Recall Z-1993-2020

Covidien Llc · initiated April 30, 2020

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Dover 400 ml 100% Silicone Silver-Hydrogel Coated Premium Urine Meter Foley Tray,18 Fr/Ch (6.0 mm), 5 mL Catheter Pre-connected to 2000 mL Drainage Bag with Temperature Sensor, Sterile, Product Code: P4P18XTS - Product Usage: for Drainage of the Urinary bladder and simultaneous monitoring of temperature. Prescription use only.

Codes / lots: Lot Numbers: 1919708264

Reason for recall

Temperature sensing catheters failed to meet the established calibrated temperature sensing tolerance of +/- 0.2C. Failure to correctly measure patient body temperatures could lead to an incorrect diagnosis and insufficient treatment

FDA-determined cause: Nonconforming Material/Component

Action taken

Cardinal Health issued letter Urgent Medical Device Recall EVENT-2020-02846 on 4/30/20 via overnight mail. Letter states reason for recall, health risk, and action to take: 1. CHECK all storage and usage locations to confirm whether you have any units of the affected product codes and lot numbers 2. SEGREGATE and QUARANTINE all on-hand product that is confirmed to be labeled per Exhibit A. 3. PLEASE RETURN the enclosed acknowledgment form via facsimile (847-689-9101 or 614-652-9648) or email (GMB-FieldCorrectiveAction@cardinalhealth.com) and indicate the product code, lot and quantity of product youve quarantined or discarded. Please respond regardless of whether or not you have affected product. 4. NOTIFY any customers to whom you may have distributed, or forwarded product affected by this recall. Your notification to your customers may be enhanced by including a copy of this recall notification letter. 5. CONTACT the appropriate Customer Service group to arrange for return and credit/replacement of any affected product Monday Friday between 8:00am - 5:00pm EST: Hospital800-964-5227 Federal Government800-444-1166 Distributor800-635-6021 All other Customers888-444-5440 Questions contact: Cardinal Health 800-292-9332

Details

Recall number
Z-1993-2020
Initiated
April 30, 2020
Terminated
December 14, 2023
Product code
KOD – Catheter, Urological
Quantity
120 units
Distribution
US Nationwide distribution.
Recalling firm
Covidien Llc, Mansfield, MA

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-1984-2020 Class IIDover 100% Silicone Foley Catheter with Temperature Sensor 14FR 5CC2W, Product Code:…Temperature sensing catheters failed to meet the established calibrated temperature sensing tolerance of +/-… 04/30/2020Terminated
Z-1985-2020 Class IIDover 100% Silicone Silver-Hydrogel Coated Foley Catheter with Temperature Sensor 16FR…Temperature sensing catheters failed to meet the established calibrated temperature sensing tolerance of +/-… 04/30/2020Terminated
Z-1986-2020 Class IIDover 100% Silicone Foley Catheter with Temperature 16FR 5CC2W, Product Code: 50516T -…Temperature sensing catheters failed to meet the established calibrated temperature sensing tolerance of +/-… 04/30/2020Terminated
Z-1987-2020 Class IIDover 100% Silicone Silver-Hydrogel Coated Foley Catheter with Temperature Sensor 18FR…Temperature sensing catheters failed to meet the established calibrated temperature sensing tolerance of +/-… 04/30/2020Terminated
Z-1988-2020 Class IIDover 100% Silicone Premium Urine Meter Foley Tray, 400mL, 14 Fr/Ch (4.7 mm), 5 mL…Temperature sensing catheters failed to meet the established calibrated temperature sensing tolerance of +/-… 04/30/2020Terminated
Z-1989-2020 Class IIDover 100% Silicone Premium Urine Meter Tray 400mL, 16 Fr/Ch (5.3 mm), 5 mL Foley…Temperature sensing catheters failed to meet the established calibrated temperature sensing tolerance of +/-… 04/30/2020Terminated
Z-1990-2020 Class IIDover 400mL, 100% Silicone Premium Urine Meter Foley Tray, 16 Fr/Ch (5.3 mm), 5 mL…Temperature sensing catheters failed to meet the established calibrated temperature sensing tolerance of +/-… 04/30/2020Terminated
Z-1991-2020 Class IIDover 100% Silicone Silver-Hydrogel Coated Premium Urine Meter Foley Tray 400mL…Temperature sensing catheters failed to meet the established calibrated temperature sensing tolerance of +/-… 04/30/2020Terminated
Z-1992-2020 Class IIDover 100% Silicone Silver-Hydrogel Coated Premium Urine Meter Foley Tray 400mL, 16…Temperature sensing catheters failed to meet the established calibrated temperature sensing tolerance of +/-… 04/30/2020Terminated
Z-1994-2020 Class IIDover 400 ml 100% Silicone Silver-Hydrogel Coated Premium Urine Meter Foley Tray, 18…Temperature sensing catheters failed to meet the established calibrated temperature sensing tolerance of +/-… 04/30/2020Terminated