Recall Z-1146-2026

Edan Diagnostics · initiated October 24, 2025

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Brand Name: EDAN Product Name: Patient Monitor Model/Catalog Number: elite V5, elite V6, elite V8 Product Description: V series Patient Monitor including elite V5, elite V6 and elite V8 which can perform long-time continuous monitoring of multiple physiological parameters. Also, it is capable of storing, displaying, analyzing and controlling measurements, and it will indicate alarms in case of abnormity so that doctors and nurses can deal with them in time. The V series Patient Monitor realize the monitoring of physiological parameters by configuration with different parameter modules which include SpO2 (pulse oxygen saturation, pulse rate and SpO2 plethysmogram) with EDAN SpO2 module or Nellcor SPO2 module, NIBP (systolic pressure, diastolic pressure, mean pressure and pulse rate), TEMP, ECG, RESP(respiration), CO2, IBP, C.O. and AG(anesthetic gas), RM(respiratory mechanics), BIS(bispectral index) and ICG(impedance cardiography). Component: No

Codes / lots: EDAN Model: elite V8 UDI-DI code: 06944413801707 Serial/Lot Numbers: 260342-M17B09290001 002338-M23226990001 002338-M23226990002 002338-M23226990004 002338-M23226990003 002338-M23812870001 002338-M23C15870001 002338-M24A12130001 260342-M16506570001 260342-M15A03180001 002338-K22100150003 002338-K22100150004 002338-K22100150002 002338-K22100150001 261241-M22701020001 002338-M25705810020 002338-M25705810021 002338-M25705810011 002338-M25705810003 002338-M25705810013 002338-M25705810029 002338-M25705810033 002338-M25705810001 002338-M25705810010 002338-M25705810009 002338-M25705810004 002338-M25705810012 002338-M25705810018 002338-M25705810032 002338-M25705810017 002338-M25705810019 002338-M25705810024 002338-M25705810030 002338-M25705810031 002338-M25705810025 002338-M25705810016 002338-M25705810008 002338-M25705810006 002338-M25705810002 002338-M25705810028 002338-M25705810027 002338-M25705810022 002338-M25705810026 002338-M25705810023 002338-M25705810015 002338-M25705810014 002338-M25705810005 002338-M25705810007 002338-M25908680004 002338-M25908680003 002338-M25908680002 002338-M25908680001 EDAN Model: elite V5 UDI-DI code: 06944413802155 Serial/Lot Numbers: 260586-M17211520002…

Reason for recall

As a result of receiving an FDA It Has Come to Our Attention Letter (IHCTOA Letter) regarding potential cybersecurity issues with various of the firm device.

FDA-determined cause: Device Design

Action taken

On 10/24/2025, the firm emailed notifying customers of a product recall. On 11/06/2025, the firm emailed an "URGENT: MEDICAL DEVICE CORRECTION" Letter to customers informing them of labeling issues involving cybersecurity which arose from a communication from the U.S. Food & Drug Administration. -First issue involves an inappropriate publicly routable IP address that was used as an example in the user manual for demonstrating user-configurable IP settings and Second issue involves an inappropriate publicly IP address that was used as the factory default configuration. Customer are instructed to: 1. Immediately examine their inventory and quarantine product subject to correction. 2. If customers may have further distributed affected product, identify their customers and notify them at once of this product correction. The notification to their customers may be enhanced by including the enclosed URGENT MEDICAL DEVICE CORRECTION. 3. For affected product still in their inventory, place copy of the enclosed Notice Before Use and Corrigenda of User Manual into its package before distribution. 4. For affected product that has been distributed or is in use: (1) Confirm that they have received and acknowledged Corrigenda of User Manual. Print out a copy, attach to the User Manual. (2) Confirm the publicly routable IP address 202.114.4.* is NOT used in the User Maintain in the device. If 202.114.4.* is being used, change to appropriate IP settings of device in the Network Setup according to Instruction Instructions for IP Settings Configuration. 5. Complete the attached Acknowledgement and Receipt Form, sign it, and return a scanned copy to the sender of this notification by email. For questions - Contact: EDAN Instruments, Inc. Monday to Friday 09:00-17:00 (UTC+8) Tel:+86-0755-26898326 EDAN Diagnostics, Inc. Monday to Friday 09:00-17:00 (PST) Tel: 1-(858)-750-3066

