Recall Z-1836-2016

Intuitive Surgical, Inc. · initiated May 9, 2016

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

SEAL,IS4000 PORTS,5MM-8MM,BOX of 10; Model numbers: 470361-03, 470361-04, 470361-05, 470361-06; Part number: 470361. The da Vinci Xi Surgical System cannula sterile single use disposable cap for each cannula . When attached to the cannula, the seal maintains insufflation when the cannula is empty and when the instrument is inserted in the patient during a surgical procedure.

Codes / lots: All lots

Reason for recall

Potential for particulates in insufflation stopcocks on the da Vinci Xi Surgical System cannula seals. The particulates could be introduced into the cannula lumen when insufflation is connected during a surgical procedure.

FDA-determined cause: Manufacturing material removal

Action taken

Upon discovery of the issue, Intuitive sent Urgent Product Safety Notices to all customers by e-mail to initially alert them of the situation. This will be followed by a formal letter that is scheduled to arrive on May 23, 2016. Both notification request that users Do NOT use the da Vinci Xi 5 mm8 mm Universal Cannula Seal or the da Vinci Xi 12 mm & Stapler Universal Cannula Seal for insufflation, or open the stopcock during intracardiac procedures.

Details

Recall number
Z-1836-2016
Initiated
May 9, 2016
Posted by FDA
May 27, 2016
Terminated
November 1, 2016
Product code
GCY – Apparatus, Suction, Single Patient Use, Portable, Nonpowered
Quantity
46, 191 boxes - total both sizes
Distribution
Worldwide distribution. US, Australia, Austria, Belgium, Chile, Cyprus, Czech Republic, Denmark, Finland, France, Germany, Hong Kong, India, Ireland, Israel, Italy, Japan, Monaco, Netherlands, Norway, Puerto Rico, Portugal, Qatar, Romania, Saudi Arabia, Singapore, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, and United Kingdom.
Recalling firm
Intuitive Surgical, Inc., Sunnyvale, CA

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-1837-2016 Class IISEAL,IS4000 PORTS,12MM&STAPLER,BOX of 10; Model numbers: 470380-04, 470380-05…Potential for particulates in insufflation stopcocks on the da Vinci Xi Surgical System cannula seals. The… 05/09/2016Terminated