Recall Z-1413-2025
Baxter Healthcare Corporation · initiated February 25, 2025
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
Baxter Mobile column TruSystem 7500 U, Product Code 1730720
Codes / lots: UID/DI 00887761974388, All serial numbers manufactured until 11/14/2024.
Reason for recall
There is a software issue which causes the upper back section to not be operable/adjustable when the "emergency mode" function is enabled.
FDA-determined cause: Software design
Action taken
Baxter issued an UREGENT MEDICAL DEVICE CORRECTION notice to its consignees on 02/25/2025 by letter sent via traceable means. The notice explained the issue, hazard involved, and requested the following actions be taken: "Actions to be Taken by Customers 1. Baxter will contact you to install a software upgrade in the impacted surgical table systems. 2. If you received this communication directly from Baxter, acknowledge receipt by following the instructions on the enclosed reply form instruction sheet, even if you have no remaining inventory. Acknowledging receipt of this notification will prevent you from receiving repeat notices. If you do not complete the acknowledgement, you will receive a phone call from OnProcess Technology on behalf of Baxter to confirm your receipt of this notification. 3. If you purchased this product from a distributor or wholesaler, please note that responding on the Baxter customer portal is not applicable. If a response is requested by your distributor or wholesaler, please respond to them according to their instructions. 4. Please forward a copy of this communication to the Medical Director/Office Manager, Director of Pharmacy, Medication Safety Officer, Purchasing Director/Head of Central Supply and any other departments within your institution who use the affected product. 5. If you are a dealer, wholesaler, distributor/reseller, or original equipment manufacturer (OEM) that distributed any affected product to other facilities, please notify your customers of this Urgent Medical Device Correction in accordance with your customary procedures and check the associated box on the customer portal." "For general questions regarding this communication, please contact Baxter Technical Support at 800-445-3720 (select option 2, then option 7) between 8:00 am and 6:00 pm Eastern Time, Monday through Friday, or email Baxter at HRC_Trumpf_Tech_Support@baxter.com."
Details
- Recall number
- Z-1413-2025
- Initiated
- February 25, 2025
- Posted by FDA
- March 21, 2025
- Product code
- KXJ – Table, Radiologic
- Quantity
- 40 units
- Distribution
- US Nationwide distribution.
- Recalling firm
- Baxter Healthcare Corporation, Deerfield, IL
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-1409-2025 Class II | Baxter Operating table column TS7500 MOBIUS, Product Code 1704695There is a software issue which causes the upper back section to not be operable/adjustable when the… | 02/25/2025Open, Classified |
| Z-1410-2025 Class II | Baxter Stationary column TruSystem 7500, Product Code 1717020There is a software issue which causes the upper back section to not be operable/adjustable when the… | 02/25/2025Open, Classified |
| Z-1411-2025 Class II | Baxter Floor mounting column TruSystem 7500, Product Code 1717021There is a software issue which causes the upper back section to not be operable/adjustable when the… | 02/25/2025Open, Classified |
| Z-1412-2025 Class II | Baxter Mobile column TruSystem 7500, Product Code 1717023There is a software issue which causes the upper back section to not be operable/adjustable when the… | 02/25/2025Open, Classified |
| Z-1414-2025 Class II | Baxter Stationary column TruSystem 7500 U, Product Code 1730731There is a software issue which causes the upper back section to not be operable/adjustable when the… | 02/25/2025Open, Classified |
| Z-1415-2025 Class II | Baxter Floor mounting column TS 7500 U, Product Code 1730732There is a software issue which causes the upper back section to not be operable/adjustable when the… | 02/25/2025Open, Classified |
| Z-1416-2025 Class II | Baxter TruSystem 7500 Hybrid (MC), Product Code 1773204There is a software issue which causes the upper back section to not be operable/adjustable when the… | 02/25/2025Open, Classified |
| Z-1417-2025 Class II | Baxter TruSystem 7500 Hybrid (FC), Product Code 1854085There is a software issue which causes the upper back section to not be operable/adjustable when the… | 02/25/2025Open, Classified |
| Z-1418-2025 Class II | Baxter TruSystem 7500 Hybrid (SC), Product Code 1854086There is a software issue which causes the upper back section to not be operable/adjustable when the… | 02/25/2025Open, Classified |
| Z-1419-2025 Class II | Baxter TruSystem 7500 Hybrid Plus (FC), Product Code 1854087There is a software issue which causes the upper back section to not be operable/adjustable when the… | 02/25/2025Open, Classified |
| Z-1420-2025 Class II | Baxter TruSystem 7500 Hybrid Plus (SC), Product Code 1854088There is a software issue which causes the upper back section to not be operable/adjustable when the… | 02/25/2025Open, Classified |
| Z-1421-2025 Class II | Baxter TruSystem 7500 Hybrid MR IMRIS, Product Code 2067886There is a software issue which causes the upper back section to not be operable/adjustable when the… | 02/25/2025Open, Classified |
| Z-1422-2025 Class II | Baxter TruSystem 7500, Product Code 4091000There is a software issue which causes the upper back section to not be operable/adjustable when the… | 02/25/2025Open, Classified |