Recall Z-1694-2008

Mallinckrodt Inc · initiated April 7, 2008

Terminated

Product

The OptiVantage DH Injection System is a contrast delivery system and is designed to inject radiopaque contrast media into a patient's vascular system to obtain diagnostic images when used with computed tomography equipment. The J-Bow component is used to connect the powerhead to a suspension system.

Codes / lots: Catalog #V8400, Part No. 241078 & Catalog #V8402, Part #241079

Reason for recall

Screws can begin to loosen and shear off to the point of the J-Bow falling from the suspension system. This may cause injury to patients or hospital staff.

FDA-determined cause: Component design/selection

Action taken

On April 18, 2008, Mallincrodt Inc. issued an Urgent Device Correction letter informing customers to check their inventory, and determine if they have any defective system units. If they observe any loose screws, sheared screws and/or elongated opening(s) around the screws, they are to immediately take the system out of service and contact the firm's service representative. Covidien's customers were all asked to inspect their inventories and to complete an 'Effectiveness Check Return Form' whether or not the product is defective. However, only those system units observed to have loose screws, sheared screws, and/or elongated openings around the screws, will be scheduled to receive correction by a Covidien Service Representative.

Details

Recall number
Z-1694-2008
Initiated
April 7, 2008
Posted by FDA
August 15, 2008
Terminated
May 20, 2011
Product code
IZQ – Injector, Contrast Medium, Automatic
Quantity
783 Units
Distribution
Product is distributed in the US throughout the following states: AL, AR, AZ, CA, CO ,CT, DC, FL, GA, HI, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NJ, NM, NV, NY, OH, OR, PA, RI, SC, SD, TN, TX, VA, WA, WI, and WV.
Recalling firm
Mallinckrodt Inc, Cincinnati, OH

Official FDA recall record