Recall Z-2177-2019
Ortho Clinical Diagnostics · initiated July 8, 2019
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
VITROS 350 Chemistry System, clinical chemistry analyzer
Codes / lots: product code: 6802153 UDI: 10758750002054 Affected software : Version 9.7
Reason for recall
Potential Unsuccessful Calibration Due to Software Anomaly in VITROS System
FDA-determined cause: Software design
Action taken
Customers were notified prior to July 10, 2019. For the VITROS 250, 250AT, and 350 chemistry system: Do not use VITROS Calibrator Kit 32 to calibrate assays until an Ortho trained service personnel reconfigures your system(s) with the correct parameters for VITROS Calibrator Kit 32. Complete the enclosed Confirmation of Receipt form no later than July 15, 2019. Post this notification by each system that may use VITROS Calibrator Kit 32 until your system is reconfigured by an Ortho trained representative. Please forward this notification if the product was distributed outside of your facility. For the VITROS 4600,5600 and 5,1 FS chemistry System: If your system is configured with User Modified parameters, choose an option provided in the Interim Resolutions section listed above. Install ADD DRV 6040 or above upon availability. Complete the enclosed Confirmation of Receipt form no later than July 15, 2019. Post this notification either by each system that processes VITROS Calibrator Kit 32 or with your user documentation until DRV 6040 or above is installed on your system(s). Please forward this notification if the product was distributed outside of your facility. If you have further questions, please contact Ortho Care Technical Solutions Center at 1-800-421-3311.
Details
- Recall number
- Z-2177-2019
- Initiated
- July 8, 2019
- Terminated
- October 19, 2023
- Product code
- JJE – Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
- Quantity
- 30
- Distribution
- Worldwide distribution. US nationwide including Puerto Rico, Brazil, Canada, Chile, France, India, and Philippines.
- Recalling firm
- Ortho Clinical Diagnostics, Rochester, NY
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-2174-2019 Class II | VITROS 250 Chemistry System, clinical chemistry analyzerPotential Unsuccessful Calibration Due to Software Anomaly in VITROS System | 07/08/2019Terminated |
| Z-2175-2019 Class II | VITROS 250 Chemistry System, clinical chemistry analyzerPotential Unsuccessful Calibration Due to Software Anomaly in VITROS System | 07/08/2019Terminated |
| Z-2176-2019 Class II | VITROS 250AT Chemistry System, clinical chemistry analyzerPotential Unsuccessful Calibration Due to Software Anomaly in VITROS System | 07/08/2019Terminated |
| Z-2178-2019 Class II | VITROS Chemistry Products Calibrator Kit 32, for use with VITROS clinical chemistry…Potential Unsuccessful Calibration Due to Software Anomaly in VITROS System | 07/08/2019Terminated |
| Z-2179-2019 Class II | VITROS 5600 Integrated System, clinical chemistry analyzerPotential Unsuccessful Calibration Due to Software Anomaly in VITROS System | 07/08/2019Terminated |
| Z-2180-2019 Class II | VITROS 5600 Integrated System, clinical chemistry analyzerPotential Unsuccessful Calibration Due to Software Anomaly in VITROS System | 07/08/2019Terminated |
| Z-2181-2019 Class II | VITROS 4600 Chemistry System, clinical chemistry analyzerPotential Unsuccessful Calibration Due to Software Anomaly in VITROS System | 07/08/2019Terminated |
| Z-2182-2019 Class II | VITROS 4600 Chemistry System, clinical chemistry analyzerPotential Unsuccessful Calibration Due to Software Anomaly in VITROS System | 07/08/2019Terminated |
| Z-2183-2019 Class II | VITROS 5,1 FS Chemistry System, clinical chemistry analyzerPotential Unsuccessful Calibration Due to Software Anomaly in VITROS System | 07/08/2019Terminated |
| Z-2184-2019 Class II | VITROS 5,1 FS Chemistry System, clinical chemistry analyzerPotential Unsuccessful Calibration Due to Software Anomaly in VITROS System | 07/08/2019Terminated |
| Z-2185-2019 Class II | VITROS Chemistry Products Calibrator Kit 32, for use with VITROS clinical chemistry…Potential Unsuccessful Calibration Due to Software Anomaly in VITROS System | 07/08/2019Terminated |