Details

Recall number
Z-1146-2026
Initiated
October 24, 2025
Posted by FDA
January 22, 2026
Product code
MHX – Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
Quantity
342
Distribution
Worldwide - U.S. Nationwide distribution in the states of CA, KY, PR, FL, CT, IL, PA, RI, NC, NY, LA, TX, MI and the country of Mexico.
Recalling firm
Edan Diagnostics, San Diego, CA

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-1140-2026 Class IIBrand Name: EDAN Product Name: Fetal Monitor Model/Catalog Number: F2, F3 Product…As a result of receiving an FDA It Has Come to Our Attention Letter (IHCTOA Letter) regarding potential… 10/24/2025Open, Classified
Z-1141-2026 Class IIBrand Name: EDAN Product Name: Fetal & Maternal Monitor Model/Catalog Number: F6, F9, F6…As a result of receiving an FDA It Has Come to Our Attention Letter (IHCTOA Letter) regarding potential… 10/24/2025Open, Classified
Z-1142-2026 Class IIBrand Name: EDAN Product Name: Vital Signs Monitor Model/Catalog Number: M3 Product…As a result of receiving an FDA It Has Come to Our Attention Letter (IHCTOA Letter) regarding potential… 10/24/2025Open, Classified
Z-1143-2026 Class IIBrand Name: EDAN Product Name: Vital Signs Monitor Model/Catalog Number: iM3 Product…As a result of receiving an FDA It Has Come to Our Attention Letter (IHCTOA Letter) regarding potential… 10/24/2025Open, Classified
Z-1144-2026 Class IIBrand Name: EDAN Product Name: Vital Signs Monitor Model/Catalog Number: iM3s, iM3As…As a result of receiving an FDA It Has Come to Our Attention Letter (IHCTOA Letter) regarding potential… 10/24/2025Open, Classified
Z-1145-2026 Class IIBrand Name: EDAN Product Name: Patient Monitor Model/Catalog Number: iM50, iM60, iM70…As a result of receiving an FDA It Has Come to Our Attention Letter (IHCTOA Letter) regarding potential… 10/24/2025Open, Classified
Z-1147-2026 Class IIBrand Name: EDAN Product Name: Patient Monitor Model/Catalog Number: iM20 Product…As a result of receiving an FDA It Has Come to Our Attention Letter (IHCTOA Letter) regarding potential… 10/24/2025Open, Classified
Z-1148-2026 Class IIBrand Name: EDAN Product Name: Patient Monitor Model/Catalog Number: X8, X10, X12…As a result of receiving an FDA It Has Come to Our Attention Letter (IHCTOA Letter) regarding potential… 10/24/2025Open, Classified
Z-1149-2026 Class IIBrand Name: EDAN Product Name: Patient Monitor Model/Catalog Number: iX10, iX12, iX15…As a result of receiving an FDA It Has Come to Our Attention Letter (IHCTOA Letter) regarding potential… 10/24/2025Open, Classified
Z-1150-2026 Class IIBrand Name: EDAN Product Name: Central Monitoring System Model/Catalog Number: MFM-CMS…As a result of receiving an FDA It Has Come to Our Attention Letter (IHCTOA Letter) regarding potential… 10/24/2025Open, Classified
Z-1151-2026 Class IIBrand Name: EDAN Product Name: Central Monitoring System Model/Catalog Number: MFM-CNS…As a result of receiving an FDA It Has Come to Our Attention Letter (IHCTOA Letter) regarding potential… 10/24/2025Open, Classified
Z-1152-2026 Class IIBrand Name: EDAN Product Name: Vital Signs Monitor Model/Catalog Number: M3B Product…As a result of receiving an FDA It Has Come to Our Attention Letter (IHCTOA Letter) regarding potential… 10/24/2025Open, Classified
Z-1153-2026 Class IIBrand Name: EDAN Product Name: Vital Signs Monitor Model/Catalog Number: M3A Product…As a result of receiving an FDA It Has Come to Our Attention Letter (IHCTOA Letter) regarding potential… 10/24/2025Open, Classified
Z-1154-2026 Class IIBrand Name: EDAN Product Name: Patient Monitor Model/Catalog Number: iM8, iM8A, iM8B…As a result of receiving an FDA It Has Come to Our Attention Letter (IHCTOA Letter) regarding potential… 10/24/2025Open, Classified
Z-1155-2026 Class IIBrand Name: EDAN Product Name: Telemetry Transmitter Model/Catalog Number: iT20 Product…As a result of receiving an FDA It Has Come to Our Attention Letter (IHCTOA Letter) regarding potential… 10/24/2025Open, Classified