699-2026 Class IIBrand Name: da Vinci S, Si Grasping Retractor Instrument Product Name: da Vinci S, Si…Due to increased complaints for broken/frayed grip cables for reusable instruments. 12/09/2025Open, Classified
Z-1698-2026 Class IIBrand Name: da Vinci S, Si Mega Needle Driver Instrument Product Name: da Vinci S, Si…Due to increased complaints for broken/frayed grip cables for reusable instruments. 12/09/2025Open, Classified
Z-1697-2026 Class IIBrand Name: da Vinci S, Si Double Fenestrated Grasper Instrument Product Name: da Vinci…Due to increased complaints for broken/frayed grip cables for reusable instruments. 12/09/2025Open, Classified
Z-1696-2026 Class IIBrand Name: da Vinci S, Si Monopolar Curved Scissors Instrument Product Name: da Vinci S…Due to increased complaints for broken/frayed grip cables for reusable instruments. 12/09/2025Open, Classified
Z-1134-2026 Class IIBrand Name: Ion Endoluminal System Product Name: Ion Vision Probe Bag Model/Catalog…Due to probe bags that may have a compromised or incomplete sterile pouch seal. 12/04/2025Open, Classified
Z-1428-2026 Class IIDa Vinci 5 ASSY, DV5 CONSOLE, IS5000 Model/Catalog Number: 380730 Used for minimally…An error event following the release of a software version for the surgical system that can result in loss… 11/17/2025Open, Classified
Z-1232-2026 Class IIBrand Name: Da Vinci Product Name: da Vinci X, Xi Surgical System, & da Vinci 5 Surgical…Due to a software implementation error that allowed instruments arms failing an important diagnostic test to… 11/07/2025Open, Classified
Z-2722-2026 Class IIda Vinci SP Instrument Arm Drape, SP, 1-PACK, Part Number: 430013-13, 430013-15Robotic-assisted surgical system arm drapes may detach from the instrument arm due to an issue involving the… 05/12/2025Open, Classified
Z-1325-2025 Class IIBrand Name: Da Vinci 5 Product Name: ASSY,DV5 CONSOLE,IS5000 Foot Tray Pedals…Due to an increase in complaints concerning foot tray pedal spring failing resulting in the pedal remaining… 02/26/2025Open, Classified
Z-0924-2025 Class II8MM, Small Graptor REF 470318 Is a grasping retractor Instrument used with the da Vinci…Due to an increase in complaints for pitch cable failures related to forceps and small Graptor instruments 12/19/2024Open, Classified
Z-0923-2025 Class II8MM, Tenaculum Forceps REF 470207 Instrument used with the da Vinci X/Xi Surgical…Due to an increase in complaints for pitch cable failures related to forceps and small Graptor instruments 12/19/2024Open, Classified
Z-0883-2025 Class II8MM,FORCE BIPOLAR,IS4000 REF 471405Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws. 12/19/2024Open, Classified
Z-0882-2025 Class II8MM,LONG BIPOLAR GRASPER,IS4000 REF 471400Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws. 12/19/2024Open, Classified
Z-0881-2025 Class II8MM,CURVED BIPOLAR DISSECTOR,IS4000 REF 471344Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws. 12/19/2024Open, Classified
Z-0880-2025 Class II8MM,MEGA SUTURECUT ND,IS4000 REF 471309Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws. 12/19/2024Open, Classified
Z-0879-2025 Class II8MM,LARGE SUTURECUT NEEDLE DRIVER,IS4000 REF 471296Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws. 12/19/2024Open, Classified
Z-0878-2025 Class II8MM,FENESTRATED BIPOLAR FORCEPS,IS4000 REF 471205Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws. 12/19/2024Open, Classified
Z-0877-2025 Class II8MM,COBRA GRASPER,IS4000 REF 471190Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws. 12/19/2024Open, Classified
Z-0876-2025 Class II8MM,MARYLAND BIPOLAR FORCEPS,IS4000 REF 471172Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws. 12/19/2024Open, Classified
Z-0875-2025 Class II8MM,MICRO BIPOLAR FORCEPS,IS4000 REF 471171Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws. 12/19/2024Open, Classified
Z-0874-2025 Class II8MM,PROGRASP FORCEPS,IS4000 REF 471093Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws. 12/19/2024Open, Classified
Z-0873-2025 Class II8MM,CADIERE FORCEPS,IS4000 REF 471049Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws. 12/19/2024Open, Classified
Z-0872-2025 Class II8MM,LONG TIP FORCEPS,IS4000 REF 471048Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws. 12/19/2024Open, Classified
Z-0871-2025 Class II8MM,FORCE BIPOLAR,IS4000 REF 470405Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws. 12/19/2024Open, Classified
Z-0870-2025 Class II8MM,HORIZON SMALL CLIP APPLIER,IS4000 REF 470401Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws. 12/19/2024Open, Classified
Z-0869-2025 Class II8MM,TIP-UP FENESTRATED GRASPER,IS4000 REF 470347Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws. 12/19/2024Open, Classified
Z-0868-2025 Class II8MM,CURVED BIPOLAR DISSECTOR,IS4000 REF 470344Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws. 12/19/2024Open, Classified
Z-0867-2025 Class II8MM,MEDIUM-LARGE CLIP APPLIER,IS4000 REF 470327Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws. 12/19/2024Open, Classified
Z-0866-2025 Class II8MM,SMALL GRASPING RETRACTOR,IS4000 REF 470318Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws. 12/19/2024Open, Classified
