Recall Z-1923-2015

Roche Diagnostics Operations, Inc. · initiated May 7, 2015

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Creatinine Plus ver. 2. Intended for use on COBAS INTEGRA systems , cobas c111, and cobas c systems for the quantitative determination of the creatinine concentration in serum, plasma, and urine.

Codes / lots: Catalog Numbers: 05401470190; 03263991190; 05168589190; 03263991190.

Reason for recall

Possible drug interference when using the Trinder reaction method on multiple Cobas Analyzers. Results of Trinder tests may be falsely decreased when the tests are run with urine samples (Creatinine plus and Uric Acid applications only) or blood samples collected before sufficient time has lapsed after the administration of the drugs Acetaminophen and N-acetylcysteine.

FDA-determined cause: Under Investigation by firm

Action taken

Roche Diagnostics initiated a voluntary recall and the Urgent Medical Device Correction notices were sent via UPS Ground on 05/07/2015 to customers. Customers were notified of the interference and the method sheets and package inserts will be updated with the new interference claims. The firm plans to monitor any non-responding accounts and address those customers with follow-up attempts to distribute recall notices. The firm plans to conduct level B effectiveness checks via fax and phone. Until the new method sheets are available, the Urgent Medical Device Correction serves as the updated labeling. Roche is not requesting return of the affected product. Customers may contact Todd Siesky Corporate Communications Director, Phone: 317.521.3911, Fax: 317.521.4696, E-mail: todd.siesky@roche.com. For technical questions, contact the Customer Support Center at 1-800-428-2336

Details

Recall number
Z-1923-2015
Initiated
May 7, 2015
Posted by FDA
June 27, 2015
Terminated
March 24, 2017
Product code
JFY – Enzymatic Method, Creatinine
Quantity
6,255 analytical units distributed
Distribution
US (nationwide) including PR.
Recalling firm
Roche Diagnostics Operations, Inc., Indianapolis, IN

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-1920-2015 Class IICholesterol CHOD-PAP. In vitro diagnostic reagent system intended for use on the Hitachi…Possible drug interference when using the Trinder reaction method on multiple Cobas Analyzers. Results of… 05/07/2015Terminated
Z-1921-2015 Class IICholesterol Gen.2. The cassette COBAS Integra Cholesterol Gen.2 (CHOL2) contains an in…Possible drug interference when using the Trinder reaction method on multiple Cobas Analyzers. Results of… 05/07/2015Terminated
Z-1922-2015 Class IICreatinine Plus. Intended for use on Hitachi/MODULAR systems for the quantitative…Possible drug interference when using the Trinder reaction method on multiple Cobas Analyzers. Results of… 05/07/2015Terminated
Z-1924-2015 Class IIHDL-Cholesterol gen.3. For the quantitative determination of high-density lipoprotein…Possible drug interference when using the Trinder reaction method on multiple Cobas Analyzers. Results of… 05/07/2015Terminated
Z-1925-2015 Class IIHDL-Cholesterol plus gen.3.; and Small Supersized Intended for use on COBAS Integra 800…Possible drug interference when using the Trinder reaction method on multiple Cobas Analyzers. Results of… 05/07/2015Terminated
Z-1926-2015 Class IILactate. Intended for use on COBAS INTEGRA 800 and 400 plus systems, cobas c111…Possible drug interference when using the Trinder reaction method on multiple Cobas Analyzers. Results of… 05/07/2015Terminated
Z-1927-2015 Class IILDL-Cholesterol gen.2. The Boehringer Mannheim Direct LDL-Cholesterol test in intended…Possible drug interference when using the Trinder reaction method on multiple Cobas Analyzers. Results of… 05/07/2015Terminated
Z-1928-2015 Class IILDL-Cholesterol plus gen.2. The Boehringer Mannheim Direct LDL-Cholesterol test in…Possible drug interference when using the Trinder reaction method on multiple Cobas Analyzers. Results of… 05/07/2015Terminated
Z-1929-2015 Class IITriglycerides GPO. In vitro test for the quantitative determination of triglycerides in…Possible drug interference when using the Trinder reaction method on multiple Cobas Analyzers. Results of… 05/07/2015Terminated
Z-1930-2015 Class IITriglycerides/Glycerol Blanked. In vitro test for the quantitative determination of…Possible drug interference when using the Trinder reaction method on multiple Cobas Analyzers. Results of… 05/07/2015Terminated
Z-1931-2015 Class IIUric Acid plus ver.2 In vitro test for the quantitative determination of uric acid in…Possible drug interference when using the Trinder reaction method on multiple Cobas Analyzers. Results of… 05/07/2015Terminated