Recall Z-1920-2015
Roche Diagnostics Operations, Inc. · initiated May 7, 2015
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
Cholesterol CHOD-PAP. In vitro diagnostic reagent system intended for use on the Hitachi systems for The quantitative determination of total cholesterol in serum and plasma.
Codes / lots: Catalog Numbers: 11491458216; 11875540216; 11875523216.
Reason for recall
Possible drug interference when using the Trinder reaction method on multiple Cobas Analyzers. Results of Trinder tests may be falsely decreased when the tests are run with urine samples (Creatinine plus and Uric Acid applications only) or blood samples collected before sufficient time has lapsed after the administration of the drugs Acetaminophen and N-acetylcysteine.
FDA-determined cause: Under Investigation by firm
Action taken
Roche Diagnostics initiated a voluntary recall and the Urgent Medical Device Correction notices were sent via UPS Ground on 05/07/2015 to customers. Customers were notified of the interference and the method sheets and package inserts will be updated with the new interference claims. The firm plans to monitor any non-responding accounts and address those customers with follow-up attempts to distribute recall notices. The firm plans to conduct level B effectiveness checks via fax and phone. Until the new method sheets are available, the Urgent Medical Device Correction serves as the updated labeling. Roche is not requesting return of the affected product. Customers may contact Todd Siesky Corporate Communications Director, Phone: 317.521.3911, Fax: 317.521.4696, E-mail: todd.siesky@roche.com. For technical questions, contact the Customer Support Center at 1-800-428-2336
Details
- Recall number
- Z-1920-2015
- Initiated
- May 7, 2015
- Posted by FDA
- June 27, 2015
- Terminated
- March 24, 2017
- Product code
- CHH – Enzymatic Esterase--Oxidase, Cholesterol
- Quantity
- 6,255 analytical units distributed
- Distribution
- US (nationwide) including PR.
- Recalling firm
- Roche Diagnostics Operations, Inc., Indianapolis, IN
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-1921-2015 Class II | Cholesterol Gen.2. The cassette COBAS Integra Cholesterol Gen.2 (CHOL2) contains an in…Possible drug interference when using the Trinder reaction method on multiple Cobas Analyzers. Results of… | 05/07/2015Terminated |
| Z-1922-2015 Class II | Creatinine Plus. Intended for use on Hitachi/MODULAR systems for the quantitative…Possible drug interference when using the Trinder reaction method on multiple Cobas Analyzers. Results of… | 05/07/2015Terminated |
| Z-1923-2015 Class II | Creatinine Plus ver. 2. Intended for use on COBAS INTEGRA systems , cobas c111, and…Possible drug interference when using the Trinder reaction method on multiple Cobas Analyzers. Results of… | 05/07/2015Terminated |
| Z-1924-2015 Class II | HDL-Cholesterol gen.3. For the quantitative determination of high-density lipoprotein…Possible drug interference when using the Trinder reaction method on multiple Cobas Analyzers. Results of… | 05/07/2015Terminated |
| Z-1925-2015 Class II | HDL-Cholesterol plus gen.3.; and Small Supersized Intended for use on COBAS Integra 800…Possible drug interference when using the Trinder reaction method on multiple Cobas Analyzers. Results of… | 05/07/2015Terminated |
| Z-1926-2015 Class II | Lactate. Intended for use on COBAS INTEGRA 800 and 400 plus systems, cobas c111…Possible drug interference when using the Trinder reaction method on multiple Cobas Analyzers. Results of… | 05/07/2015Terminated |
| Z-1927-2015 Class II | LDL-Cholesterol gen.2. The Boehringer Mannheim Direct LDL-Cholesterol test in intended…Possible drug interference when using the Trinder reaction method on multiple Cobas Analyzers. Results of… | 05/07/2015Terminated |
| Z-1928-2015 Class II | LDL-Cholesterol plus gen.2. The Boehringer Mannheim Direct LDL-Cholesterol test in…Possible drug interference when using the Trinder reaction method on multiple Cobas Analyzers. Results of… | 05/07/2015Terminated |
| Z-1929-2015 Class II | Triglycerides GPO. In vitro test for the quantitative determination of triglycerides in…Possible drug interference when using the Trinder reaction method on multiple Cobas Analyzers. Results of… | 05/07/2015Terminated |
| Z-1930-2015 Class II | Triglycerides/Glycerol Blanked. In vitro test for the quantitative determination of…Possible drug interference when using the Trinder reaction method on multiple Cobas Analyzers. Results of… | 05/07/2015Terminated |
| Z-1931-2015 Class II | Uric Acid plus ver.2 In vitro test for the quantitative determination of uric acid in…Possible drug interference when using the Trinder reaction method on multiple Cobas Analyzers. Results of… | 05/07/2015Terminated |