Recall Z-2175-2019

Ortho Clinical Diagnostics · initiated July 8, 2019

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

VITROS 250 Chemistry System, clinical chemistry analyzer

Codes / lots: product code: 6801759 (Refurbished) UDI: 10758750001330 Affected software : Version 9.7

Reason for recall

Potential Unsuccessful Calibration Due to Software Anomaly in VITROS System

FDA-determined cause: Software design

Action taken

Customers were notified prior to July 10, 2019. For the VITROS 250, 250AT, and 350 chemistry system: Do not use VITROS Calibrator Kit 32 to calibrate assays until an Ortho trained service personnel reconfigures your system(s) with the correct parameters for VITROS Calibrator Kit 32. Complete the enclosed Confirmation of Receipt form no later than July 15, 2019. Post this notification by each system that may use VITROS Calibrator Kit 32 until your system is reconfigured by an Ortho trained representative. Please forward this notification if the product was distributed outside of your facility. For the VITROS 4600,5600 and 5,1 FS chemistry System: If your system is configured with User Modified parameters, choose an option provided in the Interim Resolutions section listed above. Install ADD DRV 6040 or above upon availability. Complete the enclosed Confirmation of Receipt form no later than July 15, 2019. Post this notification either by each system that processes VITROS Calibrator Kit 32 or with your user documentation until DRV 6040 or above is installed on your system(s). Please forward this notification if the product was distributed outside of your facility. If you have further questions, please contact Ortho Care Technical Solutions Center at 1-800-421-3311.

Details

Recall number
Z-2175-2019
Initiated
July 8, 2019
Terminated
October 19, 2023
Product code
JJE – Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Quantity
2
Distribution
Worldwide distribution. US nationwide including Puerto Rico, Brazil, Canada, Chile, France, India, and Philippines.
Recalling firm
Ortho Clinical Diagnostics, Rochester, NY

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-2174-2019 Class IIVITROS 250 Chemistry System, clinical chemistry analyzerPotential Unsuccessful Calibration Due to Software Anomaly in VITROS System 07/08/2019Terminated
Z-2176-2019 Class IIVITROS 250AT Chemistry System, clinical chemistry analyzerPotential Unsuccessful Calibration Due to Software Anomaly in VITROS System 07/08/2019Terminated
Z-2177-2019 Class IIVITROS 350 Chemistry System, clinical chemistry analyzerPotential Unsuccessful Calibration Due to Software Anomaly in VITROS System 07/08/2019Terminated
Z-2178-2019 Class IIVITROS Chemistry Products Calibrator Kit 32, for use with VITROS clinical chemistry…Potential Unsuccessful Calibration Due to Software Anomaly in VITROS System 07/08/2019Terminated
Z-2179-2019 Class IIVITROS 5600 Integrated System, clinical chemistry analyzerPotential Unsuccessful Calibration Due to Software Anomaly in VITROS System 07/08/2019Terminated
Z-2180-2019 Class IIVITROS 5600 Integrated System, clinical chemistry analyzerPotential Unsuccessful Calibration Due to Software Anomaly in VITROS System 07/08/2019Terminated
Z-2181-2019 Class IIVITROS 4600 Chemistry System, clinical chemistry analyzerPotential Unsuccessful Calibration Due to Software Anomaly in VITROS System 07/08/2019Terminated
Z-2182-2019 Class IIVITROS 4600 Chemistry System, clinical chemistry analyzerPotential Unsuccessful Calibration Due to Software Anomaly in VITROS System 07/08/2019Terminated
Z-2183-2019 Class IIVITROS 5,1 FS Chemistry System, clinical chemistry analyzerPotential Unsuccessful Calibration Due to Software Anomaly in VITROS System 07/08/2019Terminated
Z-2184-2019 Class IIVITROS 5,1 FS Chemistry System, clinical chemistry analyzerPotential Unsuccessful Calibration Due to Software Anomaly in VITROS System 07/08/2019Terminated
Z-2185-2019 Class IIVITROS Chemistry Products Calibrator Kit 32, for use with VITROS clinical chemistry…Potential Unsuccessful Calibration Due to Software Anomaly in VITROS System 07/08/2019Terminated