Recall Z-1432-2010
Elekta, Inc. · initiated January 11, 2007
Product
Precise Treatment Table Intended as a universal patient treatment support and positioning table for radiation therapy and simulation use.
Codes / lots: 103101, 105136, 105183, 105223, 105271, 105281, 105292, 105330, 105343, 105370, 105446, 105473, 105511, 105518, 105528, 105532, 105542, 105550, 105568, 151416, 151417, 151420, 151421, 151422, 151423, 151424, 151425, 151426, 151427, 151428, 151429, 151430, 151432, 151433, 151434, 151435, 151436, 151437, 151439, 151440, 151441, 151443, 151444, 151445, 151446, 151447, 151448, 151449, 151450, 151451, 151452, 151453, 151454, 151455, 151456, 151457, 151458, 151459, 151460, 151461, 151462, 151463, 151464, 151466, 151467, 151468, 151469, 151470, 151472, 151473, 151474, 151475, 151476, 151477, 151478, 151479, 151480, 151481, 151483, 151484, 151485, 151486, 151487, 151488, 151489, 151490, 151491, 151492, 151493, 151494, 151495, 151496, 151497, 151499, 151500, 151501, 151502, 151503, 151504, 151505, 151506, 151507, 151508, 151509, 151510, 151511, 151513, 151514, 151515, 151516, 151517, 151518, 151519, 151520, 151522, 151523, 151524, 151525, 151526, 151527, 151528, 151529, 151530, 151531, 151532, 151533, 151534, 151535, 151536, 151537, 151538, 151539, 151540, 151541, 151542, 151543, 151544, 151546, 151547, 151549, 151550, 151551, 151552, 151553, 151554, 151555, 151557, 151558, 151559…
Reason for recall
In some instances, the bolts which are supplied with the table have sheared off, resulting in inability to operate the powered table.
FDA-determined cause: Other
Action taken
Mandatory Field Change Order (FCO) 528 053 "Check for zinc plated Precise Table 'Z' mechanism fixation bolts", dated 1/11/2007 was distributed to all affected customers. The Change Order identified the product, stated the reason for the FCO, and the procedure to check and resolve the problem identified.
Details
- Recall number
- Z-1432-2010
- Initiated
- January 11, 2007
- Posted by FDA
- April 23, 2010
- Terminated
- December 28, 2010
- Product code
- JAI – Couch, Radiation Therapy, Powered
- Quantity
- 185 units
- Distribution
- Worldwide Distribution -- USA, Puerto Rico, Argentina, Australia, Austria, Bangladesh, Belgium, Canada, China, Czech Republic, Denmark, Egypt, France, Germany, Greece, Hong Kong, India, Indonesia, Iraq, Israel, Italy, Japan, Korea, Malaysia, Mexico, Netherlands, Pakistan, Peru, Poland, Russia, Serbia, Spain, Sudan, Sweden, Taiwan, Thailand, Turkey & United Kingdom.
- Recalling firm
- Elekta, Inc., Norcross, GA