Z-0865-2025 Class II8MM,MEGA SUTURECUT ND,IS4000 REF 470309Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws. 12/19/2024Open, Classified
Z-0864-2025 Class II8MM,LARGE SUTURECUT NEEDLE DRIVER,IS4000 REF 470296Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws. 12/19/2024Open, Classified
Z-0863-2025 Class II8MM,DUAL BLADE RETRACTOR,IS4000 REF 470249Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws. 12/19/2024Open, Classified
Z-0862-2025 Class II8MM,ATRIAL RETRACTOR SHORT RIGHT,IS4000 REF 470246Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws. 12/19/2024Open, Classified
Z-0861-2025 Class II8MM,LARGE HEM-O-LOK CLIP APPLIER,IS4000 REF 470230Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws. 12/19/2024Open, Classified
Z-0860-2025 Class II8MM,TENACULUM FORCEPS,IS4000 REF 470207Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws. 12/19/2024Open, Classified
Z-0859-2025 Class II8MM,FENESTRATED BIPOLAR FORCEPS,IS4000 REF 470205Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws. 12/19/2024Open, Classified
Z-0858-2025 Class II8MM,MEGA NEEDLE DRIVER,IS4000 REF 470194Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws. 12/19/2024Open, Classified
Z-0857-2025 Class II8MM,RESANO FORCEPS,IS4000 REF 470181Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws. 12/19/2024Open, Classified
Z-0856-2025 Class II8MM,MARYLAND BIPOLAR FORCEPS,IS4000 REF 470172Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws. 12/19/2024Open, Classified
Z-0855-2025 Class II8MM,MICRO BIPOLAR FORCEPS,IS4000 REF 470171Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws. 12/19/2024Open, Classified
Z-0854-2025 Class II8MM,PROGRASP FORCEPS,IS4000 REF 470093Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws. 12/19/2024Open, Classified
Z-0853-2025 Class II8MM,CADIERE FORCEPS,IS4000 REF 470049Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws. 12/19/2024Open, Classified
Z-0852-2025 Class II8MM,DEBAKEY FORCEPS,IS4000 REF 470036Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws. 12/19/2024Open, Classified
Z-0851-2025 Class II8MM,BLACK DIAMOND MICRO FORCEPS,IS4000 REF 470033Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws. 12/19/2024Open, Classified
Z-0850-2025 Class II8MM, ROUND TIP SCISSORS, IS4000 REF 470007Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws. 12/19/2024Open, Classified
Z-0849-2025 Class II8MM, LARGE NEEDLE DRIVER, IS4000 REF 470006Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws. 12/19/2024Open, Classified
Z-0848-2025 Class II8MM, POTTS SCISSORS, IS4000 REF 470001Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws. 12/19/2024Open, Classified
Z-0847-2025 Class II8MM, MONOPOLAR CURVED SCISSORS, IS4000 REF 470179Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws. 12/19/2024Open, Classified
Z-0964-2026 Class IIda Vinci 5, Surgeon Console Column Motor Connector; ASSY,DV5 CONSOLE,SSC,IS5000; Part…Surgical system Error 23062 due to connector1) Partially damaged "Potential problem with ergonomics: Disable… 08/21/2024Open, Classified
Z-2720-2024 Class IIIon Endoluminal System, REF: 380748-65There is a potential that the robotic system's instrument cart arm may contain screws that may potentially… 07/12/2024Open, Classified
Z-1518-2024 Class IIPatient Cart, Part: 380601, a component of the da Vinci SP Surgical System, Model: SP1098Surgical system may have inadequately welded top and/or bottom belts in the link 2 of the entry guide… 03/13/2024Open, Classified
Z-0406-2024 Class IIDa Vinci Single-Site Wristed Needle Driver, REF: 478115, 5 mmThere is the potential for the needle drive to become dislodged and/or broken which could result in… 10/06/2023Open, Classified
Z-2059-2023 Class IIda Vinci Xi/X Tip-Up Fenestrated Grasper, REF: 470347, VER: 12Fenestrated Graspers have a manufacturing issue has the potential to cause breakage on both upper and/or… 06/12/2023Open, Classified
Z-1293-2023 Class IIda Vinci Xi Surgical System Model IS4000 da Vinci X Surgical System Model IS4200Due to Universal Surgical Manipulator (USM) instrument carriage being loose as a result of the linear rail… 02/10/2023Open, Classified
Z-1262-2023 Class IIPatient Side Cart, REF: 380620, with instrument arms, used as part of the da Vinci X…Preventative maintenance data was used to identify instrument Arms, part of surgical systems, that allow… 01/24/2023Open, Classified
Z-0340-2023 Class IIINTUITIVE Da Vinci Xi 8mm SureForm 30 Curved-Tip Stapler Da Vinci Xi|X REF 488530…Potential staple deployment failure and device fragment generation 10/07/2022Open, Classified
Z-0133-2023 Class IIda Vinci X (IS4200) and Xi (IS4000) systems, which consists of finished devices Surgeon…Inadvertent energy delivery from surgical system instrument if 1) Force bipolar and bipolar instruments… 09/20/2022Open, Classified
Z-1460-2022 Class IIDa Vinci X/Xi Large Clip Applier, Model 470230 Da Vinci X/Xi Medium-Large Clip Applier…Due to the potential for unexpected motion caused by the partial disengagement of the instrument from the… 07/01/2022Open, Classified
