Recall Z-1432-2010

Elekta, Inc. · initiated January 11, 2007

Terminated

Product

Precise Treatment Table Intended as a universal patient treatment support and positioning table for radiation therapy and simulation use.

Codes / lots: 103101, 105136, 105183, 105223, 105271, 105281, 105292, 105330, 105343, 105370, 105446, 105473, 105511, 105518, 105528, 105532, 105542, 105550, 105568, 151416, 151417, 151420, 151421, 151422, 151423, 151424, 151425, 151426, 151427, 151428, 151429, 151430, 151432, 151433, 151434, 151435, 151436, 151437, 151439, 151440, 151441, 151443, 151444, 151445, 151446, 151447, 151448, 151449, 151450, 151451, 151452, 151453, 151454, 151455, 151456, 151457, 151458, 151459, 151460, 151461, 151462, 151463, 151464, 151466, 151467, 151468, 151469, 151470, 151472, 151473, 151474, 151475, 151476, 151477, 151478, 151479, 151480, 151481, 151483, 151484, 151485, 151486, 151487, 151488, 151489, 151490, 151491, 151492, 151493, 151494, 151495, 151496, 151497, 151499, 151500, 151501, 151502, 151503, 151504, 151505, 151506, 151507, 151508, 151509, 151510, 151511, 151513, 151514, 151515, 151516, 151517, 151518, 151519, 151520, 151522, 151523, 151524, 151525, 151526, 151527, 151528, 151529, 151530, 151531, 151532, 151533, 151534, 151535, 151536, 151537, 151538, 151539, 151540, 151541, 151542, 151543, 151544, 151546, 151547, 151549, 151550, 151551, 151552, 151553, 151554, 151555, 151557, 151558, 151559…

Reason for recall

In some instances, the bolts which are supplied with the table have sheared off, resulting in inability to operate the powered table.

FDA-determined cause: Other

Action taken

Mandatory Field Change Order (FCO) 528 053 "Check for zinc plated Precise Table 'Z' mechanism fixation bolts", dated 1/11/2007 was distributed to all affected customers. The Change Order identified the product, stated the reason for the FCO, and the procedure to check and resolve the problem identified.

Details

Recall number
Z-1432-2010
Initiated
January 11, 2007
Posted by FDA
April 23, 2010
Terminated
December 28, 2010
Product code
JAI – Couch, Radiation Therapy, Powered
Quantity
185 units
Distribution
Worldwide Distribution -- USA, Puerto Rico, Argentina, Australia, Austria, Bangladesh, Belgium, Canada, China, Czech Republic, Denmark, Egypt, France, Germany, Greece, Hong Kong, India, Indonesia, Iraq, Israel, Italy, Japan, Korea, Malaysia, Mexico, Netherlands, Pakistan, Peru, Poland, Russia, Serbia, Spain, Sudan, Sweden, Taiwan, Thailand, Turkey & United Kingdom.
Recalling firm
Elekta, Inc., Norcross, GA