Recall Z-1221-2010

Leica Microsystems, Inc. · initiated November 12, 2009

Terminated

Product

Shutter HX SP5, a component of the Leica Point Scanning Confocal Microscope Models TCS SP5 and TCS SP5 X. Laboratory instrument.

Codes / lots: Shutter HX SP5, Material No: 1531160006 Spare Part No: 15000904000093

Reason for recall

A manufacturing defect was identified that could potentially result in the laser shutter not closing as intended which could result in the shutter not completely closing disrupting the laser beam safely.

FDA-determined cause: Radiation Control for Health and Safety Act

Action taken

A Service Note was sent to service organizations on November 12, 2009. The communication asked customers to check if they must exchange their shutters and send their planned exchange date by 2009-11-20 to clsm.qc@leica-microsystems.com. Customers were also asked to check the serial numbers of the one-wire chips of the shutters. Then lastly, the were asked to exchange the shutters and confirm the exchange.

Details

Recall number
Z-1221-2010
Initiated
November 12, 2009
Posted by FDA
May 6, 2010
Terminated
May 11, 2010
Product code
REM – Spectroscopy Instrument
Quantity
15 microscopes and 8 spare part shutters
Distribution
Worldwide Distribution -- USA, Sweden, Japan, Hong Kong, Italy, the Netherlands, Singapore, UK, and Spain.
Recalling firm
Leica Microsystems, Inc., Bannockburn, IL