Recall Z-1488-2020

Philips North America, LLC · initiated December 30, 2019

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Integris Allura 9, system codes 722018 722021

Codes / lots: Serial Codes 8 114 64 906187 41 38 119 75 25 87 67 107 32 97

Reason for recall

A capacitor inside the converter of the Velara X-ray generator may fail after a large number of surges in a short period of time. When this occurs, no image acquisition is possible.

FDA-determined cause: Component design/selection

Action taken

On December 30, 2019, the firm notified customers via Medical Device Correction letter. Customers were informed of the product issue and advised to do the following: If the capacitor of the Velara X-Ray fails as described (capacitor failure after large number of surges in a short period of time), please switch off the system, take it out of service immediately, and contact your local Philips representative. The firm will be resolving the issue by replacing a Printed Circuit Board (PCB) in the convertor, which prevents the capacitor from failing. PCB replacements started in January 2020. Customers will be contacted by their local Philips representatives to schedule a date for the PCB replacement. Customers with further questions should contact the Technical Support Line: 1-800-722-9377.

Details

Recall number
Z-1488-2020
Initiated
December 30, 2019
Terminated
November 17, 2022
Product code
OWB – Interventional Fluoroscopic X-Ray System
Quantity
20
Distribution
Nationwide domestic distribution. Foreign distribution worldwide.
Recalling firm
Philips North America, LLC, Andover, MA

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-1486-2020 Class IIAllura Exper FD Series. System codes 722003 722010 722022 722005 722011 722019 722001…A capacitor inside the converter of the Velara X-ray generator may fail after a large number of surges in a… 12/30/2019Terminated
Z-1487-2020 Class IIIntegris Allura 15 & 12, monoplane (system code 722043) and biplane (system code 722044)A capacitor inside the converter of the Velara X-ray generator may fail after a large number of surges in a… 12/30/2019Terminated
Z-1489-2020 Class IIMultiDiagnost Eleva, System codes 722030 722016 722017 708032 708036 708034 708037 708038A capacitor inside the converter of the Velara X-ray generator may fail after a large number of surges in a… 12/30/2019Terminated
Z-1490-2020 Class IIOmniDiagnost Eleva System codes 708028 708027A capacitor inside the converter of the Velara X-ray generator may fail after a large number of surges in a… 12/30/2019Terminated
Z-1491-2020 Class IIPoly G Integris H5000, System code 72246A capacitor inside the converter of the Velara X-ray generator may fail after a large number of surges in a… 12/30/2019Terminated
Z-1492-2020 Class IIUroDiagnost Eleva, system code 708033A capacitor inside the converter of the Velara X-ray generator may fail after a large number of surges in a… 12/30/2019Terminated
Z-1493-2020 Class IIAllura CV20, system code 722031A capacitor inside the converter of the Velara X-ray generator may fail after a large number of surges in a… 12/30/2019Terminated
Z-1494-2020 Class IICardio Vascular-Allura Centron, system code 722400A capacitor inside the converter of the Velara X-ray generator may fail after a large number of surges in a… 12/30/2019Terminated