Recall Z-1325-2026

Medline Industries, LP · initiated December 22, 2025

Class I Open, Classified

Class I – dangerous or defective products that could cause serious health problems or death

Product

Reprocessed Siemens ACUSON AcuNav Diagnostic Ultrasound Catheters: ACUSON AcuNav Ultrasound Catheter, Use on GE Systems 10135910RH; 10043342RH ACUSON AcuNav Ultrasound Catheter, Use on Siemens Systems 10135936RH, 08255790RH

Codes / lots: 10135910RH UDI-DI 10197344026500 Lots EP250324 EP250429 EP250521 EP250609 EP250630 EP250710 EP250905; 10135936RH UDI-DI 10197344026517 Lots EP250324 EP250429 EP250521 EP250609 EP250630 EP250710 EP250905

Reason for recall

These lots of reprocessed catheters may contain small particles of residual material. If affected devices are used, there is a risk of inflammatory response, systemic infection, or granulomatous reaction or thrombus reaction which may lead to cerebral or pulmonary embolism, or deep vein thrombosis. Expansion of recall Z-2610/2614-2025

FDA-determined cause: Under Investigation by firm

Action taken

On December 22, 2025, the firm issued Urgent Medical Device Recall letters as a follow-up to an earlier action in 2025 for the same issue, but newly identified item numbers and lots of previously identified item numbers. Notably, the products associated with the new/updated action are to be destroyed by the customer on-site for credit, and not returned to the manufacturer. ***Updated 5/18/26*** The firm issued an expanded notification on 5/15/26 which expanded the action to include on additional lot of product. These devices were not expired at the time of the original notification but have since expired.

Details

Recall number
Z-1325-2026
Initiated
December 22, 2025
Posted by FDA
February 6, 2026
Product code
OWQ – Reprocessed Intravascular Ultrasound Catheter
Quantity
511
Distribution
Distribution US nationwide.
Recalling firm
Medline Industries, LP, Northfield, IL

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-1319-2026 Class IMedline ReNewal Reprocessed Webster CS Catheters: Webster CS Catheter 10 Electrodes…These lots of reprocessed catheters may contain small particles of residual material. If affected devices… 12/22/2025Open, Classified
Z-1320-2026 Class IReprocessed Abbott Inquiry Steerable Diagnostic Catheters: Abbott Inquiry Steerable…These lots of reprocessed catheters may contain small particles of residual material. If affected devices… 12/22/2025Open, Classified
Z-1321-2026 Class IReprocessed Livewire Steerable Diagnostic Electrophysiology Catheters: Livewire…These lots of reprocessed catheters may contain small particles of residual material. If affected devices… 12/22/2025Open, Classified
Z-1322-2026 Class IReprocessed St. Jude Supreme and Response Diagnostic Electrophysiology Catheters: St.…These lots of reprocessed catheters may contain small particles of residual material. If affected devices… 12/22/2025Open, Classified
Z-1323-2026 Class IMedline ReNewal Reprocessed ViewFlex Xtra ICE Catheters: ViewFlex Xtra ICE Catheter 9F x…These lots of reprocessed catheters may contain small particles of residual material. If affected devices… 12/22/2025Open, Classified
Z-1324-2026 Class IReprocessed Biosense Webster SOUNDSTAR eco Diagnostic Ultrasound Catheters: Biosense…These lots of reprocessed catheters may contain small particles of residual material. If affected devices… 12/22/2025Open, Classified