Recall Z-2286-2019
Luminex Corporation · initiated February 20, 2019
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
ARIES Group A Strep Assay , REF 50-10041, UDI # 00840487101469
Codes / lots: Lot numbers & discontinue use by dates for Part Number 50-10041: AA7313 - Immediately; AA7526 - 3/7/2019; AA7605 - 3/9/2019; AA7631 - 4/3/2019; AA7645 - 4/12/2019
Reason for recall
Through internal investigation, it was discovered that a portion of side cassette assemblies (SCAs) back side seal (BSS) did not meet design specification, resulting in partially sealed SCAs. The impact of a partial seal of the BSS could result in potential decrease of volume for each solution in the SCAs and decreased sensitivity or stability of the assay.
FDA-determined cause: Process control
Action taken
On February 20, 2019, the firm sent a "Urgent: Medical Device Recall/Field Safety Notice" via email to all consignees. The customer notification informed customers of the following: 1. As an immediate action, Luminex is requesting that customers discontinue use of and discard the products listedin the recall notice immediately or by the Discontinue Use By date. Contact Luminex Global Support Services to obtain replacement material. Material that is still within its Discontinue Use By date may be used at the customers discretion. 2. Complete the Acknowledgment and Receipt form. You must complete and return this form even if you do not have any remaining product. Luminex Technical Support can assist you in completing this form. This information is essential in order to maintain recall effectiveness information required by US FDA. 3. Although no adverse events have been reported, adverse reaction or quality problems experienced with the use of this product may be reported to the FDAs MedWatch Adverse Event Reporting program either online, by regular mail or by fax. 4. Please call the Luminex Global Support Services Team if you have any questions or concerns. Luminex Global Support Services 1-877-785-2323 (U.S. and Canada) +1-512-381-4397 (Outside U.S. and Canada) support@luminexcorp.com CAN-0233
Details
- Recall number
- Z-2286-2019
- Initiated
- February 20, 2019
- Posted by FDA
- August 15, 2019
- Product code
- PGX – Groups A, C And G Beta-Hemolytic Streptococcus Nucleic Acid Amplification System
- Distribution
- US: AK, AL, AZ,CA,CO,FL,GA,IL,KS,KY,LA,MA,MD,ME,MI,MN,MO,MS,MT,NC,NE,NJ,NM,NY,OH,ON,OR,PA,TN,TX,UT,VA,VT,WA,WI, OUS: Belgium, Canada, Thailand, Germany, Singapore, United Kingdom
- Recalling firm
- Luminex Corporation, Austin, TX
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-2282-2019 Class II | Aries Bordetella Assay, REF 50-10037, UDI # 00840487101452Through internal investigation, it was discovered that a portion of side cassette assemblies (SCAs) back… | 02/20/2019Open, Classified |
| Z-2283-2019 Class II | ARIES HSV 1&2 Assay , REF 50-10017, UDI # 00840487100295Through internal investigation, it was discovered that a portion of side cassette assemblies (SCAs) back… | 02/20/2019Open, Classified |
| Z-2284-2019 Class II | ARIES Flu A/B & RSV Assay , REF 50-10020, UDI # 00840487100158Through internal investigation, it was discovered that a portion of side cassette assemblies (SCAs) back… | 02/20/2019Open, Classified |
| Z-2285-2019 Class II | ARIES GBS Assay , REF 50-10021, UDI # 00840487100165Through internal investigation, it was discovered that a portion of side cassette assemblies (SCAs) back… | 02/20/2019Open, Classified |
| Z-2287-2019 Class II | ARIES C. difficile Assay , REF 50-10018, UDI # 00840487100059Through internal investigation, it was discovered that a portion of side cassette assemblies (SCAs) back… | 02/20/2019Open, Classified |
| Z-2288-2019 Class II | ARIES Extraction Kit; ARIES System; SYNCT , REF 50-10026, UDI # 00840487100073Through internal investigation, it was discovered that a portion of side cassette assemblies (SCAs) back… | 02/20/2019Open, Classified |