Recall Z-1148-2026
Edan Diagnostics · initiated October 24, 2025
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
Brand Name: EDAN Product Name: Patient Monitor Model/Catalog Number: X8, X10, X12 Product Description: The X8 X10 X12 Patient Monitor (hereinafter called X series) can perform long-time continuous monitoring of multiple physiological parameters. Also, it is capable of storing, displaying, analyzing and controlling measurements, and it will indicate alarms in case of abnormity so that doctors and nurses can deal with them in time. Component: No
Codes / lots: EDAN Model: X8 UDI-DI code: 06944413810907 Serial/Lot Numbers: 261586-M21116620010 261586-M21116620002 261586-M21116620007 261586-M21116620006 261586-M21116620009 261586-M21116630001 261586-M21116620008 261586-M21116620003 261586-M21116630005 261586-M21116630003 261586-M21116620004 261586-M21116620001 261586-M21116630004 261586-M21116630002 261586-M21116620005 261586-M21301970003 261586-M21301970004 261586-M21301960001 261586-M21301970005 261586-M21301960003 261586-M21301970002 261586-M21301970001 261586-M21301960002 261586-M21714500002 261586-M21714500001 261586-M21714500003 261586-M21816750005 261586-M21816750003 261586-M21816750002 261586-M21816750004 261586-M21816750001 261586-M22105750001 261586-M22105750007 261586-M22105750008 261586-M22105750010 261586-M22105750002 261586-M22105750005 261586-M22105750009 261586-M22105750006 261586-M22105750003 261586-M22105750004 261586-M22405430007 261586-M22405430001 261586-M22405430004 261586-M22405430010 261586-M22405430006 261586-M22405430009 261586-M22405430008 261586-M22405430003 261586-M22405430002 261586-M22405430005 261586-M22615270004 261586-M22615270005 261586-M22615270001 261586-M22615270002 261586-M22615270003…
Reason for recall
As a result of receiving an FDA It Has Come to Our Attention Letter (IHCTOA Letter) regarding potential cybersecurity issues with various of the firm device.
FDA-determined cause: Device Design
Action taken
On 10/24/2025, the firm emailed notifying customers of a product recall. On 11/06/2025, the firm emailed an "URGENT: MEDICAL DEVICE CORRECTION" Letter to customers informing them of labeling issues involving cybersecurity which arose from a communication from the U.S. Food & Drug Administration. -First issue involves an inappropriate publicly routable IP address that was used as an example in the user manual for demonstrating user-configurable IP settings and Second issue involves an inappropriate publicly IP address that was used as the factory default configuration. Customer are instructed to: 1. Immediately examine their inventory and quarantine product subject to correction. 2. If customers may have further distributed affected product, identify their customers and notify them at once of this product correction. The notification to their customers may be enhanced by including the enclosed URGENT MEDICAL DEVICE CORRECTION. 3. For affected product still in their inventory, place copy of the enclosed Notice Before Use and Corrigenda of User Manual into its package before distribution. 4. For affected product that has been distributed or is in use: (1) Confirm that they have received and acknowledged Corrigenda of User Manual. Print out a copy, attach to the User Manual. (2) Confirm the publicly routable IP address 202.114.4.* is NOT used in the User Maintain in the device. If 202.114.4.* is being used, change to appropriate IP settings of device in the Network Setup according to Instruction Instructions for IP Settings Configuration. 5. Complete the attached Acknowledgement and Receipt Form, sign it, and return a scanned copy to the sender of this notification by email. For questions - Contact: EDAN Instruments, Inc. Monday to Friday 09:00-17:00 (UTC+8) Tel:+86-0755-26898326 EDAN Diagnostics, Inc. Monday to Friday 09:00-17:00 (PST) Tel: 1-(858)-750-3066
Details
- Recall number
- Z-1148-2026
- Initiated
- October 24, 2025
- Posted by FDA
- January 22, 2026
- Product code
- MHX – Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
- Quantity
- 3824
- Distribution
- Worldwide - U.S. Nationwide distribution in the states of CA, KY, PR, FL, CT, IL, PA, RI, NC, NY, LA, TX, MI and the country of Mexico.
