Recall Z-0883-2025
Intuitive Surgical, Inc. · initiated December 19, 2024
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
8MM,FORCE BIPOLAR,IS4000 REF 471405
Codes / lots: Model Number: 471405 Part Number: 471405-06 UDI-DI code: 00886874120767 Batch Numbers: K10221204 K10221212 K10221218 K10230111 K10230212 K10230219 K10230226 K10230302 K10230310 K10230413 K10230420 K10230504 K10230515 K10230518 K10230525 K10230531 K10230601 K10230608 K10230615 K10230622 K10230630 K10230706 K10230713 K10230727 K10230803 K10230810 K10230817 K10230824 K10230831 K10230914 K10230928 K10231005 K10231012 K10231019 K10231027 K10231102 K10231109 K10231123 K10231130 K10231218 K10240104 K10240111 K10240118 K10240124 K10240208 K10240215 K10240219 K10240222 K10240229 K10240307 K10240327 K10240404 K10240411 K10240415 K10240418 K10240419 K10240425 K10240509 K10240516 K10240520 K10240530 K10240711 K10240728 K10240815 K10240822 K11221204 K11221212 K11221218 K11230212 K11230226 K11230302 K11230310 K11230413 K11230420 K11230515 K11230518 K11230525 K11230601 K11230608 K11230615 K11230622 K11230706 K11230713 K11230727 K11230803 K11230810 K11230914 K11231005 K11231019 K11231109 K11231116 K11240104 K11240111 K11240118 K11240208 K11240229 K11240307 K11240411 K11240418 K11240509 K11240516 K11240530 K11240711 K11240815 K11240822 K11240829 K12221204 K12221218 K12230122 K12230212 K12230226…
Reason for recall
Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws.
FDA-determined cause: Process change control
Action taken
On 12/19/2024, the firm send via email an "Urgent: Medical Device Correction" Letter informing customer that Intuitive Surgical has become aware of an increase in complaints regarding frayed or broken cables on some da Vinci X and Xi reusable instruments. Customer are instructed to: 1. Customers can continue using the products in accordance with the user manual. 2. As a reminder, when using da Vinci X and Xi reusable instruments, follow the Inspection Before Use and Warnings listed in the manual provided with your system to inspect for any broken cables. Refer to Figures C and D for examples. In addition, please refer to Appendix B for additional images for detection of grip cable failures. 3. If you observe any failed (frayed or broken) grip cables prior to use, during procedure, or during reprocessing, please stop use of instrument, remove from use and inform Intuitive via the standard complaint process. 4. Inform Intuitive of any Adverse Events*/Serious Incidents** or quality problems concerning the use of the subject device via the standard complaint process. For Question and assistance - " North and South America: (800) 876-1310, Option 3 (4 am to 5 pm PST) or mail: customerservice@intusurg.com. " Europe, Middle East, Asia and Africa: +800 0821 2020 or +41 21 821 2020 (8 am to 6 pm CET) or eucs@intusurg.com " South Korea: 02-3271-3200 (9 am to 6 pm KSTJ) or support.korea@intusurg.com " Japan: 0120-56-5635 or 03-5575-1362 (9 am to 6 pm JST) or csjapan@intusurg.com " India: +1-800-103-6952 (9 am to 6 pm IT) " Taiwan: +0800-86-8181 (9 am to 6 pm CT)
Details
- Recall number
- Z-0883-2025
- Initiated
- December 19, 2024
- Posted by FDA
- January 10, 2025
- Product code
- NAY – System, Surgical, Computer Controlled Instrument
- Quantity
- 46,759
- Distribution
- U.S. - Nation-wide. Alabama, Alaska, Arizona, Arkansas, California, Colorado, Connecticut, Delaware, District of Columbia, Florida, Georgia, Hawaii, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine, Maryland, Massachusetts, Michigan, Minnesota, Mississippi, Missouri, Montana, Nebraska, Nevada, New Hampshire, New Jersey, New Mexico, New York, North Carolina, North Dakota, Ohio…
- Recalling firm
- Intuitive Surgical, Inc., Sunnyvale, CA
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-0847-2025 Class II | 8MM, MONOPOLAR CURVED SCISSORS, IS4000 REF 470179Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws. | 12/19/2024Open, Classified |
| Z-0848-2025 Class II | 8MM, POTTS SCISSORS, IS4000 REF 470001Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws. | 12/19/2024Open, Classified |
| Z-0849-2025 Class II | 8MM, LARGE NEEDLE DRIVER, IS4000 REF 470006Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws. | 12/19/2024Open, Classified |
