Recall Z-1322-2026

Medline Industries, LP · initiated December 22, 2025

Class I Open, Classified

Class I – dangerous or defective products that could cause serious health problems or death

Product

Reprocessed St. Jude Supreme and Response Diagnostic Electrophysiology Catheters: St. Jude Medical Supreme Diagnostic Electrophysiology Catheter 4 Elec. 2-5-2mm 401211RH 401222RH 401223RH 401226RH 401227RH 401260RH 401261RH 401306RH 401309RH 401312RH 401430RH 401433RH 401435RH 401441RH 401442RH 401443RH 401448RH 401449RH 401450RH 401451RH 401474RH 401860RH 401877RH 401878RH 401890RH 401891RH 401994RH; St. Jude Medical Supreme Diagnostic Electrophysiology Catheter 4 Elec. 5-5-5mm 402012RH; St. Jude Medical Supreme Diagnostic Electrophysiology Catheter 4 Elec. 5mm 402004RH; St. Jude Medical Supreme Diagnostic Electrophysiology Catheter 6 Elec. 2-5-2mm 402010RH; St. Jude Medical Supreme Diagnostic Electrophysiology Catheter 6 Elec. 5-5-5mm 402008RH 402009RH

Codes / lots: 401211RH UDI-DI 10197344043736 Lot EP250317; 401222RH UDI-DI 10197344043712 Lot EP250317; 401223RH UDI-DI 10197344043705 Lot EP250317; 401226RH UDI-DI 10197344043699 Lot EP250307 EP250317; 401227RH UDI-DI 10197344043682 Lots EP250307 EP250317; 401260RH UDI-DI 10197344043668 Lot EP250317; 401261RH UDI-DI 10197344043651 Lot EP250317; 401306RH UDI-DI 10197344043620¿ Lot EP250317; 401309RH UDI-DI 10197344043613 Lot EP250317; 401312RH UDI-DI 10197344043583¿ Lot EP250307; 401430RH UDI-DI 10197344043491 Lots EP250307 EP250317; 401433RH UDI-DI 10197344043484 Lots EP250307 EP250317; 401435RH UDI-DI 10197344043460 Lots EP250307 EP250317; 401441RH UDI-DI 10197344043439 Lots EP250307 EP250317; 401442RH UDI-DI 10197344044245 Lot EP250307; 401443RH UDI-DI 10197344044252 Lots EP250307 EP250317; 401448RH UDI-DI 10197344044283 Lot EP250307; 401449RH UDI-DI 10197344044290 Lot EP250307; 401450RH UDI-DI 10197344044306 Lots EP250307 EP250317; 401451RH UDI-DI 10197344044313 EP250307 EP250317; 401474RH UDI-DI 10197344044351 Lot EP250307; 401860RH UDI-DI 10197344044382 Lots EP250307 EP250317; 401877RH UDI-DI 10197344044436 Lots EP250307 EP250317; 401878RH UDI-DI 10197344044443 Lot EP250307; 401890RH…

Reason for recall

These lots of reprocessed catheters may contain small particles of residual material. If affected devices are used, there is a risk of inflammatory response, systemic infection, or granulomatous reaction or thrombus reaction which may lead to cerebral or pulmonary embolism, or deep vein thrombosis. Expansion of recall Z-2610/2614-2025

FDA-determined cause: Under Investigation by firm

Action taken

On December 22, 2025, the firm issued Urgent Medical Device Recall letters as a follow-up to an earlier action in 2025 for the same issue, but newly identified item numbers and lots of previously identified item numbers. Notably, the products associated with the new/updated action are to be destroyed by the customer on-site for credit, and not returned to the manufacturer. ***Updated 5/18/26*** The firm issued an expanded notification on 5/15/26 which expanded the action to include on additional lot of product. These devices were not expired at the time of the original notification but have since expired.

Details

Recall number
Z-1322-2026
Initiated
December 22, 2025
Posted by FDA
February 6, 2026
Product code
NLH – Catheter, Recording, Electrode, Reprocessed
Quantity
48
Distribution
Distribution US nationwide.
Recalling firm
Medline Industries, LP, Northfield, IL

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-1319-2026 Class IMedline ReNewal Reprocessed Webster CS Catheters: Webster CS Catheter 10 Electrodes…These lots of reprocessed catheters may contain small particles of residual material. If affected devices… 12/22/2025Open, Classified
Z-1320-2026 Class IReprocessed Abbott Inquiry Steerable Diagnostic Catheters: Abbott Inquiry Steerable…These lots of reprocessed catheters may contain small particles of residual material. If affected devices… 12/22/2025Open, Classified
Z-1321-2026 Class IReprocessed Livewire Steerable Diagnostic Electrophysiology Catheters: Livewire…These lots of reprocessed catheters may contain small particles of residual material. If affected devices… 12/22/2025Open, Classified
Z-1323-2026 Class IMedline ReNewal Reprocessed ViewFlex Xtra ICE Catheters: ViewFlex Xtra ICE Catheter 9F x…These lots of reprocessed catheters may contain small particles of residual material. If affected devices… 12/22/2025Open, Classified
Z-1324-2026 Class IReprocessed Biosense Webster SOUNDSTAR eco Diagnostic Ultrasound Catheters: Biosense…These lots of reprocessed catheters may contain small particles of residual material. If affected devices… 12/22/2025Open, Classified
Z-1325-2026 Class IReprocessed Siemens ACUSON AcuNav Diagnostic Ultrasound Catheters: ACUSON AcuNav…These lots of reprocessed catheters may contain small particles of residual material. If affected devices… 12/22/2025Open, Classified