- Recalling firm
- Edan Diagnostics, San Diego, CA
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-1140-2026 Class II | Brand Name: EDAN Product Name: Fetal Monitor Model/Catalog Number: F2, F3 Product…As a result of receiving an FDA It Has Come to Our Attention Letter (IHCTOA Letter) regarding potential… | 10/24/2025Open, Classified |
| Z-1141-2026 Class II | Brand Name: EDAN Product Name: Fetal & Maternal Monitor Model/Catalog Number: F6, F9, F6…As a result of receiving an FDA It Has Come to Our Attention Letter (IHCTOA Letter) regarding potential… | 10/24/2025Open, Classified |
| Z-1142-2026 Class II | Brand Name: EDAN Product Name: Vital Signs Monitor Model/Catalog Number: M3 Product…As a result of receiving an FDA It Has Come to Our Attention Letter (IHCTOA Letter) regarding potential… | 10/24/2025Open, Classified |
| Z-1143-2026 Class II | Brand Name: EDAN Product Name: Vital Signs Monitor Model/Catalog Number: iM3 Product…As a result of receiving an FDA It Has Come to Our Attention Letter (IHCTOA Letter) regarding potential… | 10/24/2025Open, Classified |
| Z-1144-2026 Class II | Brand Name: EDAN Product Name: Vital Signs Monitor Model/Catalog Number: iM3s, iM3As…As a result of receiving an FDA It Has Come to Our Attention Letter (IHCTOA Letter) regarding potential… | 10/24/2025Open, Classified |
| Z-1145-2026 Class II | Brand Name: EDAN Product Name: Patient Monitor Model/Catalog Number: iM50, iM60, iM70…As a result of receiving an FDA It Has Come to Our Attention Letter (IHCTOA Letter) regarding potential… | 10/24/2025Open, Classified |
| Z-1146-2026 Class II | Brand Name: EDAN Product Name: Patient Monitor Model/Catalog Number: elite V5, elite V6…As a result of receiving an FDA It Has Come to Our Attention Letter (IHCTOA Letter) regarding potential… | 10/24/2025Open, Classified |
| Z-1147-2026 Class II | Brand Name: EDAN Product Name: Patient Monitor Model/Catalog Number: iM20 Product…As a result of receiving an FDA It Has Come to Our Attention Letter (IHCTOA Letter) regarding potential… | 10/24/2025Open, Classified |
| Z-1149-2026 Class II | Brand Name: EDAN Product Name: Patient Monitor Model/Catalog Number: iX10, iX12, iX15…As a result of receiving an FDA It Has Come to Our Attention Letter (IHCTOA Letter) regarding potential… | 10/24/2025Open, Classified |
| Z-1150-2026 Class II | Brand Name: EDAN Product Name: Central Monitoring System Model/Catalog Number: MFM-CMS…As a result of receiving an FDA It Has Come to Our Attention Letter (IHCTOA Letter) regarding potential… | 10/24/2025Open, Classified |
| Z-1151-2026 Class II | Brand Name: EDAN Product Name: Central Monitoring System Model/Catalog Number: MFM-CNS…As a result of receiving an FDA It Has Come to Our Attention Letter (IHCTOA Letter) regarding potential… | 10/24/2025Open, Classified |
| Z-1152-2026 Class II | Brand Name: EDAN Product Name: Vital Signs Monitor Model/Catalog Number: M3B Product…As a result of receiving an FDA It Has Come to Our Attention Letter (IHCTOA Letter) regarding potential… | 10/24/2025Open, Classified |
| Z-1153-2026 Class II | Brand Name: EDAN Product Name: Vital Signs Monitor Model/Catalog Number: M3A Product…As a result of receiving an FDA It Has Come to Our Attention Letter (IHCTOA Letter) regarding potential… | 10/24/2025Open, Classified |
| Z-1154-2026 Class II | Brand Name: EDAN Product Name: Patient Monitor Model/Catalog Number: iM8, iM8A, iM8B…As a result of receiving an FDA It Has Come to Our Attention Letter (IHCTOA Letter) regarding potential… | 10/24/2025Open, Classified |
| Z-1155-2026 Class II | Brand Name: EDAN Product Name: Telemetry Transmitter Model/Catalog Number: iT20 Product…As a result of receiving an FDA It Has Come to Our Attention Letter (IHCTOA Letter) regarding potential… | 10/24/2025Open, Classified |