Recall Z-0756-2020

GE Healthcare, LLC · initiated September 27, 2019

Class I Terminated

Class I – dangerous or defective products that could cause serious health problems or death

Product

CARESCAPE Respiratory Module E-sCAiO

Codes / lots: Mfg. Lot or Serial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…

Reason for recall

Displayed End Tidal Oxygen (EtO2) and Fraction of Inspired Oxygen (FiO2) values may be incorrect which could cause up to 50% measurement error in the EtO2/FiO2 values.

FDA-determined cause: Process control

Action taken

The firm issued the Urgent Medical Device Correction Letter dated 09/27/2019. The letter describes the failure and requests replacement of the affected module with another module. If a replacement is not available, the consignee is directed to follow the instructions below to ensure that your respiratory module O2 reading is within specifications (a) before any new patient case is started and (b) for prolonged patient cases, at a minimum of once a day: 1) Connect the module to the host device; 2) Ensure the module is fully warmed up by having it powered on for minimum 20 minutes; 3) Configure the O2 reading and waveform to the host device screen as defined in the host device instructions; 4) Feed 100% O2 gas flow from the ventilator to the module and observe the O2 reading and waveform; and 5) Ensure the O2 reading is within specifications 100% +/-3% and there is no erratic behavior in the waveform. 6) If you are unable to complete the above steps successfully, remove the respiratory module from clinical use and quarantine. Contact GE Healthcare Service or your local Service Representative. If you have any questions or concerns regarding this notification, please contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.

Details

Recall number
Z-0756-2020
Initiated
September 27, 2019
Posted by FDA
January 10, 2020
Terminated
September 26, 2023
Product code
CCK – Analyzer, Gas, Carbon-Dioxide, Gaseous-Phase
Quantity
161 units
Distribution
US OUS: Australia, Austria, Belgium, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Czech Republic, Finland, France, Germany, Hong Kong, India, Indonesia, Ireland, Italy, Japan, Korea (Republic of), Malaysia, Mexico, Myanmar, Netherlands, New Zealand, Peru, Poland, Republic of Serbia, Romania, Russia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, Uganda, and…
Recalling firm
GE Healthcare, LLC, Waukesha, WI

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0754-2020 Class ICARESCAPE Respiratory Module E-sCODisplayed End Tidal Oxygen (EtO2) and Fraction of Inspired Oxygen (FiO2) values may be incorrect which could… 09/27/2019Terminated
Z-0755-2020 Class ICARESCAPE Respiratory Module E-sCOVXDisplayed End Tidal Oxygen (EtO2) and Fraction of Inspired Oxygen (FiO2) values may be incorrect which could… 09/27/2019Terminated
Z-0757-2020 Class ICARESCAPE Respiratory Module E-sCAiOVDisplayed End Tidal Oxygen (EtO2) and Fraction of Inspired Oxygen (FiO2) values may be incorrect which could… 09/27/2019Terminated
Z-0758-2020 Class ICARESCAPE Respiratory Module E-sCAiOEDisplayed End Tidal Oxygen (EtO2) and Fraction of Inspired Oxygen (FiO2) values may be incorrect which could… 09/27/2019Terminated
Z-0759-2020 Class ICARESCAPE Respiratory Module E-sCAiOVEDisplayed End Tidal Oxygen (EtO2) and Fraction of Inspired Oxygen (FiO2) values may be incorrect which could… 09/27/2019Terminated
Z-0760-2020 Class IAirway Gas Option N-CAiODisplayed End Tidal Oxygen (EtO2) and Fraction of Inspired Oxygen (FiO2) values may be incorrect which could… 09/27/2019Terminated