Z-1459-2022 Class IIDa Vinci S/Si Large Clip Applier, Model 420230 Da Vinci S/Si Medium-Large Clip Applier…Due to the potential for unexpected motion caused by the partial disengagement of the instrument from the… 07/01/2022Open, Classified
Z-1484-2022 Class IIThe E-100 Generator (PN 374848-09) is a bipolar electrosurgical unit (ESU) designed to…Due to preventative maintenance (PM) not being performed on E-100 generators which is a bipolar… 06/22/2022Open, Classified
Z-0999-2022 Class IIda Vinici X da Vinci Xi Sureform 45 REF 480445 da Vinci X da Vinci Xi Sureform 45…As a result of complaint data review, it identified tissue pushout events occurring during firing of the… 03/17/2022Open, Classified
Z-1210-2022 Class IIDAVINCI | Energy SynchroSeal da Vinci X' da Vinci Xi', Single Use Instrument 8mm, REF…Placing excessive tissue in the instrument jaws of the devices prior to sealing and transection can result… 03/03/2022Open, Classified
Z-1209-2022 Class IIVESSEL SEALER EXTEND, Single-Use Instrument EndoWrist Instruments 8mm, da Vinci X' da…Placing excessive tissue in the instrument jaws of the devices prior to sealing and transection can result… 03/03/2022Open, Classified
Z-0702-2022 Class IIda Vinci Xi and X 12-8 mm Conductive Cannula Reducer, REF: 470381-11Conductive Cannula Reducer metal tip may get dislodged from the plastic shaft. If unnoticed, additional… 12/23/2021Open, Classified
Z-0419-2022 Class IIda Vinci SP Surgical systems with the following Model Name/ Model Number / UDI:…Issue was identified during internal engineering evaluation. Use of the system with the affected software… 09/22/2021Open, Classified
Z-0275-2022 Class IIThe da Vinci Xi and X surgical systems (IS4000 and IS4200) each consist of three main…Incorrect expiration date on its package labeling (shipper box and inner carton label). 09/21/2021Terminated
Z-0106-2022 Class IIIon Flexison Biopsy Needle, 19G, Extension Range : 0-3 cm, Ref: 490104-09 and 49004-10…Potential for the sheath tip on the biopsy needle to separate from the sheath shaft. 08/06/2021Terminated
Z-0489-2021 Class IIBox Label: ION Fully Articulating Catheter, REF 490105, VER 08 - Product Usage: used to…The distal tip ring of the fully articulating catheter may become dislodged during the procedure. 10/23/2020Terminated
Z-2950-2020 Class IIION Reprocessing Accessories Kit, IF1000 (Part Number: 490024-03) UDI:00886874119129.…One type of the ION Catheter Reprocessing Covers (PN 490114-02) has the potential to become disengaged or… 06/10/2020Open, Classified
Z-2949-2020 Class IIIon Catheter Reprocessing Cover, IF1000 (Part Number:490114-02) UDI:00886874116845.One type of the ION Catheter Reprocessing Covers (PN 490114-02) has the potential to become disengaged or… 06/10/2020Open, Classified
Z-2329-2020 Class IISTAINLESS STEEL REPROCESSING TRAY, 2D CAMERA HEAD, IS4000, Model No. 400499-03 - Product…Small but detectable holes in the sterilization wrap used with the firm's 8 mm endoscope sterilization trays… 12/19/2019Terminated
Z-2328-2020 Class IISTAINLESS STEEL REPROCESSING TRAY, 8MM ENDOSCOPE, IS4000, Model No. 400498-02 - Product…Small but detectable holes in the sterilization wrap used with the firm's 8 mm endoscope sterilization trays… 12/19/2019Terminated
Z-2131-2020 Class IIda Vinci SP Camera; PN 430060-33; Model Number: 430060-33. For endoscopic surgery.Some cameras were found to have residual soil adhered to the tip of the camera after reprocessing which may… 08/19/2019Open, Classified
Z-0927-2020 Class IIda Vinci SP Surgical System, SP1098, 00886874114605 Product Usage: used with an…The firm become aware of a potentially defective capacitors on circuit boards installed on certain da Vinci… 07/26/2019Terminated
Z-0926-2020 Class IIda Vinci X Surgical System Surgical System, IS4200, 00886874114216 Product Usage: used…The firm become aware of a potentially defective capacitors on circuit boards installed on certain da Vinci… 07/26/2019Terminated
Z-0925-2020 Class IIda Vinci Xi Surgical System Surgical System, IS4000, 00886874114216 - Product Usage:…The firm become aware of a potentially defective capacitors on circuit boards installed on certain da Vinci… 07/26/2019Terminated
Z-1891-2020 Class IIdaVinci XI Sureform 60 Reload, 60 mm black 6-Row 4.6 mm, Ref: 48360T-08 , UDI:…The firm became aware of a manufacturing variation in which black and green reloads in select Stapler… 07/01/2019Open, Classified
Z-1890-2020 Class IIdaVinci XI Sureform 60 reload, 60 mm Green 6- Row, 4.6 mm, REF 48360G-08, UDI:…The firm became aware of a manufacturing variation in which black and green reloads in select Stapler… 07/01/2019Open, Classified
Z-0283-2021 Class IIda Vinci Surgical System, IS 4000; Endoscope Controller, Model Number: 372601-16 (UDI:…One da Vinci Xi Endoscope Controller in the field was improperly calibrated during servicing. The result of… 06/17/2019Terminated
Z-1897-2020 Class IIda Vinci X Surgical System, IS4200 - Product Usage: The da Vinci Xi surgical system…Failure of Universal Surgical Manipulators due to a high friction points, resulting in an increase of arm… 04/11/2019Terminated
