Recall Z-1492-2020

Philips North America, LLC · initiated December 30, 2019

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

UroDiagnost Eleva, system code 708033

Codes / lots: Serial numbers 64 48 2 60

Reason for recall

A capacitor inside the converter of the Velara X-ray generator may fail after a large number of surges in a short period of time. When this occurs, no image acquisition is possible.

FDA-determined cause: Component design/selection

Action taken

On December 30, 2019, the firm notified customers via Medical Device Correction letter. Customers were informed of the product issue and advised to do the following: If the capacitor of the Velara X-Ray fails as described (capacitor failure after large number of surges in a short period of time), please switch off the system, take it out of service immediately, and contact your local Philips representative. The firm will be resolving the issue by replacing a Printed Circuit Board (PCB) in the convertor, which prevents the capacitor from failing. PCB replacements started in January 2020. Customers will be contacted by their local Philips representatives to schedule a date for the PCB replacement. Customers with further questions should contact the Technical Support Line: 1-800-722-9377.

Details

Recall number
Z-1492-2020
Initiated
December 30, 2019
Terminated
November 17, 2022
Product code
IXR – Table, Radiographic, Tilting
Quantity
4
Distribution
Nationwide domestic distribution. Foreign distribution worldwide.
Recalling firm
Philips North America, LLC, Andover, MA

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-1486-2020 Class IIAllura Exper FD Series. System codes 722003 722010 722022 722005 722011 722019 722001…A capacitor inside the converter of the Velara X-ray generator may fail after a large number of surges in a… 12/30/2019Terminated
Z-1487-2020 Class IIIntegris Allura 15 & 12, monoplane (system code 722043) and biplane (system code 722044)A capacitor inside the converter of the Velara X-ray generator may fail after a large number of surges in a… 12/30/2019Terminated
Z-1488-2020 Class IIIntegris Allura 9, system codes 722018 722021A capacitor inside the converter of the Velara X-ray generator may fail after a large number of surges in a… 12/30/2019Terminated
Z-1489-2020 Class IIMultiDiagnost Eleva, System codes 722030 722016 722017 708032 708036 708034 708037 708038A capacitor inside the converter of the Velara X-ray generator may fail after a large number of surges in a… 12/30/2019Terminated
Z-1490-2020 Class IIOmniDiagnost Eleva System codes 708028 708027A capacitor inside the converter of the Velara X-ray generator may fail after a large number of surges in a… 12/30/2019Terminated
Z-1491-2020 Class IIPoly G Integris H5000, System code 72246A capacitor inside the converter of the Velara X-ray generator may fail after a large number of surges in a… 12/30/2019Terminated
Z-1493-2020 Class IIAllura CV20, system code 722031A capacitor inside the converter of the Velara X-ray generator may fail after a large number of surges in a… 12/30/2019Terminated
Z-1494-2020 Class IICardio Vascular-Allura Centron, system code 722400A capacitor inside the converter of the Velara X-ray generator may fail after a large number of surges in a… 12/30/2019Terminated