Recall Z-1991-2020
Covidien Llc · initiated April 30, 2020
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
Dover 100% Silicone Silver-Hydrogel Coated Premium Urine Meter Foley Tray 400mL, Premium, Urine Meter, 16 Fr/Ch (5.3 mm), 5 mL Catheter Pre-connected to 2000 mL Drainage Bag with Luer-Lock Sampling, Temperature Sensor and Adhesive Catheter Securement, Sterile, Product Code P4P16XTSD - Product Usage: for Drainage of the Urinary bladder and simultaneous monitoring of temperature. Prescription use only.
Codes / lots: Lot Numbers: 1828205964, 1911414164
Reason for recall
Temperature sensing catheters failed to meet the established calibrated temperature sensing tolerance of +/- 0.2C. Failure to correctly measure patient body temperatures could lead to an incorrect diagnosis and insufficient treatment
FDA-determined cause: Nonconforming Material/Component
Action taken
Cardinal Health issued letter Urgent Medical Device Recall EVENT-2020-02846 on 4/30/20 via overnight mail. Letter states reason for recall, health risk, and action to take: 1. CHECK all storage and usage locations to confirm whether you have any units of the affected product codes and lot numbers 2. SEGREGATE and QUARANTINE all on-hand product that is confirmed to be labeled per Exhibit A. 3. PLEASE RETURN the enclosed acknowledgment form via facsimile (847-689-9101 or 614-652-9648) or email (GMB-FieldCorrectiveAction@cardinalhealth.com) and indicate the product code, lot and quantity of product youve quarantined or discarded. Please respond regardless of whether or not you have affected product. 4. NOTIFY any customers to whom you may have distributed, or forwarded product affected by this recall. Your notification to your customers may be enhanced by including a copy of this recall notification letter. 5. CONTACT the appropriate Customer Service group to arrange for return and credit/replacement of any affected product Monday Friday between 8:00am - 5:00pm EST: Hospital800-964-5227 Federal Government800-444-1166 Distributor800-635-6021 All other Customers888-444-5440 Questions contact: Cardinal Health 800-292-9332
Details
- Recall number
- Z-1991-2020
- Initiated
- April 30, 2020
- Terminated
- December 14, 2023
- Product code
- KOD – Catheter, Urological
- Quantity
- 46828 units
- Distribution
- US Nationwide distribution.
- Recalling firm
- Covidien Llc, Mansfield, MA
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-1984-2020 Class II | Dover 100% Silicone Foley Catheter with Temperature Sensor 14FR 5CC2W, Product Code:…Temperature sensing catheters failed to meet the established calibrated temperature sensing tolerance of +/-… | 04/30/2020Terminated |
| Z-1985-2020 Class II | Dover 100% Silicone Silver-Hydrogel Coated Foley Catheter with Temperature Sensor 16FR…Temperature sensing catheters failed to meet the established calibrated temperature sensing tolerance of +/-… | 04/30/2020Terminated |
| Z-1986-2020 Class II | Dover 100% Silicone Foley Catheter with Temperature 16FR 5CC2W, Product Code: 50516T -…Temperature sensing catheters failed to meet the established calibrated temperature sensing tolerance of +/-… | 04/30/2020Terminated |
| Z-1987-2020 Class II | Dover 100% Silicone Silver-Hydrogel Coated Foley Catheter with Temperature Sensor 18FR…Temperature sensing catheters failed to meet the established calibrated temperature sensing tolerance of +/-… | 04/30/2020Terminated |
| Z-1988-2020 Class II | Dover 100% Silicone Premium Urine Meter Foley Tray, 400mL, 14 Fr/Ch (4.7 mm), 5 mL…Temperature sensing catheters failed to meet the established calibrated temperature sensing tolerance of +/-… | 04/30/2020Terminated |
| Z-1989-2020 Class II | Dover 100% Silicone Premium Urine Meter Tray 400mL, 16 Fr/Ch (5.3 mm), 5 mL Foley…Temperature sensing catheters failed to meet the established calibrated temperature sensing tolerance of +/-… | 04/30/2020Terminated |
| Z-1990-2020 Class II | Dover 400mL, 100% Silicone Premium Urine Meter Foley Tray, 16 Fr/Ch (5.3 mm), 5 mL…Temperature sensing catheters failed to meet the established calibrated temperature sensing tolerance of +/-… | 04/30/2020Terminated |
| Z-1992-2020 Class II | Dover 100% Silicone Silver-Hydrogel Coated Premium Urine Meter Foley Tray 400mL, 16…Temperature sensing catheters failed to meet the established calibrated temperature sensing tolerance of +/-… | 04/30/2020Terminated |
| Z-1993-2020 Class II | Dover 400 ml 100% Silicone Silver-Hydrogel Coated Premium Urine Meter Foley Tray,18…Temperature sensing catheters failed to meet the established calibrated temperature sensing tolerance of +/-… | 04/30/2020Terminated |
| Z-1994-2020 Class II | Dover 400 ml 100% Silicone Silver-Hydrogel Coated Premium Urine Meter Foley Tray, 18…Temperature sensing catheters failed to meet the established calibrated temperature sensing tolerance of +/-… | 04/30/2020Terminated |