Z-1896-2020 Class IIda Vinci Xi Surgical System, IS4000 - Product Usage: The da Vinci Xi surgical system…Failure of Universal Surgical Manipulators due to a high friction points, resulting in an increase of arm… 04/11/2019Terminated
Z-2260-2020 Class IIda Vinci SP surgical systemIntuitive has become aware that the da Vinci SP system may trigger a mechanical vibration of the instrument… 04/01/2019Terminated
Z-0841-2019 Class IIdaVinci Harmonic ACE Curved ShearsDuring an internal inspection, Intuitive found that the sterile packaging of the da Vinci S/Si Harmonic ACE… 01/16/2019Open, Classified
Z-1105-2020 Class IIDa Vinci Xi/X EndoWrist Monopolar Curved scissors (MCS), model number 470179-19, UDI:…This recall is being initiated because lots of EndoWrist Monopolar Curved Scissor instruments are not… 12/04/2018Terminated
Z-1961-2020 Class IIda Vinci X Surgical SystemDue to a manufacturing variation in the arm that can result in a persistent recoverable fault error (error… 11/21/2018Terminated
Z-1960-2020 Class IIda Vinci Xi Surgical SystemDue to a manufacturing variation in the arm that can result in a persistent recoverable fault error (error… 11/21/2018Terminated
Z-1906-2020 Class IIda Vinci SP Instrument Arm Drape. REF 430013-12 QTY 1 PouchThe instrument drape may tear while draping the Patient Side Cart presenting a breach in the sterile barrier. 10/25/2018Terminated
Z-0845-2020 Class IIThe da Vinci SP surgical system is made of three main components: the Surgeon Console…The firm became aware of a problem with the surgical system where the cannula mount button may be damaged… 10/23/2018Terminated
Z-3025-2018 Class IIda Vinci X/Xi EndoWrist 8mm Monopolar Curved Scissors Model No. 470179-14 Product Usage:…Monopolar Curved Scissor (MCS) instruments may be susceptible to developing cracks near the proximal edge of… 08/15/2018Open, Classified
Z-3024-2018 Class IIda Vinci S/Si EndoWrist 8mm Monopolar Curved Scissors Model No. 420179-16 Product Usage:…Monopolar Curved Scissor (MCS) instruments may be susceptible to developing cracks near the proximal edge of… 08/15/2018Open, Classified
Z-3254-2018 Class IIda Vinci Si/X/Xi Surgical System Redundant Medical Grade Power Supply (RMGPS) Units The…Certain Redundant Medical Grade Power Supply (RMGPS) units specific to Surgeon Console and Patient Side… 07/16/2018Terminated
Z-0249-2020 Class IIEndoWrist Suction Irrigator, Single Use Instrument. Part Number: 480299-04. An accessory…The firm has identified the potential for a silicone particle to be generated within the Suction Irrigator… 06/04/2018Open, Classified
Z-2835-2018 Class IIThe da Vinci Xi Surgical System (IS4000) Model Name: IS4000, Endoscope Controller. Model…An Endoscope Controller has been improperly calibrated during manufacturing. This improper calibration may… 05/17/2018Open, Classified
Z-2381-2018 Class IIda Vinci Xi EndoWrist Stapler 45 Reload Green, Model 48445G-03, 12 reloads/box.Intuitive Surgical, Inc. (ISI) has recently become aware of an issue with the da Vinci Xi Stapler 45 Blue… 04/23/2018Terminated
Z-2380-2018 Class IIda Vinci Xi EndoWrist Stapler 45 Reload Blue, Model 48645B-03, 12 reloads/box.Intuitive Surgical, Inc. (ISI) has recently become aware of an issue with the da Vinci Xi Stapler 45 Blue… 04/23/2018Terminated
Z-2341-2018 Class IIda Vinci X Surgical System, Model IS4200. Product Usage: The Intuitive Surgical…Specific da Vinci X Surgical Systems (IS4200) were shipped with defective Arm Controller Joint (ACJ) boards… 03/07/2018Terminated
Z-0404-2018 Class IIdaVinci X EndoWrist(R) Stapler 45, REF 470298 Product Usage: The Intuitive Surgical…Specific da Vinci Xi EndoWrist Stapler 45 instruments are potentially impacted by a variation in the… 09/19/2017Terminated
Z-2630-2017 Class IIda Vinci Xi Surgical System Product Usage: The da Vinci Xi Surgical System (IS4000) is…da Vinci Xi(R) Surgical System Patient Side Carts (PSC) were shipped to the field with potentially… 06/01/2017Terminated
Z-1819-2017 Class IIda Vinci Xi EndoWrist Suction Irrigator, 8 mm instrument; PN 480299-03; and SOFTWARE…Intuitive Surgical has become aware that in specific scenarios with system software P6 and the da Vinci Xi… 03/31/2017Terminated
Z-1729-2017 Class IISTAPLER,IS4000; da Vinci Xi EndoWrist Stapler Release Kit (SRK); model number 470298. da…Intuitive Surgical has become aware and further investigation has determined that misinterpretation of… 03/15/2017Terminated
Z-1728-2017 Class IISTAPLER,IS4000; da Vinci Xi EndoWrist Stapler Release Kit (SRK); model number 470298. da…Intuitive Surgical has become aware and further investigation has determined that misinterpretation of… 03/15/2017Terminated
Z-1727-2017 Class IISTAPLER,IS4000; da Vinci Xi EndoWrist Stapler Release Kit (SRK); model number 470298. da…Intuitive Surgical has become aware and further investigation has determined that misinterpretation of… 03/15/2017Terminated
Z-1726-2017 Class IISTAPLER,IS4000; da Vinci Xi EndoWrist Stapler Release Kit (SRK); model number 470298. da…Intuitive Surgical has become aware and further investigation has determined that misinterpretation of… 03/15/2017Terminated
