Mallinckrodt Inc

Medical device manufacturer in Bedminster, NJ.

2510(k) clearances
0PMA approvals
16recalls

Device types

510(k) clearances

510(k)DeviceDecision
K003075Knightstar 330 VentilatorMNT · Traditional 12/11/2001ST
K003315Shiley Tracheosoft XLT Extended Length Tracheostomy Tube and Disposable Inner CannulaJOH · Special 11/08/2000SE

Recalls

RecallProductInitiated
Z-1745-2022 Class IIINOmax DSIR -(delivery system) delivers INOMAX (nitric oxide for inhalation) therapy gas…The INOmax DSIR Delivery System, manufactured by Mallinckrodt Pharmaceuticals, is not compatible with the… 06/23/2022Open, Classified
Z-1123-2012 Coiled 60" Low Pressure Extension Tube 400 PSI Max; packaged in Plastic and Tyvec pouch…The machine used to package the product was missing a tool, which could prevent the packaging from sealing… 02/10/2012Terminated
Z-1122-2012 Mallinckrodt Optistar Elite 60 mL Syringe with Lateral Flow Needle, Liebel-Flarsheim…The machine used to package the product was missing a tool, which could prevent the packaging from sealing… 02/10/2012Terminated
Z-1121-2012 Mallinckrodt Optistar Elite 90" Y-Tubing with Single Check Valve, Liebel-Flarsheim…The machine used to package the product was missing a tool, which could prevent the packaging from sealing… 02/10/2012Terminated
Z-1120-2012 Mallinckrodt Optistar Elite 60" Y-Tubing with Single Check Valve, Liebel-Flarsheim…The machine used to package the product was missing a tool, which could prevent the packaging from sealing… 02/10/2012Terminated
Z-1119-2012 Liebel-Flarsheim Optistar MR 25 mL Syringe w/lateral flow needle, Mallinckrodt…The machine used to package the product was missing a tool, which could prevent the packaging from sealing… 02/10/2012Terminated
Z-1118-2012 Liebel-Flarsheim Optistar MR 60 mL Syringe w/lateral flow needle, Mallinckrodt…The machine used to package the product was missing a tool, which could prevent the packaging from sealing… 02/10/2012Terminated
Z-0815-2012 Handi-Fil Disposable Syringe Fill Tube P/N 302050, Plastic & Tyvec pouch, Mallinckrodt…The firm received complaints from their customers stating that they found open seals on two of their… 12/20/2011Terminated
Z-0830-2012 Illumena 150mL Linden Luer Syringe with Handi-Fil Straw (p/n 900103) & Illumena 150mL…Syringes packaged and labeled as the Illumena 150mL Linden Luer Syringe with Handi-Fil Straw, product… 10/11/2011Terminated
Z-2859-2011 Angiomat Illumena Syringe, 150mL Linden Luer Syringe with Handi-Fil Straw (p/n 900103)…Syringes packaged and labeled as the Illumena 150mL Linden Luer Syringe with Handi-Fil Straw, product… 05/20/2011Terminated
Z-1904-2010 Hydradjust IV DR Urological Table Sedecal Generator (p/n 750732 - 64W generator & p/n…Covidien received complaints of the Sedecal Generator failing and producing smoke at customer sites. The… 06/14/2010Terminated
Z-2081-2012 Class IIThe Handi-Fil Disposable Syringe Fill Tube is a sterile plastic tube, approximately 10…A report was received from a Japanese distributor regarding a complaint from a hospital customer stating… 02/01/2010Terminated
Z-1828-2008 Covidien/ Mallinckrodt CT9000, CT9000 ADV (J-Bow component 241300), contrast media…The tubing metal begins to crack and thin. When this occurs it causes the tubing material to crack around… 04/14/2008Terminated
Z-1694-2008 The OptiVantage DH Injection System is a contrast delivery system and is designed to…Screws can begin to loosen and shear off to the point of the J-Bow falling from the suspension system. This… 04/07/2008Terminated
Z-0056-2008 Angiomat Illumena Injector with Auxiliary Console, Mallinckrodt, Inc., Hazelwood, MO.Synchronization problems: units that have both a main programmable console and an auxiliary programmable… 08/24/2007Terminated
Z-0288-2007 OptiVantage DH Power Injection System.The flow rate programmed on the console may be changed without automatically updating the flow rate… 11/07/2006Terminated