Recall Z-1722-2024
Zimmer, Inc. · initiated March 20, 2024
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
ZPLP (Zimmer Periarticular Locking Plate System) Distal Lateral Fibular Plate, Right, 4 Holes, 80 MM, Nonsterile- Indicated for temporary internal fixation and stabilization of osteotomies and fractures Item Number: 00-2357-017-04
Codes / lots: GTIN: 00889024055865 Lot Numbers: 64370265 64370266 64370267 64370268 64370269 64370270 64370271 64370279 64370280 64370281 64370289 64370290 64370291 64370292 64370293 64370299 64370301 64370303 64370305 64370307 64475227 64475228 64475252 64475276 64475298 64475306 64475320 64475324 64475359 64475360 64475379 64475383 64528443 64528464 64528481 64528482 64528483 64528484 64528485 64528486 64528487 64528488 64528489 64528490 64528491 64528492 64528493 64528494 64528495 64552845 64552846 64552847 64552848 64552892 64552893 64552895 64552896 64552897 64552898 64552899 64552907 64552909 64552912 64552914 64552921 64553013 64553015 64553020 64553022 64553024 64553029 64553032 64553034 64553037 64563320 64563321 64563324 64563325 64563327 64572555 64572556 64572557 64572558 64572559 64572560 64572561 64572562 64572564 64572565 64572566 64572567 64572568 64572569 64572570 64572571 64572572 64572573 64572574 64572576 64572577 64572578 64572579 64572580 64572581 64572582 64572584 64588261 64588262 64588263 64588264 64588265 64588266 64588267 64751424 64751425 64751427 64751429 64751430 64751431 64751432 64751433 64751435 64751437 64751438 64751439 64751441 64751443 64751444 64751446…
Reason for recall
Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The improperly mated screw may not be readily recognizable by the user since the screw may not correctly lock, may result in delay in surgery or loss of fixation resulting further intervention
FDA-determined cause: Under Investigation by firm
Action taken
Zimmer issued Urgent Medical Device Recall Letter to Distributors Risk Managers and Surgeons on March 20, 2024 via mail/email. Letter states reason for recall, health risk and action to take: Our records indicate that you may have received one or more of the affected products. The affected units were distributed between March 2019 and January 2024. Risk Manager Responsibilities: 1. Review this notification and ensure that affected personnel are aware of the contents. 2. If you have affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility. 3. If the product has been further distributed, provide your customers with the recall notice for hospitals or surgeons and ensure documentation. 4. Complete Attachment 1 Certificate of Acknowledgement and send to CorporateQuality.PostMarket@zimmerbiomet.com. This form shall be returned even if you do not have affected products at your facility. 5. Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facility s documentation. 6. If you have further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to CorporateQuality.PostMarket@zimmerbiomet.com. Surgeon Responsibilities: 1. Review this notification for awareness of the contents. 2. There are no specific patient monitoring instructions related to this recall that are recommended beyond your existing follow-up schedule. 3. Complete Attachment 1 Certificate of Acknowledgement and send to CorporateQuality.PostMarket@zimmerbiomet.com. 4. Retain a copy of the acknowledgeme
Details
- Recall number
- Z-1722-2024
- Initiated
- March 20, 2024
- Posted by FDA
- April 30, 2024
- Product code
- HWC – Screw, Fixation, Bone
- Distribution
- Worldwide distribution - US Nationwide and the countries of Brazil, Canada, China, Hong Kong, India, Japan, New Zealand, Seoul Korea, Singapore, Switzerland, Taiwan.
- Recalling firm
- Zimmer, Inc., Warsaw, IN
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-1709-2024 Class II | ZPLP (Zimmer Periarticular Locking Plate System) Distal Lateral Fibular Plate, Right, 4…Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The… | 03/20/2024Open, Classified |
| Z-1710-2024 Class II | ZPLP (Zimmer Periarticular Locking Plate System) Distal Lateral Fibular Plate, Right, 6…Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The… | 03/20/2024Open, Classified |
| Z-1711-2024 Class II | ZPLP (Zimmer Periarticular Locking Plate System) Distal Lateral Fibular Plate, Right, 8…Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The… | 03/20/2024Open, Classified |
| Z-1712-2024 Class II | ZPLP (Zimmer Periarticular Locking Plate System) Distal Lateral Fibular Plate, Right, 10…Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The… | 03/20/2024Open, Classified |
| Z-1713-2024 Class II | ZPLP (Zimmer Periarticular Locking Plate System) Distal Lateral Fibular Plate, Right, 12…Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The… | 03/20/2024Open, Classified |
| Z-1714-2024 Class II | ZPLP (Zimmer Periarticular Locking Plate System) Distal Lateral Fibular Plate, Right, 16…Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The… | 03/20/2024Open, Classified |
| Z-1715-2024 Class II | ZPLP Distal Lateral Fibular Plate, Left, 4 Holes, 80 MM, Sterile- Indicated for…Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The… | 03/20/2024Open, Classified |
| Z-1716-2024 Class II | ZPLP(Zimmer Periarticular Locking Plate System) Distal Lateral Fibular Plate, Left, 6…Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The… | 03/20/2024Open, Classified |
| Z-1717-2024 Class II | ZPLP (Zimmer Periarticular Locking Plate System) Distal Lateral Fibular Plate, Left, 8…Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The… | 03/20/2024Open, Classified |
| Z-1718-2024 Class II | ZPLP(Zimmer Periarticular Locking Plate System) Distal Lateral Fibular Plate, Left, 10…Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The… | 03/20/2024Open, Classified |
| Z-1719-2024 Class II | ZPLP (Zimmer Periarticular Locking Plate System) Distal Lateral Fibular Plate, Left, 12…Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The… | 03/20/2024Open, Classified |
| Z-1720-2024 Class II | ZPLP Distal Lateral Fibular Plate, Left, 14 Holes, 210 MM, Sterile- Indicated for…Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The… | 03/20/2024Open, Classified |
| Z-1721-2024 Class II | ZPLP (Zimmer Periarticular Locking Plate System) Distal Lateral Fibular Plate, Left, 16…Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The… | 03/20/2024Open, Classified |
| Z-1723-2024 Class II | ZPLP Distal Lateral Fibular Plate, Right, 6 Holes, 106 MM, Nonsterile- Indicated for…Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The… | 03/20/2024Open, Classified |
| Z-1724-2024 Class II | ZPLP Distal Lateral Fibular Plate, Right, 8 Holes, 132 MM, Nonsterile- Indicated for…Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The… | 03/20/2024Open, Classified |
| Z-1725-2024 Class II | ZPLP Distal Lateral Fibular Plate, Right, 10 Holes, 158 MM, Nonsterile- Indicated for…Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The… | 03/20/2024Open, Classified |
| Z-1726-2024 Class II | ZPLP Distal Lateral Fibular Plate, Left, 4 Holes, 80 MM, Nonsterile- Indicated for…Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The… | 03/20/2024Open, Classified |
| Z-1727-2024 Class II | ZPLP Distal Lateral Fibular Plate, Left, 6 Holes, 106 MM, Nonsterile- Indicated for…Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The… | 03/20/2024Open, Classified |
| Z-1728-2024 Class II | ZPLP Distal Lateral Fibular Plate, Left, 8 Holes, 132 MM, Nonsterile-Indicated for…Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The… | 03/20/2024Open, Classified |
| Z-1729-2024 Class II | ZPLP Distal Lateral Fibular Plate, Left, 10 Holes, 158 MM, Nonsterile-Indicated for…Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The… | 03/20/2024Open, Classified |