Recall Z-1123-2012

Mallinckrodt Inc · initiated February 10, 2012

Terminated

Product

Coiled 60" Low Pressure Extension Tube 400 PSI Max; packaged in Plastic and Tyvec pouch, with 100 pouches per case; Liebel-Flarsheim Company, Cincinnati, OH 45237. Made in Mexico. Sterile, The purpose of this product is the injection of contrast media by means of power injector system during a diagnostic procedure.

Codes / lots: Catalog #: 601195; Lot #:0001101, 0001102, 0001103, 0004086, 0008405, 0008406, 0008407, 0008408, 0015031, 0015032, 0015033, 0015034, 0015035, 0022157, 0022158, 0022159, 0022160, 0022161, 0029487, 0029488, 0029489, 0029490, 0029491, 0036242, 0036243, 0036244, 0036245, 0036246, 0039037, 0043271, 0043272, 0043273, 0043274, 0043275, 0053046, 0057333, 0057334, 0057335, 0057336, 0057613, 0057614, 0064338, 0064339, 0064340, 0064341, 0064342, 0071225, 0071226, 0071227, 0071228, 0071229, 0078225, 0078226, 0078227, 0078228, 0078229, 0081107, 0085369, 0085370, 0085371, 0085372, 0088166, 0092279, 0092280, 0095183, 0099103, 0099104, 0099105, 0106068, 0106069, 0106070, 0106071, 0113415, 0113416, 0113417, 0113418, 0120203, 0120204, 0120205, 0127197, 0127198, 0127199, 0134337, 0134338, 0134339, 0141203, 0141204, 0141205, 0148232, 0148233, 0148234, 0169313, 0176210, 0176211, 0182133, 0182134, 0190241, 0190242, 0190243, 0197169, 0197170, 0197171, 0197172, 0204150, 0204151, 0204152, 0204153, 0211150, 0211151, 0211152, 0211153, 0211154, 0218187, 0218188, 0218189, 0218190, 0225129, 0225130, 0225131, 0225132, 0232144, 0232145, 0239174, 0239175, 0239176, 0239177, 0246191, 0246192, 0246193, 0246194…

Reason for recall

The machine used to package the product was missing a tool, which could prevent the packaging from sealing properly, causing the sterility of the product to be compromised.

FDA-determined cause: Packaging process control

Action taken

Covidien sent an Urgent Device Recall letter dated February 10, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to discontinue use of the affected product immediately. Customers were to use the attached business response form to report the quantity of each lot that is currently in their stock. If customers did not have any of these lots in stock, they should indicate by checking the appropriate box on the business report form. When completed, customers should email the completed form to covidien2600@stericycle.com or fax to 888-257-7910. For any questions customers should call 866-761-9571.

Details

Recall number
Z-1123-2012
Initiated
February 10, 2012
Posted by FDA
February 29, 2012
Terminated
December 24, 2013
Product code
DXT – Injector And Syringe, Angiographic
Distribution
Worldwide Distribution - USA including Distributed worldwide: USA including the states of: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY, Puerto Rico and the countries of Canada, Australia, Belgium, Brazil, Chile, Ecuador, Guatemala, Mexico…
Recalling firm
Mallinckrodt Inc, Cincinnati, OH

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-1118-2012 Liebel-Flarsheim Optistar MR 60 mL Syringe w/lateral flow needle, Mallinckrodt…The machine used to package the product was missing a tool, which could prevent the packaging from sealing… 02/10/2012Terminated
Z-1119-2012 Liebel-Flarsheim Optistar MR 25 mL Syringe w/lateral flow needle, Mallinckrodt…The machine used to package the product was missing a tool, which could prevent the packaging from sealing… 02/10/2012Terminated
Z-1120-2012 Mallinckrodt Optistar Elite 60" Y-Tubing with Single Check Valve, Liebel-Flarsheim…The machine used to package the product was missing a tool, which could prevent the packaging from sealing… 02/10/2012Terminated
Z-1121-2012 Mallinckrodt Optistar Elite 90" Y-Tubing with Single Check Valve, Liebel-Flarsheim…The machine used to package the product was missing a tool, which could prevent the packaging from sealing… 02/10/2012Terminated
Z-1122-2012 Mallinckrodt Optistar Elite 60 mL Syringe with Lateral Flow Needle, Liebel-Flarsheim…The machine used to package the product was missing a tool, which could prevent the packaging from sealing… 02/10/2012Terminated