Z-1725-2017 Class IISTAPLER,IS4000; da Vinci Xi EndoWrist Stapler Release Kit (SRK); model number 470298. da…Intuitive Surgical has become aware and further investigation has determined that misinterpretation of… 03/15/2017Terminated
Z-1724-2017 Class IISTAPLER,IS4000; da Vinci Xi EndoWrist Stapler Release Kit (SRK); model number 470298. da…Intuitive Surgical has become aware and further investigation has determined that misinterpretation of… 03/15/2017Terminated
Z-1723-2017 Class IISTAPLER,IS4000; da Vinci Xi EndoWrist Stapler Release Kit (SRK); model number 470298. da…Intuitive Surgical has become aware and further investigation has determined that misinterpretation of… 03/15/2017Terminated
Z-1722-2017 Class IISTAPLER,IS4000; da Vinci Xi EndoWrist Stapler Release Kit (SRK); model number 470298. da…Intuitive Surgical has become aware and further investigation has determined that misinterpretation of… 03/15/2017Terminated
Z-1721-2017 Class IISTAPLER,IS4000; da Vinci Xi EndoWrist Stapler Release Kit (SRK); model number 470298. da…Intuitive Surgical has become aware and further investigation has determined that misinterpretation of… 03/15/2017Terminated
Z-1720-2017 Class IISTAPLER,IS4000; da Vinci Xi EndoWrist Stapler Release Kit (SRK); model number 470298. da…Intuitive Surgical has become aware and further investigation has determined that misinterpretation of… 03/15/2017Terminated
Z-1719-2017 Class IISTAPLER,IS4000; da Vinci Xi EndoWrist Stapler Release Kit (SRK); model number 470298. da…Intuitive Surgical has become aware and further investigation has determined that misinterpretation of… 03/15/2017Terminated
Z-1718-2017 Class IISTAPLER,IS4000; da Vinci Xi EndoWrist Stapler Release Kit (SRK); model number 470298. da…Intuitive Surgical has become aware and further investigation has determined that misinterpretation of… 03/15/2017Terminated
Z-1717-2017 Class IISTAPLER,IS4000; da Vinci Xi EndoWrist Stapler Release Kit (SRK); model number 470298. da…Intuitive Surgical has become aware and further investigation has determined that misinterpretation of… 03/15/2017Terminated
Z-1716-2017 Class IISTAPLER,IS4000; da Vinci Xi EndoWrist Stapler Release Kit (SRK); model number 470298. da…Intuitive Surgical has become aware and further investigation has determined that misinterpretation of… 03/15/2017Terminated
Z-1715-2017 Class IISTAPLER,IS4000; da Vinci Xi EndoWrist Stapler Release Kit (SRK); model number 470298. da…Intuitive Surgical has become aware and further investigation has determined that misinterpretation of… 03/15/2017Terminated
Z-1714-2017 Class IISTAPLER,IS4000; da Vinci Xi EndoWrist Stapler Release Kit (SRK); model number 470298. da…Intuitive Surgical has become aware and further investigation has determined that misinterpretation of… 03/15/2017Terminated
Z-0315-2017 Class IIda Vinci Xi" Surgical System, model number IS4000, A70_P5x with P5 Software; General and…Intuitive Surgical has identified a software anomaly in the da Vinci Xi P5 software that can result in… 10/05/2016Terminated
Z-2677-2016 Class IICURVED-TIP STAPLER 30,IS4000; Model number 470530; General and Plastic Surgery: The…Field failures were reported related to the da Vinci Xi Surgical System EndoWrist Stapler 45 and 30… 07/14/2016Terminated
Z-2676-2016 Class IISTAPLER 30,IS4000; Model number 470430; General and Plastic Surgery: The Intuitive…Field failures were reported related to the da Vinci Xi Surgical System EndoWrist Stapler 45 and 30… 07/14/2016Terminated
Z-2675-2016 Class IISTAPLER,IS4000; Model number 470298; General and Plastic Surgery: The Intuitive Surgical…Field failures were reported related to the da Vinci Xi Surgical System EndoWrist Stapler 45 and 30… 07/14/2016Terminated
Z-2350-2016 Class IICURVED-TIP STAPLER 30,IS4000, model number 470530; Gastroenterology/Urology: The…An increase in the number of da Vinci EndoWrist Stapler complaints where the customer experienced a Stapler… 07/14/2016Terminated
Z-2349-2016 Class IISTAPLER 30, IS4000, model number 470430; Gastroenterology/Urology: The Intuitive…An increase in the number of da Vinci EndoWrist Stapler complaints where the customer experienced a Stapler… 07/14/2016Terminated
Z-2348-2016 Class IISTAPLER 45, IS4000, model number 470298 Gastroenterology/Urology: The Intuitive Surgical…An increase in the number of da Vinci EndoWrist Stapler complaints where the customer experienced a Stapler… 07/14/2016Terminated
Z-1837-2016 Class IISEAL,IS4000 PORTS,12MM&STAPLER,BOX of 10; Model numbers: 470380-04, 470380-05…Potential for particulates in insufflation stopcocks on the da Vinci Xi Surgical System cannula seals. The… 05/09/2016Terminated
Z-1836-2016 Class IISEAL,IS4000 PORTS,5MM-8MM,BOX of 10; Model numbers: 470361-03, 470361-04, 470361-05…Potential for particulates in insufflation stopcocks on the da Vinci Xi Surgical System cannula seals. The… 05/09/2016Terminated
Z-1568-2016 Class IIEndoWrist Stapler 45 and Stapler 30 instruments used on the da Vinci Xi systems (IS4000)…Potential for unexpected motion of the Xi Stapler jaws on the da Vinci Xi System with p5 software relating… 03/25/2016Terminated