| Z-0850-2025 Class II | 8MM, ROUND TIP SCISSORS, IS4000 REF 470007Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws. | 12/19/2024Open, Classified |
| Z-0851-2025 Class II | 8MM,BLACK DIAMOND MICRO FORCEPS,IS4000 REF 470033Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws. | 12/19/2024Open, Classified |
| Z-0852-2025 Class II | 8MM,DEBAKEY FORCEPS,IS4000 REF 470036Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws. | 12/19/2024Open, Classified |
| Z-0853-2025 Class II | 8MM,CADIERE FORCEPS,IS4000 REF 470049Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws. | 12/19/2024Open, Classified |
| Z-0854-2025 Class II | 8MM,PROGRASP FORCEPS,IS4000 REF 470093Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws. | 12/19/2024Open, Classified |
| Z-0855-2025 Class II | 8MM,MICRO BIPOLAR FORCEPS,IS4000 REF 470171Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws. | 12/19/2024Open, Classified |
| Z-0856-2025 Class II | 8MM,MARYLAND BIPOLAR FORCEPS,IS4000 REF 470172Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws. | 12/19/2024Open, Classified |
| Z-0857-2025 Class II | 8MM,RESANO FORCEPS,IS4000 REF 470181Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws. | 12/19/2024Open, Classified |
| Z-0858-2025 Class II | 8MM,MEGA NEEDLE DRIVER,IS4000 REF 470194Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws. | 12/19/2024Open, Classified |
| Z-0859-2025 Class II | 8MM,FENESTRATED BIPOLAR FORCEPS,IS4000 REF 470205Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws. | 12/19/2024Open, Classified |
| Z-0860-2025 Class II | 8MM,TENACULUM FORCEPS,IS4000 REF 470207Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws. | 12/19/2024Open, Classified |
| Z-0861-2025 Class II | 8MM,LARGE HEM-O-LOK CLIP APPLIER,IS4000 REF 470230Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws. | 12/19/2024Open, Classified |
| Z-0862-2025 Class II | 8MM,ATRIAL RETRACTOR SHORT RIGHT,IS4000 REF 470246Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws. | 12/19/2024Open, Classified |
| Z-0863-2025 Class II | 8MM,DUAL BLADE RETRACTOR,IS4000 REF 470249Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws. | 12/19/2024Open, Classified |
| Z-0864-2025 Class II | 8MM,LARGE SUTURECUT NEEDLE DRIVER,IS4000 REF 470296Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws. | 12/19/2024Open, Classified |
| Z-0865-2025 Class II | 8MM,MEGA SUTURECUT ND,IS4000 REF 470309Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws. | 12/19/2024Open, Classified |
| Z-0866-2025 Class II | 8MM,SMALL GRASPING RETRACTOR,IS4000 REF 470318Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws. | 12/19/2024Open, Classified |
| Z-0867-2025 Class II | 8MM,MEDIUM-LARGE CLIP APPLIER,IS4000 REF 470327Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws. | 12/19/2024Open, Classified |
| Z-0868-2025 Class II | 8MM,CURVED BIPOLAR DISSECTOR,IS4000 REF 470344Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws. | 12/19/2024Open, Classified |
| Z-0869-2025 Class II | 8MM,TIP-UP FENESTRATED GRASPER,IS4000 REF 470347Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws. | 12/19/2024Open, Classified |
| Z-0870-2025 Class II | 8MM,HORIZON SMALL CLIP APPLIER,IS4000 REF 470401Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws. | 12/19/2024Open, Classified |
| Z-0871-2025 Class II | 8MM,FORCE BIPOLAR,IS4000 REF 470405Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws. | 12/19/2024Open, Classified |
| Z-0872-2025 Class II | 8MM,LONG TIP FORCEPS,IS4000 REF 471048Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws. | 12/19/2024Open, Classified |
| Z-0873-2025 Class II | 8MM,CADIERE FORCEPS,IS4000 REF 471049Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws. | 12/19/2024Open, Classified |
| Z-0874-2025 Class II | 8MM,PROGRASP FORCEPS,IS4000 REF 471093Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws. | 12/19/2024Open, Classified |
| Z-0875-2025 Class II | 8MM,MICRO BIPOLAR FORCEPS,IS4000 REF 471171Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws. | 12/19/2024Open, Classified |
| Z-0876-2025 Class II | 8MM,MARYLAND BIPOLAR FORCEPS,IS4000 REF 471172Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws. | 12/19/2024Open, Classified |