Recall Z-0853-2011

Accuray Inc · initiated July 21, 2009

Terminated

Product

RoboCouch Patient Support System, model number 025007, a subsystem of the Cyberknife Robotic Radiosurgery System, manufactured by Accuray Incorporated, Sunnyvale, CA. Intended for use in the support and positioning of a patient during Radiosurgery and radiotherapy procedures and other medical procedures when precise positioning is required.

Codes / lots: RoboCouch model 025007

Reason for recall

Mechanical Failure-- Patient Support System's table top may unexpectedly descend several inches.

FDA-determined cause: Device Design

Action taken

Consignees were notified verbally about the issue when an inspection was scheduled.

Details

Recall number
Z-0853-2011
Initiated
July 21, 2009
Posted by FDA
January 11, 2011
Terminated
January 12, 2011
Product code
JAI – Couch, Radiation Therapy, Powered
Quantity
26
Distribution
Worldwide Distribution -- USA, including states of CA, FL, VA, AL, CT, MA, NC, TN, PA, RI, and NE and countries of Turkey, France, UK, Saudi Arabia, and Switzerland.
Recalling firm
Accuray Inc, Sunnyvale, CA