Recall Z-2232-2008
Hamilton Co · initiated May 7, 2008
Product
Hamilton Soft Grip Pipettes, 1 ml fixed volume, model number 55019-19 Product is indicated for manual pipetting.
Codes / lots: serial numbers 64133, 64137, 64139, 64145, 64148, 64156, 64163, 64171, 64175, 64180, 64189, 64192, 64193, 64193, 64198, 64199, 64200, 64202, 64203, 64205, 64211, 64212, 64216, 64224, 64225, 64921, 65072, 65081, 65099, 65105, 65107, 65110, 65140, 65141, 65151, 65159, 65160, 65163, 65164, 65172, 65486, 65501, 65517, 65540, 65585, 65599, 65619, 65629, 65658, 65699, 65934, 65951, 65969, 65972, 65992, 66764, 66783, 66788, 66793, 66892, 66897, 66902, 66972, 66982, 66991, 66999, 67001, 67031, 67035, 67044, 67054, 67055, 67059, 67082, 67084, 67091, 67094, 67098, 67120, 67130, 67145, 67162, 67167, 67170, 67175, 67176, 67190, 67191, 67198, 67224, 67237, 67247, 67258, 67260, 67269, 67278, 67280, 67282, 67283, and 67285
Reason for recall
Manufacturing issue with spring may restrict product to drawing less than 1/2 of its stated measured capacity
FDA-determined cause: Device Design
Action taken
Recall initiated May 7 2008. Notification issued via certified overnight mail explains the problem and instructs customers to inventory the affected pipettes. The firm will contact its customers and supply replacement lower-body assemblies, repair instructions and return information so that affected parts can be returned to Hamilton. A fax-back form is supplied with each notification. Responses will be tracked. Contact Hamilton Co. at 1-775-558-3000 for assistance.
Details
- Recall number
- Z-2232-2008
- Initiated
- May 7, 2008
- Posted by FDA
- September 17, 2008
- Terminated
- February 8, 2011
- Product code
- JQW – Station, Pipetting And Diluting, For Clinical Use
- Quantity
- 48 units
- Distribution
- Product distributed throughout the US and to Puerto Rico, New Zealand, Australia, Japan and Canada.
- Recalling firm
- Hamilton Co, Reno, NV
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-2233-2008 | Hamilton Soft Grip Pipettes, 1 ml adjustable volume, model number 55019-40 Product is…Manufacturing issue with spring may restrict product to drawing less than 1/2 of its stated measured capacity | 05/07/2008Terminated |