Recall Z-1203-2009

Hamilton Co · initiated December 1, 2008

Terminated

Product

300 microliter CO-RE tips for use with Microlab STAR series instruments, Part Number 235902, 5,760 tips per case, manufactured by Hamilton Company, Reno, NV. The products are a disposable product that are used with the Microlab STAR series instruments, which are microtiter diluting and dispensing devices.

Codes / lots: Lot Number: 2873900.

Reason for recall

Some tips may be occluded and are unable to aspirate or dispense liquid.

FDA-determined cause: Process control

Action taken

An "Urgent: Product Recall Letter" dated December 1, 2008 was issued to consignees via certified mail and e-mail. The letter described the problem and potential effects of use. Consignees were instructed to complete the enclosed "Hamilton Company Product Recall Questionnaire" and return it via e-mail or fax (775-856-7259) to the Hamilton Company. After receipt of the completed questionnaire, the firm's Customer Service Department will contact consignees to provide Returned Material Authorization information for returning affected product. Direct questions to the Hamilton Company at 775-858-3000, ext. 236.

Details

Recall number
Z-1203-2009
Initiated
December 1, 2008
Posted by FDA
April 17, 2009
Terminated
July 21, 2009
Product code
JTC – Device, Microtiter Diluting/Dispensing
Quantity
28 cases
Distribution
Nationwide Distribution -- MA, CA, MN, CT, and NC.
Recalling firm
Hamilton Co, Reno, NV