Z-1377-2016 Class IIda Vinci S Surgical System (IS4000) user manual English and OUS. Product Usage: The…Intuitive Surgical has found that the use of instruments other than the Cardiac Probe Grasper (e.g. Large… 03/16/2016Terminated
Z-1376-2016 Class IIda Vinci S Surgical System (IS1200/IS2000/IS3000) user manual, instrument & accessory.…Intuitive Surgical has found that the use of instruments other than the Cardiac Probe Grasper (e.g. Large… 03/16/2016Terminated
Z-1191-2016 Class IIda Vinci Si (IS3000) and da Vinci Xi (IS4000) Surgeon Side Console (SSC)Touchpad; Model… 02/16/2016Terminated
Z-0653-2016 Class II30 deg, 8mm Endoscope, IS4000: Model numbers: 470027-34, 470027-40, 470027-41. General…This correction to the da Vinci Xi I&A User Manual communicates proper use of the da Vinci Xi 8 mm… 11/23/2015Terminated
Z-0652-2016 Class II0 deg, 8mm Endoscope, IS4000: Model numbers: 470026-34, 470026-40, 470026-41. General…This correction to the da Vinci Xi I&A User Manual communicates proper use of the da Vinci Xi 8 mm… 11/23/2015Terminated
Z-2782-2015 Class IIFluorescence Imaging Procedure Kits used with the da Vinci Si and Xi Surgical Systems…B. Braun 2 gang 4-way stopcocks in the Fluorescence Imaging Procedure Kit may experience damage to the… 09/09/2015Terminated
Z-2734-2015 Class IIda Vinci Xi Surgical System P4 software;with Xi Stapler; model number IS4000 : General…Complaints regarding the Exposed Knife Blade recoverable fault for the Xi Stapler with Xi System P4 version… 08/17/2015Terminated
Z-2758-2015 Class IIFenestrated Bipolar Forceps; used with the da Vinci Si Surgical System (IS3000); Model…In certain circumstances involving select da Vinci Si Single-Site instruments, the jaws of grip actuated… 08/08/2015Terminated
Z-2757-2015 Class IICurved Needle Driver; used with the da Vinci Si Surgical System (IS3000); Model numbers:…In certain circumstances involving select da Vinci Si Single-Site instruments, the jaws of grip actuated… 08/08/2015Terminated
Z-2756-2015 Class IIMaryland Bipolar Forceps; used with the da Vinci Si Surgical System (IS3000); Model…In certain circumstances involving select da Vinci Si Single-Site instruments, the jaws of grip actuated… 08/08/2015Terminated
Z-2755-2015 Class IICrocodile Grasper; used with the da Vinci Si Surgical System (IS3000); Model numbers:…In certain circumstances involving select da Vinci Si Single-Site instruments, the jaws of grip actuated… 08/08/2015Terminated
Z-2754-2015 Class IIFundus Grasper; used with the da Vinci Si Surgical System (IS3000); Model numbers:…In certain circumstances involving select da Vinci Si Single-Site instruments, the jaws of grip actuated… 08/08/2015Terminated
Z-2753-2015 Class IINeedle Driver; used with the da Vinci Si Surgical System (IS3000); Model numbers:…In certain circumstances involving select da Vinci Si Single-Site instruments, the jaws of grip actuated… 08/08/2015Terminated
Z-2752-2015 Class IICadiere Forceps; used with the da Vinci Si Surgical System (IS3000); Model numbers:…In certain circumstances involving select da Vinci Si Single-Site instruments, the jaws of grip actuated… 08/08/2015Terminated
Z-2751-2015 Class IIMedium -Large Clip Applier; used with the da Vinci Si Surgical System (IS3000); Model…In certain circumstances involving select da Vinci Si Single-Site instruments, the jaws of grip actuated… 08/08/2015Terminated
Z-2750-2015 Class IIMaryland Dissector; used with the da Vinci Si Surgical System (IS3000); Model numbers:…In certain circumstances involving select da Vinci Si Single-Site instruments, the jaws of grip actuated… 08/08/2015Terminated
Z-1949-2015 Class IIINSTRUMENT ARM DRAPE, IS4000, 20 PACK; used on the da Vinci Xi Surgical System. The…Field Safety Notification to advise of an issue that may occur involving the combination of the da Vinci Xi… 06/16/2015Terminated
Z-1812-2015 Class IIThoracic Grasper, 5 MM, for use with da Vinci S (IS2000) and SI (IS3000) Surgical…The coating of the 5 mm Thoracic Grasper main tube may experience damage with the possibility of fragments… 05/21/2015Terminated
Z-1393-2015 Class IIda Vinci Si Surgical System IS3000, Disposable Accessory Kit, 4 ARM, 5 Pack. Microtek…Cloudy/waxy appearance with potential to transfer waxy substance to the patient, and potential for drape to… 03/16/2015Terminated
Z-1392-2015 Class IIda Vinci Si Surgical System IS3000, Disposable Accessory Kit, 3 ARM, 5 Pack. Microtek…Cloudy/waxy appearance with potential to transfer waxy substance to the patient, and potential for drape to… 03/16/2015Terminated
Z-1391-2015 Class IIda Vinci Si Surgical System IS3000, Camera Arm Drape, 20 Pack. Microtek Medical…Cloudy/waxy appearance with potential to transfer waxy substance to the patient, and potential for drape to… 03/16/2015Terminated
Z-1390-2015 Class IIda Vinci S Surgical System IS2000, Disposable Accessory Kit, 4 ARM, 5 Pack. Microtek…Cloudy/waxy appearance with potential to transfer waxy substance to the patient, and potential for drape to… 03/16/2015Terminated
Z-1389-2015 Class IIda Vinci S Surgical System IS2000, Disposable Accessory Kit, 3 ARM, 5 Pack. Microtek…Cloudy/waxy appearance with potential to transfer waxy substance to the patient, and potential for drape to… 03/16/2015Terminated
