Recall Z-1118-2015

Intuitive Surgical, Inc. · initiated January 21, 2015

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

HARMONIC ACE Curved Shears Instructions for Use, 8MM, IS2000/IS3000; for use with the da Vinci Surgical System - IS2000 and IS3000. Intended for soft tissue incisions when bleeding control and minimal thermal injury are desired. It is used in conjunction with the da Vinci Surgical System

Codes / lots: PN 400275, 420275, and 420272.

Reason for recall

Internal labeling review found that the IFU for the Harmonic Ace incorrectly instructs the user on how to manually open the grips.

FDA-determined cause: Error in labeling

Action taken

A Field Safety Notice entitled, Urgent Medical Device Correction, and dated, January 21, 2015, were sent to customers. The letter identified the problem and the reason for the recall as well as the risk to health. Customers are asked to ensure that all affected personnel are informed of the notice. Additionally, customers are to discard their current Harmonic Ace IFUs. Customers are to refer to the specified manual, reference guide, and instruction card mentioned in the letter. A copy of the notice is to be retained with the da Vinci Standard, S or Si user manual. Intuitive Surgical will remove the Harmonic ACE IFU from all future shipments of the product. Representatives will be available by phone to answer questions related to the Correction.

Details

Recall number
Z-1118-2015
Initiated
January 21, 2015
Posted by FDA
February 13, 2015
Terminated
June 1, 2015
Product code
NAY – System, Surgical, Computer Controlled Instrument
Quantity
12,467 - both sizes
Distribution
Worldwide Distribution -- US, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Czech Republic, Denmark, Dominican Republic, Ecuador, Finland, France, Germany, India, Israel, Italy, Japan, Lebanon, Mexico, Monaco, Netherlands, Pakistan, Panama, Puerto Rico, Qatar, Romania, Russian Federation, Saudi Arabia, Singapore, Slovakia, South Korea, Spain, Sweden, Switzerland, Taiwan…
Recalling firm
Intuitive Surgical, Inc., Sunnyvale, CA

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-1117-2015 Class IIHARMONIC ACE Curved Shears Instructions for Use, 5MM, IS1200; for use with the da Vinci…Internal labeling review found that the IFU for the Harmonic Ace incorrectly instructs the user on how to… 01/21/2015Terminated