Recall Z-1727-2024
Zimmer, Inc. · initiated March 20, 2024
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
ZPLP Distal Lateral Fibular Plate, Left, 6 Holes, 106 MM, Nonsterile- Indicated for temporary internal fixation and stabilization of osteotomies and fractures Item Number: 00-2357-018-06
Codes / lots: GTIN: 00889024055919 Lot Numbers: 64403506 64403507 64403508 64403509 64403510 64403511 64403512 64403513 64403514 64403515 64403516 64403517 64403518 64403519 64403520 64403521 64403522 64403523 64403524 64403525 64403526 64423743 64423744 64423745 64496313 64496314 64496315 64496316 64496317 64496318 64496319 64496320 64497330 64497331 64497332 64497333 64497334 64497335 64497336 64497337 64497338 64497411 64497412 64497413 64497414 64497423 64497436 64497439 64497446 64497448 64497449 64497450 64497454 64497455 64497458 64497460 64497461 64497462 64538004 64538005 64538006 64538007 64538008 64538009 64538010 64538011 64538012 64538013 64538014 64538015 64538016 64563586 64563589 64563590 64563592 64563611 64563617 64580493 64580494 64580495 64580496 64580498 64580501 64580503 64581249 64581250 64581251 64581252 64581253 64581254 64581255 64599906 64599907 64599908 64599909 64599910 64599911 64599912 64599913 64599914 64599915 64599916 64599917 64599918 64599925 64599926 64599927 64599957 64599958 64599959 64599960 64599961 64599962 64599963 64629444 64629445 64629446 64629447 64629449 64629450 64629468 64629469 64629470 64629473 64629475 64629476 64629477 64629484 64629485…
Reason for recall
Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The improperly mated screw may not be readily recognizable by the user since the screw may not correctly lock, may result in delay in surgery or loss of fixation resulting further intervention
FDA-determined cause: Under Investigation by firm
Action taken
Zimmer issued Urgent Medical Device Recall Letter to Distributors Risk Managers and Surgeons on March 20, 2024 via mail/email. Letter states reason for recall, health risk and action to take: Our records indicate that you may have received one or more of the affected products. The affected units were distributed between March 2019 and January 2024. Risk Manager Responsibilities: 1. Review this notification and ensure that affected personnel are aware of the contents. 2. If you have affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility. 3. If the product has been further distributed, provide your customers with the recall notice for hospitals or surgeons and ensure documentation. 4. Complete Attachment 1 Certificate of Acknowledgement and send to CorporateQuality.PostMarket@zimmerbiomet.com. This form shall be returned even if you do not have affected products at your facility. 5. Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facility s documentation. 6. If you have further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to CorporateQuality.PostMarket@zimmerbiomet.com. Surgeon Responsibilities: 1. Review this notification for awareness of the contents. 2. There are no specific patient monitoring instructions related to this recall that are recommended beyond your existing follow-up schedule. 3. Complete Attachment 1 Certificate of Acknowledgement and send to CorporateQuality.PostMarket@zimmerbiomet.com. 4. Retain a copy of the acknowledgeme
Details
- Recall number
- Z-1727-2024
- Initiated
- March 20, 2024
- Posted by FDA
- April 30, 2024
- Product code
- HWC – Screw, Fixation, Bone
- Distribution
- Worldwide distribution - US Nationwide and the countries of Brazil, Canada, China, Hong Kong, India, Japan, New Zealand, Seoul Korea, Singapore, Switzerland, Taiwan.
- Recalling firm
- Zimmer, Inc., Warsaw, IN
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-1709-2024 Class II | ZPLP (Zimmer Periarticular Locking Plate System) Distal Lateral Fibular Plate, Right, 4…Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The… | 03/20/2024Open, Classified |
| Z-1710-2024 Class II | ZPLP (Zimmer Periarticular Locking Plate System) Distal Lateral Fibular Plate, Right, 6…Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The… | 03/20/2024Open, Classified |
| Z-1711-2024 Class II | ZPLP (Zimmer Periarticular Locking Plate System) Distal Lateral Fibular Plate, Right, 8…Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The… | 03/20/2024Open, Classified |
| Z-1712-2024 Class II | ZPLP (Zimmer Periarticular Locking Plate System) Distal Lateral Fibular Plate, Right, 10…Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The… | 03/20/2024Open, Classified |
| Z-1713-2024 Class II | ZPLP (Zimmer Periarticular Locking Plate System) Distal Lateral Fibular Plate, Right, 12…Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The… | 03/20/2024Open, Classified |
| Z-1714-2024 Class II | ZPLP (Zimmer Periarticular Locking Plate System) Distal Lateral Fibular Plate, Right, 16…Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The… | 03/20/2024Open, Classified |
| Z-1715-2024 Class II | ZPLP Distal Lateral Fibular Plate, Left, 4 Holes, 80 MM, Sterile- Indicated for…Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The… | 03/20/2024Open, Classified |
| Z-1716-2024 Class II | ZPLP(Zimmer Periarticular Locking Plate System) Distal Lateral Fibular Plate, Left, 6…Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The… | 03/20/2024Open, Classified |
| Z-1717-2024 Class II | ZPLP (Zimmer Periarticular Locking Plate System) Distal Lateral Fibular Plate, Left, 8…Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The… | 03/20/2024Open, Classified |
| Z-1718-2024 Class II | ZPLP(Zimmer Periarticular Locking Plate System) Distal Lateral Fibular Plate, Left, 10…Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The… | 03/20/2024Open, Classified |
| Z-1719-2024 Class II | ZPLP (Zimmer Periarticular Locking Plate System) Distal Lateral Fibular Plate, Left, 12…Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The… | 03/20/2024Open, Classified |
| Z-1720-2024 Class II | ZPLP Distal Lateral Fibular Plate, Left, 14 Holes, 210 MM, Sterile- Indicated for…Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The… | 03/20/2024Open, Classified |
| Z-1721-2024 Class II | ZPLP (Zimmer Periarticular Locking Plate System) Distal Lateral Fibular Plate, Left, 16…Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The… | 03/20/2024Open, Classified |
| Z-1722-2024 Class II | ZPLP (Zimmer Periarticular Locking Plate System) Distal Lateral Fibular Plate, Right, 4…Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The… | 03/20/2024Open, Classified |
| Z-1723-2024 Class II | ZPLP Distal Lateral Fibular Plate, Right, 6 Holes, 106 MM, Nonsterile- Indicated for…Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The… | 03/20/2024Open, Classified |
| Z-1724-2024 Class II | ZPLP Distal Lateral Fibular Plate, Right, 8 Holes, 132 MM, Nonsterile- Indicated for…Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The… | 03/20/2024Open, Classified |
| Z-1725-2024 Class II | ZPLP Distal Lateral Fibular Plate, Right, 10 Holes, 158 MM, Nonsterile- Indicated for…Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The… | 03/20/2024Open, Classified |
| Z-1726-2024 Class II | ZPLP Distal Lateral Fibular Plate, Left, 4 Holes, 80 MM, Nonsterile- Indicated for…Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The… | 03/20/2024Open, Classified |
| Z-1728-2024 Class II | ZPLP Distal Lateral Fibular Plate, Left, 8 Holes, 132 MM, Nonsterile-Indicated for…Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The… | 03/20/2024Open, Classified |
| Z-1729-2024 Class II | ZPLP Distal Lateral Fibular Plate, Left, 10 Holes, 158 MM, Nonsterile-Indicated for…Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The… | 03/20/2024Open, Classified |