Z-1388-2015 Class IIda Vinci S / Si Surgical System IS2000 / IS3000, Camera Arm Drape, 20 Pack. Microtek…Cloudy/waxy appearance with potential to transfer waxy substance to the patient, and potential for drape to… 03/16/2015Terminated
Z-1387-2015 Class IIda Vinci S / Si Surgical System IS2000 / IS3000, Instrument Arm Drape, 20 Pack. Microtek…Cloudy/waxy appearance with potential to transfer waxy substance to the patient, and potential for drape to… 03/16/2015Terminated
Z-1386-2015 Class IIda Vinci Si Surgical System IS3000, Monitor Drape, 20 Pack. Microtek Medical Equipment…Cloudy/waxy appearance with potential to transfer waxy substance to the patient, and potential for drape to… 03/16/2015Terminated
Z-1385-2015 Class IIda Vinci Si Surgical System IS3000, Camera Head Drape, 20 Pack. Microtek Medical…Cloudy/waxy appearance with potential to transfer waxy substance to the patient, and potential for drape to… 03/16/2015Terminated
Z-1384-2015 Class IIda Vinci S Surgical System IS2000, Vision-Cart Monitor Drape, 20 Pack. Microtek Medical…Cloudy/waxy appearance with potential to transfer waxy substance to the patient, and potential for drape to… 03/16/2015Terminated
Z-1383-2015 Class IIda Vinci S Surgical System IS2000, Patient-Cart Monitor Drape, 20 Pack. Microtek Medical…Cloudy/waxy appearance with potential to transfer waxy substance to the patient, and potential for drape to… 03/16/2015Terminated
Z-1382-2015 Class IIda Vinci S Surgical System IS2000, Camera Drape, 20 Pack. Microtek Medical Equipment…Cloudy/waxy appearance with potential to transfer waxy substance to the patient, and potential for drape to… 03/16/2015Terminated
Z-1381-2015 Class IIda Vinci S Surgical System IS2000, Camera Arm Drape, 20 Pack. Microtek Medical Equipment…Cloudy/waxy appearance with potential to transfer waxy substance to the patient, and potential for drape to… 03/16/2015Terminated
Z-1380-2015 Class IIda Vinci S Surgical System IS2000, Arm Drape, 20 Pack. Microtek Medical Equipment Drapes…Cloudy/waxy appearance with potential to transfer waxy substance to the patient, and potential for drape to… 03/16/2015Terminated
Z-1304-2015 Class IIda Vinci Xi Surgical System, IS4000; ASSY,USM,IS4000, Material number: 380647-20 and…Correction due to a fault that causes distal SUJ (DSUJ) failures that indicates a mismatch between the… 03/03/2015Terminated
Z-1303-2015 Class IIda Vinci Xi Surgical System, IS4000; ASSY,DISTAL SUJ,OUTER,IS4000, Material number:…Correction due to a fault that causes distal SUJ (DSUJ) failures that indicates a mismatch between the… 03/03/2015Terminated
Z-1302-2015 Class IIda Vinci Xi Surgical System, IS4000; Assy, Distal SUJ, Inner, IS4000 Material number:…Correction due to a fault that causes distal SUJ (DSUJ) failures that indicates a mismatch between the… 03/03/2015Terminated
Z-1226-2015 Class IIVacuum Source Tubing Set for the EndoWrist Stabilizer for da Vinci S (IS2000) and Si…There is the possibility that the Vacuum Source Tubing set for the Endo Wrist Stabilizer may have the outer… 02/16/2015Terminated
Z-1142-2015 Class IIEndoWrist One Vessel Sealer, 8MM, IS4000, 6 pack used in conjunction with the da Vinci…Intuitive Surgical has identified the potential for a certain, small portion of Vessel Sealers to exhibit… 02/11/2015Terminated
Z-1141-2015 Class IIEndoWrist One Vessel Sealer, IS3000, 6 pack used in conjunction with the da Vinci…Intuitive Surgical has identified the potential for a certain, small portion of Vessel Sealers to exhibit… 02/11/2015Terminated
Z-1118-2015 Class IIHARMONIC ACE Curved Shears Instructions for Use, 8MM, IS2000/IS3000; for use with the da…Internal labeling review found that the IFU for the Harmonic Ace incorrectly instructs the user on how to… 01/21/2015Terminated
Z-1117-2015 Class IIHARMONIC ACE Curved Shears Instructions for Use, 5MM, IS1200; for use with the da Vinci…Internal labeling review found that the IFU for the Harmonic Ace incorrectly instructs the user on how to… 01/21/2015Terminated
Z-1048-2015 Class IIda Vinci Si Surgical System, IS3000; - Wall chart for EndoWrist Stapler 45 System.…The Wall chart has been updated because it was noted that Wall Chart (PN 551524-03 Rev A) was inconsistent… 01/15/2015Terminated
Z-0940-2015 Class IIIS3000 Vision Side Cart (VSC) used in conjunction with the da Vinci Si Surgical System…Intuitive Surgical is initiating a voluntary correction related to da Vinci Si System (IS3000) Vision Side… 12/05/2014Terminated
Z-0835-2015 Class IIFluorescence Imaging Procedure Kit, designed for use with Firefly Fluorescence Imaging…The storage conditions for indocyanine green (ICG) are not on the individual Fluorescence Imaging Procedure… 11/21/2014Terminated
Z-0834-2015 Class IIFluorescence Imaging Procedure Kit, designed for use with Firefly Fluorescence Imaging…The storage conditions for indocyanine green (ICG) are not on the individual Fluorescence Imaging Procedure… 11/21/2014Terminated
