Recall Z-0757-2020

GE Healthcare, LLC · initiated September 27, 2019

Class I Terminated

Class I – dangerous or defective products that could cause serious health problems or death

Product

CARESCAPE Respiratory Module E-sCAiOV

Codes / lots: Mfg. Lot or Serial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…

Reason for recall

Displayed End Tidal Oxygen (EtO2) and Fraction of Inspired Oxygen (FiO2) values may be incorrect which could cause up to 50% measurement error in the EtO2/FiO2 values.

FDA-determined cause: Process control

Action taken

The firm issued the Urgent Medical Device Correction Letter dated 09/27/2019. The letter describes the failure and requests replacement of the affected module with another module. If a replacement is not available, the consignee is directed to follow the instructions below to ensure that your respiratory module O2 reading is within specifications (a) before any new patient case is started and (b) for prolonged patient cases, at a minimum of once a day: 1) Connect the module to the host device; 2) Ensure the module is fully warmed up by having it powered on for minimum 20 minutes; 3) Configure the O2 reading and waveform to the host device screen as defined in the host device instructions; 4) Feed 100% O2 gas flow from the ventilator to the module and observe the O2 reading and waveform; and 5) Ensure the O2 reading is within specifications 100% +/-3% and there is no erratic behavior in the waveform. 6) If you are unable to complete the above steps successfully, remove the respiratory module from clinical use and quarantine. Contact GE Healthcare Service or your local Service Representative. If you have any questions or concerns regarding this notification, please contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.

Details

Recall number
Z-0757-2020
Initiated
September 27, 2019
Posted by FDA
January 10, 2020
Terminated
September 26, 2023
Product code
CCK – Analyzer, Gas, Carbon-Dioxide, Gaseous-Phase
Quantity
37 units
Distribution
US OUS: Australia, Austria, Belgium, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Czech Republic, Finland, France, Germany, Hong Kong, India, Indonesia, Ireland, Italy, Japan, Korea (Republic of), Malaysia, Mexico, Myanmar, Netherlands, New Zealand, Peru, Poland, Republic of Serbia, Romania, Russia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, Uganda, and…
Recalling firm
GE Healthcare, LLC, Waukesha, WI

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0754-2020 Class ICARESCAPE Respiratory Module E-sCODisplayed End Tidal Oxygen (EtO2) and Fraction of Inspired Oxygen (FiO2) values may be incorrect which could… 09/27/2019Terminated
Z-0755-2020 Class ICARESCAPE Respiratory Module E-sCOVXDisplayed End Tidal Oxygen (EtO2) and Fraction of Inspired Oxygen (FiO2) values may be incorrect which could… 09/27/2019Terminated
Z-0756-2020 Class ICARESCAPE Respiratory Module E-sCAiODisplayed End Tidal Oxygen (EtO2) and Fraction of Inspired Oxygen (FiO2) values may be incorrect which could… 09/27/2019Terminated
Z-0758-2020 Class ICARESCAPE Respiratory Module E-sCAiOEDisplayed End Tidal Oxygen (EtO2) and Fraction of Inspired Oxygen (FiO2) values may be incorrect which could… 09/27/2019Terminated
Z-0759-2020 Class ICARESCAPE Respiratory Module E-sCAiOVEDisplayed End Tidal Oxygen (EtO2) and Fraction of Inspired Oxygen (FiO2) values may be incorrect which could… 09/27/2019Terminated
Z-0760-2020 Class IAirway Gas Option N-CAiODisplayed End Tidal Oxygen (EtO2) and Fraction of Inspired Oxygen (FiO2) values may be incorrect which could… 09/27/2019Terminated