Z-0448-2015 Class II8MM, TIP-UP FENESTRATED GRASPER; Intended to assist in the accurate control of Intuitive…Deviations in reprocessing steps from those stated in the reprocessing instructions can cause surface… 11/03/2014Terminated
Z-0447-2015 Class II8MM, CURVED BIPOLAR DISSECTOR; Intended to assist in the accurate control of Intuitive…Deviations in reprocessing steps from those stated in the reprocessing instructions can cause surface… 11/03/2014Terminated
Z-0446-2015 Class II8MM,MEDIUM-LARGE CLIP APPLIER ; Intended to assist in the accurate control of Intuitive…Deviations in reprocessing steps from those stated in the reprocessing instructions can cause surface… 11/03/2014Terminated
Z-0445-2015 Class II8MM, SMALL GRASPING RETRACTOR ; Intended to assist in the accurate control of Intuitive…Deviations in reprocessing steps from those stated in the reprocessing instructions can cause surface… 11/03/2014Terminated
Z-0444-2015 Class II8MM, MEGA SUTURECUT ND; Intended to assist in the accurate control of Intuitive Surgical…Deviations in reprocessing steps from those stated in the reprocessing instructions can cause surface… 11/03/2014Terminated
Z-0443-2015 Class II8MM, DUAL BLADE RETRACTOR; Intended to assist in the accurate control of Intuitive…Deviations in reprocessing steps from those stated in the reprocessing instructions can cause surface… 11/03/2014Terminated
Z-0442-2015 Class II8MM, ATRIAL RETRACTOR SHORT RIGHT ; Intended to assist in the accurate control of…Deviations in reprocessing steps from those stated in the reprocessing instructions can cause surface… 11/03/2014Terminated
Z-0441-2015 Class II8MM, LARGE HEM-O-LOK CLIP APPLIER; Intended to assist in the accurate control of… 11/03/2014Terminated
Z-0440-2015 Class II8MM, CARDIAC PROBE GRASPER; Intended to assist in the accurate control of Intuitive…Deviations in reprocessing steps from those stated in the reprocessing instructions can cause surface… 11/03/2014Terminated
Z-0439-2015 Class II8MM, TENACULUM FORCEPS; Intended to assist in the accurate control of Intuitive Surgical…Deviations in reprocessing steps from those stated in the reprocessing instructions can cause surface… 11/03/2014Terminated
Z-0438-2015 Class II8MM FENESTRATED BIPOLAR FORCEPS,; Intended to assist in the accurate control of…Deviations in reprocessing steps from those stated in the reprocessing instructions can cause surface… 11/03/2014Terminated
Z-0437-2015 Class II8MM PERMANENT CAUTERY SPATULA,; Intended to assist in the accurate control of Intuitive…Deviations in reprocessing steps from those stated in the reprocessing instructions can cause surface… 11/03/2014Terminated
Z-0436-2015 Class II8MM PERMANENT CAUTERY HOOK,; Intended to assist in the accurate control of Intuitive…Deviations in reprocessing steps from those stated in the reprocessing instructions can cause surface… 11/03/2014Terminated
Z-0435-2015 Class II8MM RESANO FORCEPS,; Intended to assist in the accurate control of Intuitive Surgical…Deviations in reprocessing steps from those stated in the reprocessing instructions can cause surface… 11/03/2014Terminated
Z-0434-2015 Class II8MM MONOPOLAR CURVED SCISSORS; Intended to assist in the accurate control of Intuitive…Deviations in reprocessing steps from those stated in the reprocessing instructions can cause surface… 11/03/2014Terminated
Z-0433-2015 Class II8MM,MARYLAND BIPOLAR FORCEPS; Intended to assist in the accurate control of Intuitive…Deviations in reprocessing steps from those stated in the reprocessing instructions can cause surface… 11/03/2014Terminated
Z-0432-2015 Class II8MM,PROGRASP FORCEPS; Intended to assist in the accurate control of Intuitive Surgical…Deviations in reprocessing steps from those stated in the reprocessing instructions can cause surface… 11/03/2014Terminated
Z-0431-2015 Class II8MM,MICRO BIPOLAR FORCEPS; Intended to assist in the accurate control of Intuitive…Deviations in reprocessing steps from those stated in the reprocessing instructions can cause surface… 11/03/2014Terminated
Z-0430-2015 Class II8MM,LONG TIP FORCEPS; Intended to assist in the accurate control of Intuitive Surgical…Deviations in reprocessing steps from those stated in the reprocessing instructions can cause surface… 11/03/2014Terminated
Z-0429-2015 Class II8MM,BLACK DIAMOND MICRO FORCEPS; Intended to assist in the accurate control of Intuitive…Deviations in reprocessing steps from those stated in the reprocessing instructions can cause surface… 11/03/2014Terminated
Z-0428-2015 Class II8MM,LARGE NEEDLE DRIVER; Intended to assist in the accurate control of Intuitive…Deviations in reprocessing steps from those stated in the reprocessing instructions, can cause surface… 11/03/2014Terminated
Z-0427-2015 Class II8MM,POTTS SCISSORS; Intended to assist in the accurate control of Intuitive Surgical…Deviations in reprocessing steps from those stated in the reprocessing instructions, can cause surface… 11/03/2014Terminated
Z-0074-2015 Class IIEndoWrist Stapler 45 instrument for the da Vinci Si System Model Number: IS3000 Product…Intuitive is issuing an urgent product notice to all users of the EndoWrist Stapler 45 to discontinue use… 09/18/2014Terminated